Olympus WA4KL500Recalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope "ULTRA", 5.4 mm, 0° · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus laparoscope has 131 reports, all malfunction reports, many describing a cracked, chipped or broken lens or optical parts. The reports mention no deaths or injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
131
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 131 adverse event reports that list this exact device (DI 04042761082877), received between Dec 2, 2022 and Jun 20, 2026 — classified as 131 malfunctions.

The most frequently reported specific device problem is “Break” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 4 reportsJan 2024: 3 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 4 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 8 reportsAug 2025: 6 reportsSep 2025: 5 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 10 reportsMar 2026: 11 reportsApr 2026: 5 reportsMay 2026: 5 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break69
  • Detachment of Device or Device Component18
  • Crack17
  • Material Twisted/Bent7
  • Material Fragmentation6
  • Loose or Intermittent Connection5
  • Component Missing3
  • Dent in Material2

Patient problems

  • No Clinical Signs, Symptoms or Conditions131

Recent adverse event reports

The 25 most recent of 131 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE RIGID OPTICAL LAPAROSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE EVALUATION OF THE DEVICE, CHIPPED LENS OPTICAL SYSTEM. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE COMPLAINT WAS TRACED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25570290 ↗
  • MalfunctionJun 2, 2026Dent in Material

    THE CUSTOMER RETURNED THE RIGID OPTICAL LAPAROSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE TECHNICIAN IDENTIFIED THE DEVICE HAD CHIPPED OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365406 ↗
  • MalfunctionMay 20, 2026Break

    IT WAS REPORTED THAT THE RIGID OPTICAL LAPAROSCOPE HAD A CHIP IN THE LENS. THE EVENT OCCURRED DURING INSPECTION FOR USE.THERE WERE NO REPORTS OF PATIENT HARM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25229989 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT THE RIGID OPTICAL LAPAROSCOPE HAD COMPLETELY SHATTERED LENS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162686 ↗
  • MalfunctionMay 13, 2026Break

    IT WAS REPORTED THAT THE RIGID OPTICAL LAPAROSCOPE HAD SHATTERED SCOPE APPEARED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162664 ↗
  • MalfunctionMay 11, 2026Break

    IT WAS REPORTED THAT THE RIGID OPTICAL LAPAROSCOPE HAD BUSTED FIBERS. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136890 ↗
  • MalfunctionMay 1, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED AND BROKEN LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056939 ↗
  • MalfunctionApr 24, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENSES IN THE OPTICS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24978003 ↗
  • MalfunctionApr 22, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENSES IN THE OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956582 ↗
  • MalfunctionApr 20, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENSES IN THE OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24932935 ↗
  • MalfunctionApr 15, 2026Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN THE OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24883589 ↗
  • MalfunctionApr 1, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD BROKEN OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753963 ↗
  • MalfunctionMar 30, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED OBJECT WINDOW CHIPPED MENISCUS LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24734311 ↗
  • MalfunctionMar 30, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24730124 ↗
  • MalfunctionMar 30, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE PRESENTED CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24725461 ↗
  • MalfunctionMar 30, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD A CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715233 ↗
  • MalfunctionMar 30, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24712434 ↗
  • MalfunctionMar 26, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN OPTICAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24701331 ↗
  • MalfunctionMar 26, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24697982 ↗
  • MalfunctionMar 19, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD MULTIPLE CHIPS ON OPTICAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24637481 ↗
  • MalfunctionMar 17, 2026Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24618619 ↗
  • MalfunctionMar 9, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551345 ↗
  • MalfunctionMar 6, 2026Break

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE RIGID OPTICAL LAPAROSCOPE EXHIBITED BROKEN COVER GLASS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24535203 ↗
  • MalfunctionFeb 23, 2026Loose or Intermittent Connection

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD LOOSE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24421527 ↗
  • MalfunctionFeb 23, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD MENISCUS LOOSEN APART. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24417556 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →