LigaSure LF1923Recalls & reported problems

Covidien LP · Maryland Jaw Open Sealer/Divider Coated 23cm · Class 2

All models of LigaSure (32)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien LigaSure Maryland jaw sealer is an electrosurgical instrument used to seal and divide tissue. Many of the 844 reports describe jaws that are hard to open or close, energy output problems and parts detaching, and some describe the device activating unintentionally or failing to seal vessels. The reports include 22 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
831
Deaths reported
0
Injuries reported
22
Recalls naming it
0

Summary

The FDA's MAUDE database contains 831 adverse event reports that list this exact device (DI 10884521705906), received between Apr 26, 2018 and Aug 5, 2026 — classified as 22 injuries, 809 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (287 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 7 reportsJan 2024: 19 reportsFeb 2024: 4 reportsMar 2024: 6 reportsApr 2024: 8 reportsMay 2024: 6 reportsJun 2024: 8 reportsJul 2024: 6 reportsAug 2024: 4 reportsSep 2024: 7 reportsOct 2024: 6 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 7 reportsFeb 2025: 3 reportsMar 2025: 4 reportsApr 2025: 6 reportsMay 2025: 2 reportsJun 2025: 9 reportsJul 2025: 9 reportsAug 2025: 12 reportsSep 2025: 5 reportsOct 2025: 6 reportsNov 2025: 5 reportsDec 2025: 12 reportsJan 2026: 20 reportsFeb 2026: 18 reportsMar 2026: 12 reportsApr 2026: 5 reportsMay 2026: 3 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close287
  • Energy Output Problem271
  • Retraction Problem170
  • Detachment of Device or Device Component99
  • Intermittent Energy Output55
  • Device Alarm System45
  • Protective Measures Problem41
  • Mechanical Problem40

Patient problems

  • No Clinical Signs, Symptoms or Conditions586
  • No Consequences Or Impact To Patient161
  • Insufficient Information42
  • Hemorrhage/Blood Loss/Bleeding38
  • No Information14
  • No Patient Involvement13
  • Unintended Radiation Exposure7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3

Recent adverse event reports

The 25 most recent of 831 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Energy Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50142798 ↗
  • MalfunctionJun 29, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A LAPAROTOMY PROCEDURE, THE DEVICE HAD NO SEAL. THERE WAS NO REGRASP ALERT, WITH NO EVIDENCE OF ENERGY DELIVERY DESPITE END TONES BEING HEARD AFTER APPROXIMATELY 2 GOOD SEALS ON BELOW 7 MM THICKNESS OF TISSUE. THE JAWS WERE CLEANED EACH TIME AFTER USE OR WHEN ESCHAR FORMATION WAS PRESENT. ANOTHER DEVICE WAS REPORTED TO HAVE BEEN USED SUCCESSFULLY WITH THE SAME GENERATOR. PATIENT HAD MINIMAL BLEEDING. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25648665 ↗
  • MalfunctionMay 27, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE THE DEVICE¿S SEALING WAS INADEQUATE OR PARTIAL, WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD. THE SURGICAL TIME WAS EXTENDED BUT ¿NOT VERY LONG¿. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25297049 ↗
  • InjuryMay 13, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A DIXON/LOW ANTERIOR RESECTION (LAR) PROCEDURE WITH A USE OF THE GENERATOR, TWO UNITS OF LIGASURE HANDPIECES DID NOT COAGULATE PROPERLY. PROPER SEALING CYCLE COMPLETED TONES WERE HEARD WITH END TONE PRESENT. IT WAS USED ON TISSUE BUNDLE AND WAS NOT ABLE TO COMPLETE ANY GOOD SEAL. PROCEDURE WAS COMPLETED USING A NON-MEDTRONIC DEVICE. BLEEDING OCCURRED WITH BLOOD LOSS OF LESS THAN 250 CC AND A TRANSFUSION WAS REQUIRED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25158510 ↗
  • InjuryMay 13, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A DIXON/LOW ANTERIOR RESECTION (LAR) PROCEDURE WITH A USE OF THE GENERATOR, TWO UNITS OF LIGASURE HANDPIECES DID NOT COAGULATE PROPERLY. PROPER SEALING CYCLE COMPLETED TONES WERE HEARD WITH END TONE PRESENT. IT WAS USED ON TISSUE BUNDLE AND WAS NOT ABLE TO COMPLETE ANY GOOD SEAL. PROCEDURE WAS COMPLETED USING A NON-MEDTRONIC DEVICE. BLEEDING OCCURRED WITH BLOOD LOSS OF LESS THAN 250 CC AND A TRANSFUSION WAS REQUIRED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25158473 ↗
  • MalfunctionApr 27, 2026Failure to Cut; Difficult to Open or Close; Intermittent Energy Output

