Skip to content

Olympus WA4KL100Recalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope "ULTRA", 10 mm, 0° · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus rigid video laparoscope is a surgical telescope. The 39 reports mostly describe the scope breaking or cracking — tips, light posts and internal lenses coming loose, detaching or going missing — along with a few poor image and burned tip reports. One injury was reported and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
39
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 04042761082846), received between Jun 26, 2019 and May 4, 2026 — classified as 1 injury, 38 malfunctions.

The most frequently reported specific device problem is “Break” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break21
  • Detachment of Device or Device Component7
  • Crack3
  • Component Missing2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Component1
  • Dent in Material1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • Burn, Thermal1

Recent adverse event reports

The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 4, 2026Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073669 ↗
  • MalfunctionMar 23, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED DETACHMENT OF THE INTERNAL LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24664585 ↗
  • MalfunctionFeb 17, 2026Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24367705 ↗
  • MalfunctionFeb 16, 2026Loose or Intermittent Connection

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE HAD A LOOSE LENS MENISCUS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24345370 ↗
  • MalfunctionFeb 6, 2026Overheating of Device

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A BURNED TIP DURING A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A COMPETITOR¿S DEVICE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24280542 ↗
  • MalfunctionFeb 6, 2026Break

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A BROKEN LIGHT POST. THE ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24277896 ↗
  • MalfunctionJan 29, 2026Break

    IT WAS REPORTED THAT THE RIGID VIDEO LAPAROSCOPE HAD A DAMAGED PART CONNECTED TO THE CAMERA. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24209094 ↗
  • MalfunctionDec 23, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID VIDEO LAPAROSCOPE EXHIBITED LENS INSIDE THE OPTICAL SYSTEM IS BROKEN. / DISTAL END LENS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877742 ↗
  • MalfunctionDec 17, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23830063 ↗
  • MalfunctionDec 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23806453 ↗
  • MalfunctionNov 11, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23523916 ↗
  • MalfunctionOct 16, 2025Break; Break

    IT WAS REPORTED THE SUBJECT DEVICE HAD BROKEN LIGHT EMITTING SURFACE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23316776 ↗
  • MalfunctionJun 25, 2025Crack; Crack

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BLACK SPLIT HOOD. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22331856 ↗
  • MalfunctionApr 29, 2025Crack; Crack

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21927810 ↗
  • MalfunctionApr 18, 2025Break; Break

    IT WAS REPORTED THAT THE RIGID SCOPE HAD BROKEN LIGHT POST. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21868874 ↗
  • MalfunctionFeb 19, 2025Melted; Melted

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21411945 ↗
  • MalfunctionFeb 3, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21293729 ↗
  • MalfunctionOct 4, 2024Break; Break

    IT WAS REPORTED THE TELESCOPE HAD A BROKEN LENS. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20375545 ↗
  • MalfunctionSep 23, 2024Break; Break

    IT WAS REPORTED, THE TELESCOPE "ULTRA", 10 MM, 0° EXHIBITED A BROKEN COMPONENT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20277704 ↗
  • MalfunctionJul 5, 2024Break; Break

    IT WAS REPORTED THE RIGID VIDEO LAPAROSCOPE WAS BROKEN. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19678825 ↗
  • MalfunctionJul 3, 2024Break; Break

    IT WAS REPORTED, THE TELESCOPE WAS MISSING AN ADAPTER PIECE FROM THE LIGHT ATTACHMENT. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19667758 ↗
  • MalfunctionApr 25, 2024Break; Break

    IT WAS REPORTED, THE TELESCOPE'S LIGHT GUIDE CONNECTER WAS DAMAGED. THE ISSUE OCCURRED, DURING CLEAN UP AFTER A THERAPEUTIC SURGERY. THERE WERE NO REPORTS OF PATIENT HARM, NO DELAY. AND THE PROCEDURE WAS ABLE TO BE COMPLETED SUCCESSFULLY. AS THE MALFUNCTION OCCURRED, AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19181766 ↗
  • MalfunctionFeb 25, 2024Break; Break

    IT WAS REPORTED THAT THE TELESCOPE "ULTRA", 10 MM, 0° HAD A BROKEN PIECE THAT YOU LOOK THROUGH. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18774837 ↗
  • MalfunctionDec 18, 2023Defective Component; Defective Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE HAD A BAD FIBER OPTIC LIGHT CORD. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18345079 ↗
  • MalfunctionDec 13, 2023Break; Break

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE TELESCOPE "ULTRA", 10 MM, 0° HAD NO BLURRY IMAGE, ONLY A FEW FIBERS WERE BROKEN. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18317197 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page