Olympus WB91051WRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus electrosurgical unit cuts and seals tissue using electrical energy. Many reports describe electrical faults, communication problems, and output or activation failures, and some mention burns or bleeding. The records include 1,920 reports, 135 injuries and 1 death, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,891
- Deaths reported
- 1
- Injuries reported
- 135
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,891 adverse event reports that list this exact device (DI 04042761076838), received between Aug 6, 2018 and Aug 11, 2026 — classified as 1 death, 135 injuries, 1,755 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (597 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem597
- Communication or Transmission Problem541
- Output Problem359
- Activation Problem228
- Adverse Event Without Identified Device or Use Problem60
- Image Display Error/Artifact57
- Electrical Shorting49
- Unexpected Shutdown48
Patient problems
- No Clinical Signs, Symptoms or Conditions1,683
- No Consequences Or Impact To Patient62
- Hemorrhage/Blood Loss/Bleeding27
- Insufficient Information24
- No Patient Involvement24
- Burn(s)19
- No Known Impact Or Consequence To Patient18
- Perforation11
Recent adverse event reports
The 25 most recent of 1,891 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Communication or Transmission Problem
No additional information reported from the customer.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50204554 ↗MalfunctionAug 6, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50168169 ↗MalfunctionAug 5, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50143768 ↗MalfunctionAug 3, 2026Circuit Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50118370 ↗MalfunctionAug 3, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50113717 ↗MalfunctionJun 29, 2026Electrical /Electronic Property Problem
THE CUSTOMER RETURNED THE GENERATOR TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION IT WAS FOUND THAT THE DEVICE HAD AN ERROR E433. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25659390 ↗MalfunctionJun 25, 2026Electrical /Electronic Property Problem
THE GENERATOR WAS RETURNED TO THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, THE ERROR CODE E433 WAS FOUND BECAUSE THE HIGH VOLTAGE POWER SUPPLY BOARD (HVPS) AND THE GENERATOR BOARD WERE DEFECTIVE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25628438 ↗MalfunctionJun 25, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR TO OLYMPUS SERVICE CENTER AS AN ASSET RETURN WITH NO REPORTED ISSUE. DURING THE DEVICE INSPECTION, THE SERVICE REPAIR TEAM INDICATED THAT AN ERROR E433 WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25626779 ↗MalfunctionJun 24, 2026Communication or Transmission Problem; Output Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR ESG-400 TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE FIELD SERVICE ENGINEER (FSE), IDENTIFIED THAT AN ERROR E006 WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25613326 ↗MalfunctionJun 24, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE GENERATOR TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, ERROR CODE E433 WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25608517 ↗MalfunctionJun 24, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE GENERATOR TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, ERROR CODE E486 WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25607728 ↗MalfunctionJun 19, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE GENERATOR TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE FIELD SERVICE ENGINEER (FSE) INDICATED THAT AN E433 ERROR WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25564462 ↗MalfunctionJun 18, 2026Electrical /Electronic Property Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR ESG-400 TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE REGIONAL REPAIR CENTER (RAC) INDICATED THAT ERROR 433 WAS FOUND. IT WAS DETERMINED THAT THE HIGH VOLTAGE POWER SUPPLY BOARD NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25552838 ↗MalfunctionJun 17, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS THE SERVICE REPAIR TECHNICIAN INDICATED THAT A RANDOM ERROR E433 WAS DETECTED IN THE ERROR HISTORY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25535592 ↗MalfunctionJun 10, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR ESG-400 TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE FIELD SERVICE ENGINEER IDENTIFIED THE DEVICE HAD AN E433 ERROR CODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25463197 ↗MalfunctionJun 8, 2026Communication or Transmission Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR ESG-400 TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING INSPECTION THE DEVICE EXHIBITED AN ERROR E006. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25438270 ↗MalfunctionJun 3, 2026Electrical /Electronic Property Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, ERROR E090 OCCURRENCE HISTORY HAD BEEN RECORDED MULTIPLE TIMES IN THE ERROR LOG WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE E090 WAS DUE TO A COMPONENT FAILURE OF THE INTERNAL ELECTRONIC BOARD. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25376418 ↗MalfunctionMay 29, 2026Electrical /Electronic Property Problem
THE CUSTOMER RETURNED THE ELECTROSURGICAL GENERATOR ESG-400 TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT AN ERROR E090 WAS FOUND. IT WAS DETERMINED THAT THE GENERATOR BOARD NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25338084 ↗MalfunctionMay 29, 2026Electrical /Electronic Property Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25337482 ↗MalfunctionMay 29, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE ELECTROSURGICAL GENERATOR DISPLAYED E486 IN ERROR LOG. THE ISSUE OCCURRED INSPECTION OF THE DEVICE FOR USE BEFORE PATIENT IN A ROOM. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333949 ↗MalfunctionMay 29, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE ELECTROSURGICAL GENERATOR ESG-400 HAD AN E006 ERROR. THE ISSUE WAS OBSERVED DURING SET UP/INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333759 ↗MalfunctionMay 28, 2026Electrical /Electronic Property Problem
THE ELECTROSURGICAL GENERATOR ESG-400 WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A BEEPING SOUND WHEN THE FOOTSWITCH WAS ACTIVATED. THIS ISSUE DOES NOT CONSTITUTE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS IDENTIFIED DURING TESTING AT THE SERVICE CENTER, WHERE ERROR CODE E433 WAS OBSERVED DUE TO DEFECTIVE FOOTSWITCH CABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25320277 ↗MalfunctionMay 27, 2026Communication or Transmission Problem
THE ELECTROSURGICAL GENERATOR ESG-400 WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF ERROR E197. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, AN ERROR CODE E433 DUE TO A FAILURE ON THE HIGH VOLTAGE POWER SUPPLY BOARD/POWER BOARD WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25308431 ↗MalfunctionMay 27, 2026Communication or Transmission Problem
THE CUSTOMER DID NOT REPORT A MALFUNCTION. HOWEVER, DURING THE INSPECTION OF THE DEVICE, A 433 ERROR CODE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25302085 ↗MalfunctionMay 21, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE ELECTROSURGICAL GENERATOR DISPLAYED AN E006 ERROR (INSUFFICIENT CONDUCTIVITY). THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE DURING SEDATION WHILE DEVICE INSIDE THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH NO DELAY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25243297 ↗
FDA clearance and classification
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Olympus WB91051W
Has the Olympus WB91051W been recalled?
No FDA recall record in our data names this exact device identifier (DI 04042761076838), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Olympus WB91051W?
The FDA's MAUDE database lists 1,891 adverse event reports for this device, received between Aug 6, 2018 and Aug 11, 2026. In those reports, the most common specific device problems are “Electrical /Electronic Property Problem” (597), “Communication or Transmission Problem” (541) and “Output Problem” (359); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (27), “Burn(s)” (19) and “Perforation” (11). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Olympus WB91051W?
1 report is classified as a death, 135 as injuries and 1,755 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Olympus WB91051W, and is it still sold?
The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model WB91051W). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES874 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
- Vasoview Hemopro VH-3500Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories4 recalls618 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page