    IT WAS REPORTED THAT HALFWAY THROUGH AN OPEN NEPHRECTOMY PROCEDURE, THE DEVICE DID NOT CUT SUFFICIENTLY, DIFFICULT OR IMPOSSIBLE TO CLOSE AND DIFFICULT OR IMPOSSIBLE TO OPEN WHEN APPLIED TO BOWEL FAT BUNDLE. THE DEVICE WAS NOT GET STUCK ON TISSUE AND CLEANED EVERY TIME IT WAS OUT OF THE PORT. THE KNIFE BLADE DID NOT EXTEND AND DID NOT PROTRUDE BEYOND THE EDGE OF THE JAW. ANOTHER LIGASURE DEVICE WAS REPORTED TO HAVE BEEN USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24995345 ↗
  • MalfunctionApr 17, 2026Device Remains Activated; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE¿S PURPLE ACTIVATION BUTTON GOT STUCK AND COULD NOT BE BACK TO THE ORIGINAL PLACE AFTER GETTING STUCK. THE DEVICE WAS STILL ACTIVATING WHEN THE BUTTON GOT STUCK ON THE PRESSED POSITION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24906812 ↗
  • MalfunctionApr 14, 2026Detachment of Device or Device Component

    ACCORDING TO THE REPORTER DURING A RADICAL HYSTERECTOMY PROCEDURE, DURING LYMPH NODE DISSECTION A METAL COMPONENT FROM THE FIRST LIGASURE DEVICE THAT PRODUCES THE HANDLE¿S CLICKING SOUND BROKE OFF AND DETACHED, WHILE E404 OCCURRED CONTINUOUSLY ON THE SECOND LIGASURE DEVICE THAT ALSO OCCURRED FREQUENTLY WITH OTHER TISSUES. ANOTHER LIGASURE DEVICES WERE USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872432 ↗
  • MalfunctionApr 6, 2026Device Alarm System; Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE HAD INSUFFICIENT SEALING/ONLY PARTIALLY AS THERE WAS AN ENERGY OUTPUT AND SEALING COMPLETION SOUND COULD BE CONFIRMED. CLEANED THE AREA AROUND THE JAW EVERY TIME IT GOT DIRTY. ANOTHER LIGASURE WAS APPROPRIATE FOR USE WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793939 ↗
  • MalfunctionApr 1, 2026Difficult to Open or Close; Intermittent Energy Output

    IT WAS REPORTED THAT DURING LIGAMENT DISSECTION ON A TOTAL HYSTERECTOMY, THE JAWS OF THE HANDPIECE DID NOT OPEN EVEN WHEN THE HANDLE WAS RELEASED. IT WAS NOT STUCK ON TISSUE WHEN THE ISSUE HAPPENED. WHEN THEY TRIED OPENING AND CLOSING THE DEVICE OUTSIDE THE SURGICAL FIELD, THE JAW OPENED BUT IT WAS HANDLED BY REPLACING IT WITH AN ANOTHER PRODUCT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751313 ↗
  • MalfunctionMar 31, 2026Device Alarm System; Energy Output Problem; Intermittent Energy Output

    IT WAS REPORTED THAT DURING A PYLORUS-PRESERVING PANCREATODUODENECTOMY(PPPD), WHEN THE DEVICE BUTTON WAS PRESSED, A RED LIGHT APPEARED ON THE DEVICE AND THE DEVICE HAD INADEQUATE OR PARTIAL SEAL DESPITE EVIDENCE OF ENERGY DELIVERY AND END TONES BEING HEARD AFTER SEVERAL SUCCESSFUL SEALS WERE COMPLETED. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24739402 ↗
  • MalfunctionMar 23, 2026Energy Output Problem; Intermittent Energy Output

