LigaSure LF1937Recalls & reported problems

Covidien LP · Maryland Jaw Laparoscopic Sealer/Divider Coated · Class 2

All models of LigaSure (32)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien LigaSure Maryland jaw sealer is a laparoscopic surgical instrument that seals and divides tissue with electrical energy. Its 2,067 adverse event reports mostly describe energy output problems, jaws that are difficult to open or close, and tissue retraction problems, and some mention bleeding. The reports include 3 deaths and 87 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,038
Deaths reported
3
Injuries reported
86
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,038 adverse event reports that list this exact device (DI 10884521705913), received between Jan 9, 2018 and Sep 24, 2026 — classified as 3 deaths, 86 injuries, 1,949 malfunctions.

The most frequently reported specific device problem is “Energy Output Problem” (862 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 36 reportsJan 2024: 47 reportsFeb 2024: 35 reportsMar 2024: 30 reportsApr 2024: 22 reportsMay 2024: 16 reportsJun 2024: 12 reportsJul 2024: 14 reportsAug 2024: 21 reportsSep 2024: 23 reportsOct 2024: 19 reportsNov 2024: 25 reportsDec 2024: 19 reportsJan 2025: 22 reportsFeb 2025: 8 reportsMar 2025: 16 reportsApr 2025: 14 reportsMay 2025: 12 reportsJun 2025: 25 reportsJul 2025: 19 reportsAug 2025: 18 reportsSep 2025: 7 reportsOct 2025: 16 reportsNov 2025: 12 reportsDec 2025: 16 reportsJan 2026: 11 reportsFeb 2026: 24 reportsMar 2026: 20 reportsApr 2026: 8 reportsMay 2026: 3 reportsJun 2026: 5 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202647
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Energy Output Problem862
  • Difficult to Open or Close645
  • Retraction Problem315
  • Device Alarm System143
  • Intermittent Energy Output121
  • Detachment of Device or Device Component120
  • Mechanical Problem100
  • Failure to Cut73

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,307
  • No Consequences Or Impact To Patient403
  • Hemorrhage/Blood Loss/Bleeding165
  • Insufficient Information161
  • No Information33
  • No Patient Involvement25
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available16
  • Blood Loss13

Recent adverse event reports

The 25 most recent of 2,038 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Device Alarm System; Energy Output Problem

    It was reported that during a colon per diverticulitis procedure with a use of the device to seal peritoneal tissue, the handpiece had inadequate seal despite evidence of energy delivery and audible end tones. It also alarmed with an error indicating interruption of use before completion of the vessel seal. No successful seals were achieved before the issue occurred. The device was replaced after the first activations. There was no patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885157 ↗
  • MalfunctionAug 6, 2026Difficult to Open or Close; Intermittent Energy Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50150776 ↗
  • MalfunctionAug 5, 2026Failure to Deliver Energy; Difficult to Open or Close; Intermittent Energy Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143018 ↗
  • MalfunctionJul 28, 2026Failure to Seal

    Ligasure would not seal.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50053356 ↗
  • MalfunctionJun 30, 2026Self-Activation or Keying

    IT WAS REPORTED THAT DURING A PROCEDURE, A DEVICE ACTIVATED WITHOUT THE ACTIVATION BUTTON BEING PUSHED. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25666055 ↗
  • MalfunctionJun 9, 2026Device Alarm System; Energy Output Problem

    IT WAS REPORTED THAT DURING USE ON A LAPAROSCOPIC ROUX-EN-Y GASTRIC BYPASS, THE DEVICE HAD A REGRASP ALERT WITH NO SEAL AND NO EVIDENCE OF ENERGY DELIVERY, WHILE END TONES WERE HEARD. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. NO JAWS CLEANING WAS PERFORMED AND THERE WERE ZERO GOOD SEALS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25449837 ↗
  • MalfunctionJun 4, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING USE ON APPENDECTOMY, FIRST SEALING WAS INADEQUATE OR PARTIAL DESPITE EVIDENCE OF ENERGY DELIVERY AND END TONES BEING HEARD. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR, HOWEVER THE SECOND DEVICE HAD THE SAME ISSUE OF INADEQUATE SEALING. THE THIRD DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25392861 ↗
  • MalfunctionJun 4, 2026Energy Output Problem; Intermittent Energy Output

    IT WAS REPORTED THAT DURING USE ON APPENDECTOMY, FIRST SEALING WAS INADEQUATE OR PARTIAL DESPITE EVIDENCE OF ENERGY DELIVERY AND END TONES BEING HEARD. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR, HOWEVER THE SECOND DEVICE HAD THE SAME ISSUE OF INADEQUATE SEALING. THE THIRD DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25392612 ↗
  • MalfunctionJun 2, 2026Difficult to Open or Close

