VAPR 225028Recalls & reported problems
Depuy Mitek, LLC · VAPR TRIPOLAR 90 Degrees Suction Electrode 90 Degrees Suction with Integrated Handpiece and Hand Controls · Class 2
All models of VAPR (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Mitek VAPR Tripolar suction electrode appears in 893 adverse event reports, with 19 injuries and no deaths. Many describe fluid flow problems and the electrode not stopping on its own after activation, with some mentioning burns and objects left in patients. Some describe it continuing to activate during surgery until it was replaced. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 874
- Deaths reported
- 0
- Injuries reported
- 18
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 874 adverse event reports that list this exact device (DI 10886705023103), received between Nov 29, 2017 and Jul 27, 2026 — classified as 18 injuries, 856 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (308 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem308
- Device Remains Activated141
- Electrical /Electronic Property Problem82
- Detachment of Device or Device Component70
- Output Problem67
- Break63
- Suction Failure59
- Physical Resistance/Sticking35
Patient problems
- No Clinical Signs, Symptoms or Conditions721
- No Consequences Or Impact To Patient100
- Not Applicable35
- Burn(s)10
- Foreign Body In Patient9
- Unspecified Tissue Injury7
- No Known Impact Or Consequence To Patient3
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 874 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2026Output Problem
It was reported that during a Shoulder Debridement surgery, it was noted that the VAPR device suddenly stopped working and it was not possible to be used. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50006541 ↗MalfunctionJul 27, 2026Device Remains Activated; Physical Resistance/Sticking
It was reported that during a Knee Surgery, it was noted the VAPR device start button did not release after being pressed. The device kept working and did not stop, even with the footswitch. There was no fault reported with the footswitch. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information was provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50005708 ↗MalfunctionJun 24, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE GENERATED SPARKS. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25611054 ↗MalfunctionJun 24, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE GENERATED SPARKS. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25611050 ↗MalfunctionJun 16, 2026Electrical /Electronic Property Problem; Power Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE GENERATED SPARKS AND HAD LOW POWER. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN (3 TOTAL). ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 3 OF 3 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523600 ↗MalfunctionJun 16, 2026Electrical /Electronic Property Problem; Power Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE GENERATED SPARKS AND HAD LOW POWER. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN (3 TOTAL). ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523325 ↗MalfunctionJun 16, 2026Electrical /Electronic Property Problem; Power Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE GENERATED SPARKS AND HAD LOW POWER. CHANGED ANOTHER TWO TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN (3 TOTAL). ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523323 ↗MalfunctionJun 10, 2026Electrical /Electronic Property Problem; Blocked Connection
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, THE VAPR DEVICE WAS OPENED AND CONNECTED, BUT IT WAS NOT RECOGNIZED AND USED BECAUSE IT MAY HAD BEEN SHORT-CIRCUITED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT. THE DEVICE WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25455313 ↗MalfunctionJun 4, 2026Electrical /Electronic Property Problem; Infusion or Flow Problem
IT WAS REPORTED THAT DURING ARTHROSCOPIC SURGERY, THE VAPR DEVICE WAS NOT WORKING CORRECTLY. TISSUE WAS NOT VAPORIZING, IT WAS REMOVING CLUMPS THEN GETTING STUCK. SO UNABLE TO PROCEED USING IT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 20 MINUTE SURGICAL DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25389290 ↗MalfunctionJun 1, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING ROTATOR CUFF REPAIR, THE VAPR DEVICE GENERATED SPARKS. CHANGED ANOTHER TWO VAPR DEVICES TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25346943 ↗MalfunctionJun 1, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING ROTATOR CUFF REPAIR, THE VAPR DEVICE GENERATED SPARKS. CHANGED ANOTHER TWO VAPR DEVICES TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25346937 ↗MalfunctionMay 28, 2026Infusion or Flow Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR SURGERY, THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE BECAME CLOGGED IMMEDIATELY AFTER IT WAS FIRST USED, MAKING IT IMPOSSIBLE TO USE PROPERLY. THE SURGEON MANAGED TO CONTINUE USING THE DEVICE AND COMPLETED THE SURGERY. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25318547 ↗MalfunctionMay 27, 2026Device Remains Activated