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, ON THE LIVER TISSUE, THE DEVICE WAS INITIALLY CONNECTED TO THE GENERATOR, BUT AFTER TWO HOURS OF THE PROCEDURE, THE SEAL WAS NOT COMPLETED, EVEN WITH THE MANDIBLE BEING CLEANED FREQUENTLY. SEALING WAS INADEQUATE/PARTIAL (WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD). SWITCHED TO ANOTHER GENERATOR, BUT THE PROBLEM PERSISTED. DUE TO THE PATIENT'S BLEEDING, THE FORCEPS WERE REPLACED. TESTED THE GENERATOR AGAIN, WHICH SHOWED ERROR 306, AND THE FORCEPS STILL DID NOT SEAL PROPERLY. THEY REPLACED IT WITH ANOTHER GENERATOR AND THE SEAL WAS PERFECT AGAIN.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24662231 ↗
  • MalfunctionMar 20, 2026Communication or Transmission Problem; Difficult to Open or Close

    IT WAS REPORTED THAT DURING PULMONARY RESECTION, THE OUTPUT BUTTON BECAME STUCK IN THE DEPRESSED POSITION AFTER BEING PRESSED. THE DEVICE WAS NOT APPLIED TO TISSUE AT THE TIME OF THE INCIDENT. THE KNIFE BLADE WAS NOT EXTENDED AND IT DID NOT PROTRUDE BEYOND THE EDGE OF THE JAWS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649372 ↗
  • MalfunctionMar 20, 2026Energy Output Problem; Intermittent Energy Output

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE HAD NO SEAL, AS THERE WAS NO RE-GRASP ALERT OR ANY OTHER ERROR REPORTED, WITH NO EVIDENCE OF ENERGY DELIVERY AND NO END TONES HEARD. THE JAWS WERE CLEANED THREE TIMES BEFORE THE ISSUE OCCURRED, AND APPROXIMATELY FOUR GOOD SEALS TOOK PLACE PRIOR TO THE ISSUE. ANOTHER DEVICE WAS USED DURING THE PROCEDURE AND WORKED WITHOUT ANY PROBLEMS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24648896 ↗
  • MalfunctionMar 20, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING SURGERY, THE LIGASURE VESSEL SEALING DEVICE WAS DIFFICULT OR IMPOSSIBLE TO OPEN DUE TO A JAW ISSUE. THE DEVICE WAS NOT PLUGGED INTO A GENERATOR WHEN THE EVENT OCCURRED. THE SURGERY WAS COMPLETED WITH A NEW PRODUCT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24648281 ↗
  • MalfunctionMar 18, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING SURGERY WITH THE LIGASURE VESSEL SEALING DEVICE, A SMALL METAL PART LOOSENED FROM THE INSTRUMENT AND FELL OFF, BUT DID NOT FALL INTO THE PATIENT CAVITY AS IT WAS NOTICED BY THE NURSE. ANOTHER LIGASURE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24634483 ↗
  • MalfunctionMar 17, 2026Peeled/Delaminated; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PROCEDURE, THE DEVICE INSULATION CERAMIC COMPONENT HAD FELL OFF. THE PIECE DID NOT FELL INTO THE PATIENT'S CAVITY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24616841 ↗
  • MalfunctionMar 16, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING A PROCEDURE, THE OPENING AND CLOSING OPERATION OF THE HANDPIECE REQUIRED MORE FORCE THAN USUAL. MORE RESISTANCE THAN USUAL WAS FELT. THE HANDPIECE WAS NOT APPLIED TO THE TISSUE WHEN THE ISSUE OCCURRED. THE KNIFE BLADE WAS NOT EXTENDED NOR PROTRUDED BEYOND THE EDGE OF THE JAWS. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24607107 ↗
  • MalfunctionMar 13, 2026Device Alarm System; Unintended Power Up

    IT WAS REPORTED THAT DURING USE THE DEVICE HAD AN INCOMPLETE SEAL CYCLE MESSAGE APPEARED ON THE SCREEN, A SEALING SOUND WAS GENERATED WITHOUT A SEALING COMPLETION SOUND, AND THE DEVICE ACTIVATED WHEN UNINTENDED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24597875 ↗
  • MalfunctionMar 12, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING HEPATECTOMY PROCEDURE, THE DEVICE HAD INADEQUATE OR PARTIAL SEALING WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD. MINIMAL BLEEDING OCCURRED AFTER CUTTING DESPITE EVIDENCE OF ENERGY DELIVERY. THE JAWS WERE CLEANED FREQUENTLY DURING THE PROCEDURE, AND APPROXIMATELY 10-15 GOOD SEALS WERE ACHIEVED BEFORE THE ISSUE OCCURRED WHILE SEALING A VESSEL OF STANDARD TISSUE THICKNESS. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24579015 ↗
  • MalfunctionMar 11, 2026Peeled/Delaminated; Detachment of Device or Device Component; Difficult to Open or Close