    IT WAS REPORTED THAT ONCE THE SURGERY BEGAN, THE DEVICE WAS USED WHICH IMMEDIATELY STARTED HAVING DIFFICULT OR IMPOSSIBLE TO OPEN. THE DEVICE¿S JAWS WERE FORCED OPEN TO REMOVE THE DEVICE. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25360641 ↗
  • MalfunctionMay 27, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING USE, THE DEVICE'S JAW WOULD NOT OPEN AND CLOSE. THE DEVICE IMMEDIATELY REMOVED FROM THE SURGICAL FIELD AND REPLACED IT WITH A NEW ONE WHICH WORKED FINE. THE DEVICE HAD NO PATIENT CONTACT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25301042 ↗
  • MalfunctionMay 26, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, THE PHYSICIAN PRESSED TO SEAL, AND THE CLAMP REMAINED CLOSED ON THE VESSEL FOR MORE THAN 1 MINUTE WHILE WAITING FOR THE GENERATOR TO TAKE THE READING, WHEREAS THE PROCESS WAS EXPECTED TO TAKE APPROXIMATELY 4 SECONDS. THE DEVICE HAD INADEQUATE/PARTIAL SEALING WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25279844 ↗
  • MalfunctionMay 18, 2026Retraction Problem; Difficult to Open or Close; Intermittent Energy Output

    IT WAS REPORTED THAT APPROXIMATELY 60 MINUTES INTO THE SURGERY THE DEVICE FUNCTIONED CORRECTLY AT THE START, THE DEVICE PRESENTED RE SISTANCE WHEN CLOSING AND OPENING THE JAWS. THE CLAMP NO LONGER OPENED PROPERLY, AND THE BLADE STUCK IN THE MIDDLE OF THE CLAMPS (DID NOT RETRACT) WHEN SEALING SECONDARY VESSELS IN UTERO AFTER APPROXIMATELY 5 TO 6 SEALS. THE KNIFE BLADE DID NOT PROTRUDE BEYOND THE EDGE OF THE JAWS AND WAS NOT EXPOSED. CLEANING WAS ROUTINE AFTER EACH USE AND EACH TIME, IT WAS REMOVED FROM THE CAVITY IT WAS CLEANED WITH WET GAUZE. A NEW DEVICE WAS SUPPLIED USING THE SAME GENERATOR, AND THE SURGERY PROCEEDED WITHOUT ANY ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25200513 ↗
  • MalfunctionApr 20, 2026Retraction Problem

    IT WAS REPORTED THAT DURING LAPAROSCOPIC HYSTERECTOMY, THE CUTTING BUTTON OF THE HANDPIECE CANNOT RETURN BACK TO ORIGINAL POSITION. THERE WAS A 10-MINUTE DELAY ON THE PROCEDURE TO REPLACE THE HANDPIECE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24933644 ↗
  • MalfunctionApr 14, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING A LUMBAR-AORTIC LYMPH NODE DISSECTION, THE DEVICE WAS DIFFICULT/IMPOSSIBLE TO OPEN. THE DEVICE GOT STUCK ON TISSUE ON LYMPH NODE AND REMOVED BY FORCING OF THE JAWS. THE KNIFE BLADE WAS NOT EXTENDED AND DID NOT PROTRUDE BEYOND THE EDGE OF THE JAWS. ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873351 ↗
  • MalfunctionApr 6, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE WITH A USE OF THE LIGASURE DEVICE, THERE WAS AN ENERGY ACTIVATION OR SEALING ISSUE DESCRIBED AS SEALING BEING INADEQUATE OR PARTIAL, WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24791840 ↗
  • MalfunctionApr 3, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING PROCEDURE THE DEVICE HAD NO SEAL, WITH NO EVIDENCE OF ENERGY DELIVERY AND END TONES BEING HEARD. THE DEVICE WAS PLUGGED INTO A GENERATOR AT THE TIME OF THE EVENT, AND ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO BLEEDING REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24776436 ↗
  • MalfunctionApr 1, 2026Difficult to Open or Close

    IT WAS REPORTED THAT IMMEDIATELY AFTER THE START OF THE PROCEDURE, THE JAWS OF THE HANDPIECE WAS DIFFICULT TO OPEN. IT WAS NOT ATTACHED TO THE TISSUE WHEN THE ISSUE OCCURRED. ANOTHER HANDPIECE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24755642 ↗
  • MalfunctionApr 1, 2026Self-Activation or Keying

    IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE ACTIVATED WITHOUT THE ACTIVATION BUTTON BEING PUSHED. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24755276 ↗
  • MalfunctionApr 1, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE USING THE DEVICE, THERE WAS AN ENERGY ACTIVATION ISSUE AND SEALING WAS INADEQUATE OR PARTIAL, WITH EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD. ANOTHER DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753649 ↗
  • MalfunctionApr 1, 2026Unintended Power Up; Difficult or Delayed Activation

    IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE DID NOT WORK WELL WHEN CONNECTED THE THE UNIT. ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24751352 ↗
  • MalfunctionMar 31, 2026Device Alarm System; Energy Output Problem

    IT WAS REPORTED THAT DURING USE, THE DEVICE HAD NO SEAL (NO EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD). THERE WAS ALSO A RE-GRASP ALERT. NO GOOD SEALS WERE COMPLETED. ANOTHER LIGASURE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24740463 ↗
  • MalfunctionMar 31, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE WITH A USE OF THE LIGASURE VESSEL SEALING DEVICE, THERE WAS AN ENERGY ACTIVATION ISSUE AND A SEALING ISSUE, SPECIFICALLY NO SEAL WAS ACHIEVED AS THERE WAS NO EVIDENCE OF ENERGY DELIVERY AND END TONES WERE HEARD. THE DEVICE WAS PLUGGED INTO A GENERATOR AT THE TIME OF THE EVENT, AND ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24739752 ↗
  • MalfunctionMar 31, 2026Device Alarm System; Energy Output Problem

    IT WAS REPORTED THAT DURING USE OF THE LIGASURE VESSEL SEALING DEVICE, THE DEVICE HAD NO SEAL (NO EVIDENCE OF ENERGY DELIVERY AND END TONES HEARD). THE DEVICE WAS PLUGGED INTO A GENERATOR AT THE TIME OF THE EVENT AND ANOTHER LIGASURE DEVICE WAS USED SUCCESSFULLY WITH THE SAME GENERATOR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24739633 ↗
  • MalfunctionMar 30, 2026Difficult to Open or Close; Intermittent Energy Output

    IT WAS REPORTED THAT IN THE MIDDLE OF A SURGERY USING THE DEVICE, THE SCISSORS JAMMED, PREVENTING THE BLADES FROM OPENING AND KEEPING THEM CLOSED, WHICH MADE THE DEVICE IMPOSSIBLE TO USE. THE BLADES JAMMED WHEN THE SCISSORS WERE ACTIVATED, AND THE DEVICE WAS DESCRIBED AS DIFFICULT OR IMPOSSIBLE TO OPEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24712890 ↗
  • MalfunctionMar 25, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING A PROCEDURE USING THE DEVICE, WHEN ENERGY ACTIVATION WAS ATTEMPTED, THE DEVICE PRODUCED A COAGULATION TONE BUT DID NOT DELIVER ENERGY, AND THERE WAS NO EVIDENCE OF ENERGY DELIVERY OR SEAL FORMATION DESPITE END TONES BEING HEARD. THE DEVICE WAS PLUGGED INTO A GENERATOR WITH SOFTWARE VERSION 4. THE JAWS WERE CLEAN AND THE ISSUE WAS PRESENT FROM THE START OF USE, WITH NO SEALS ACHIEVED AND NO TISSUE SEALED. THE TISSUE INTENDED FOR SEALING WAS APPROXIMATELY 5MM THICK. NO BLEEDING OCCURRED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24686777 ↗

FDA clearance and classification

Cleared through Special 510(k) K170869 on Apr 21, 2017 (decision: Substantially Equivalent). Submitted by Covidien as “LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coated”. The FDA review took 29 days. FDA record ↗

The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Common questions about the LigaSure LF1937

Has the LigaSure LF1937 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521705913), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LigaSure LF1937?

The FDA's MAUDE database lists 2,038 adverse event reports for this device, received between Jan 9, 2018 and Sep 24, 2026. In those reports, the most common specific device problems are “Energy Output Problem” (862), “Difficult to Open or Close” (645) and “Retraction Problem” (315); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (165) and “Blood Loss” (13). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LigaSure LF1937?

3 reports are classified as deaths, 86 as injuries and 1,949 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LigaSure LF1937 FDA approved?

It was cleared through Special 510(k) K170869 on Apr 21, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the LigaSure LF1937, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand LigaSure, model LF1937). Its record shows it as in commercial distribution.