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE WOULD STILL RUN WITHOUT PRESSING THE BUTTON IN THE MANUAL CONTROL MODE. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25306956 ↗MalfunctionMay 27, 2026Device Remains Activated
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE WOULD STILL RUN WITHOUT PRESSING THE BUTTON IN THE MANUAL CONTROL MODE. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25306890 ↗MalfunctionMay 27, 2026Device Remains Activated
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE WOULD STILL RUN WITHOUT PRESSING THE BUTTON IN THE MANUAL CONTROL MODE. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER VAPR TRIPOLAR 90 SUCTION ELECT DEVICE AND THE SAME PROBLEM HAPPENED AGAIN. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 2 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25306885 ↗MalfunctionMay 27, 2026Infusion or Flow Problem; Output Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE SUCTION OF THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE STOPPED WORKING IMMEDIATELY AFTER USE, MAKING IT IMPOSSIBLE TO USE NORMALLY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT, OR SURGICAL DELAY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25303200 ↗MalfunctionMay 26, 2026Infusion or Flow Problem; Output Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE SUCTION FUNCTION ON THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE STOPPED WORKING IMMEDIATELY AFTER USE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25277490 ↗MalfunctionMay 14, 2026Electrical /Electronic Property Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE VAPR TRIPOLAR 90 SUCTION ELECT DEVICE FELL APART. IT WAS INITIALLY REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE DEVICE OUTPUT SHORTED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25171857 ↗MalfunctionMay 8, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE METAL PLATE ON THE HEAD OF THE WAND DISENGAGED IN THE PATIENT. THE PROCEDURE STOPPED AND METAL TIP REMOVED FROM SHOULDER. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 15 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25129098 ↗MalfunctionMay 8, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE THAT THE SURGEON INSERTED THE ELECTRODE INTO THE JOINT THROUGH THE ANTERIOR PORTAL DURING AN AND THE TIP WAS ALREADY HALF BROKEN OFF. THERE WERE NO FRAGMENTS LEFT INSIDE THE BODY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25128823 ↗MalfunctionApr 29, 2026Electrical /Electronic Property Problem; Overheating of Device; Infusion or Flow Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE THAT THE ELECTRODE WAS ONLY GENERATING HEAT AND BUBBLES BUT GETTING NO SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25035477 ↗MalfunctionApr 28, 2026Infusion or Flow Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE SUCTION FUNCTION STOPPED WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25022327 ↗MalfunctionApr 24, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE FOR SHOULDER REPAIR IT WAS OBSERVED THAT THE METALLIC TIP OF THE VAPR TRIPOLAR 90 SUCTION ELECT ELECTRODE DEVICE SLOWLY DISCONNECTED FROM THE HEAD OF THE TIP. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO CONTINUE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT, OR SURGICAL DELAY REPORTED. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24982233 ↗MalfunctionApr 23, 2026Output Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THAT THE VAPR TRIPOLAR 90 SUCTION ELECTRODE DEVICE SUDDENLY STOP WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24971969 ↗MalfunctionApr 22, 2026Infusion or Flow Problem; Output Problem
IT WAS REPORTED THAT DURING UNKNOWN SURGERY WITH A VAPR TRIPOLAR 90 SUCTION ELECTRODE, THE PHYSICIAN NOTICED THAT THE WANDS SUCTION FAILED TO WORK AFTER ABOUT 30 SECONDS OF USE. THE TIP OF THE PROBE WAS CLEARED OF DEBRIS, AND THE SUCTION TUBING WAS TURNED TO THE MAXIMUM SETTING, HOWEVER THE WAND WAS STILL NOT PROVIDING ANY SUCTION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT OR USER OR SIGNIFICANT DELAYS. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THE DEVICE IS NOT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24953500 ↗
FDA clearance and classification
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the VAPR 225028
Has the VAPR 225028 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705023103), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VAPR 225028?
The FDA's MAUDE database lists 874 adverse event reports for this device, received between Nov 29, 2017 and Jul 27, 2026. In those reports, the most common specific device problems are “Infusion or Flow Problem” (308), “Device Remains Activated” (141) and “Electrical /Electronic Property Problem” (82); the most common patient problems are “Not Applicable” (35), “Burn(s)” (10) and “Foreign Body In Patient” (9). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VAPR 225028?
None of the reports are classified as deaths, 18 as injuries and 856 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VAPR 225028, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand VAPR, model 225028). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,891 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
- Vasoview Hemopro VH-3500Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories4 recalls618 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page