    IT WAS REPORTED THAT DURING DISSECTION OF AORTIC LYMPH NODE ON TOTAL HYSTERECTOMY PROCEDURE, THE DEVICE'S JAW COULD NOT BE OPENED OR CLOSED, AND THE TIP AT THE END OF THE JAW WAS PEELED OFF. ANOTHER LIGASURE WAS APPROPRIATE FOR USE WITH THE SAME GENERATOR. NOTHING FELL ON THE PATIENT'S CAVITY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24569454 ↗
  • MalfunctionMar 3, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING LYMPH NODE DISSECTION ON A MODIFIED RADICAL HYSTERECTOMY, A PART OF THE HANDPIECE GOT DETACHED. WHEN THE LEVER WAS PULLED, THE METAL PART INSIDE THE HANDLE AT THE TIME THE FIRST CLICKING SOUND WAS HEARD. IT FELL OFF INTO THE ABDOMINAL CAVITY, BUT IT WAS RETRIEVED AND PLACED INSIDE THE RETURNED PRODUCT. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. X-RAY OR ANY ADDITIONAL INTERVENTION WAS NOT NEEDED TO RETRIEVE THE BROKEN PIECE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24494991 ↗
  • MalfunctionFeb 24, 2026Retraction Problem; Protective Measures Problem

    IT WAS REPORTED THAT, DURING A PROCEDURE, WHEN THE CUTTING TRIGGER WAS PULLED, IT WAS FIXATED AND DID NOT RETURN. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429170 ↗
  • MalfunctionFeb 23, 2026Difficult to Open or Close; Intermittent Energy Output

    IT WAS REPORTED THAT DURING THE RESECTION OF THE PERIPANCREATIC TISSUE ON PANCREATICODUODENECTOMY, THE DEVICE'S JAW DID NOT OPEN EVEN WHEN THE HANDLE WAS RELEASED AFTER ABOUT 3 HOURS OF USE. THE JAW WOULD NOT OPEN UNLESS THE HANDLE WAS PUSHED FORWARD. THE DEVICE WAS NOT ATTACHED TO THE TISSUE. THE DEVICE WAS CLEANED EVERY TIME IT WAS RETURNED TO THE NURSE AND ABOUT ONCE EVERY 15 MINUTES. THE KNIFE BLADE EXTENDED AND DID NOT PROTRUDE BEYOND THE EDGE OF THE JAWS. ANOTHER LIGASURE WAS PROPERLY USED WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416333 ↗
  • MalfunctionFeb 23, 2026Energy Output Problem; Failure to Cut

    IT WAS REPORTED THAT DURING A PROCEDURE WITH A USE OF THE LIGASURE VESSEL SEALING DEVICE, THE DEVICE WAS NOT CUTTING SUFFICIENTLY AND THERE WAS NO SEAL, AS EVIDENCED BY A LACK OF ENERGY DELIVERY AND END TONES BEING HEARD. THE DEVICE WAS PLUGGED INTO A GENERATOR (FT10, SOFTWARE VERSION 5.0), AND ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416179 ↗

FDA clearance and classification

Cleared through Special 510(k) K170869 on Apr 21, 2017 (decision: Substantially Equivalent). Submitted by Covidien as “LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Common questions about the LigaSure LF1923

Has the LigaSure LF1923 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521705906), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LigaSure LF1923?

The FDA's MAUDE database lists 831 adverse event reports for this device, received between Apr 26, 2018 and Aug 5, 2026. In those reports, the most common specific device problems are “Difficult to Open or Close” (287), “Energy Output Problem” (271) and “Retraction Problem” (170); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (38) and “Unintended Radiation Exposure” (7). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LigaSure LF1923?

None of the reports are classified as deaths, 22 as injuries and 809 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LigaSure LF1923 FDA approved?

It was cleared through Special 510(k) K170869 on Apr 21, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the LigaSure LF1923, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand LigaSure, model LF1923). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.