Skip to content

Olympus WA4KL530Recalls & reported problems

OLYMPUS Winter & Ibe GmbH · Telescope "ULTRA", 5.4 mm, 30° · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus ULTRA gynecologic telescope is a rigid laparoscope used in keyhole surgery. Its 150 adverse event reports include 4 injuries and no deaths, and many describe broken or chipped lenses, parts coming loose, and cracks, found during device evaluation with no patient involvement. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
150
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 150 adverse event reports that list this exact device (DI 04042761082884), received between Jul 9, 2021 and May 19, 2026 — classified as 4 injuries, 146 malfunctions.

The most frequently reported specific device problem is “Break” (83 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 0 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 7 reportsApr 2025: 3 reportsMay 2025: 6 reportsJun 2025: 4 reportsJul 2025: 5 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 7 reportsFeb 2026: 18 reportsMar 2026: 9 reportsApr 2026: 9 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break83
  • Detachment of Device or Device Component18
  • Crack16
  • Poor Quality Image5
  • Scratched Material5
  • Adverse Event Without Identified Device or Use Problem4
  • Material Twisted/Bent4
  • Loose or Intermittent Connection3

Patient problems

  • No Clinical Signs, Symptoms or Conditions146
  • Burn(s)2
  • Hemorrhage/Blood Loss/Bleeding2
  • Unspecified Infection2

Recent adverse event reports

The 25 most recent of 150 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED INTERNAL LENS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25217671 ↗
  • MalfunctionMay 12, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25152739 ↗
  • MalfunctionMay 6, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095914 ↗
  • MalfunctionApr 27, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD BROKEN OFF LENS MENISCUS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24999933 ↗
  • MalfunctionApr 24, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE PRESENTED DETACHED LENSES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24988678 ↗
  • MalfunctionApr 24, 2026Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24978148 ↗
  • MalfunctionApr 16, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED OPTICAL LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24905275 ↗
  • MalfunctionApr 15, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPS ON THE LENS IN THE OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885605 ↗
  • MalfunctionApr 14, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872250 ↗
  • MalfunctionApr 10, 2026Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845245 ↗
  • MalfunctionApr 9, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD A CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24829974 ↗
  • MalfunctionApr 8, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE PRESENTED CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24818534 ↗
  • MalfunctionMar 31, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE PRESENTED CHIPPED INTERNAL BARREL ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742362 ↗
  • MalfunctionMar 26, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN OPTICAL SYSTEM THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24701289 ↗
  • MalfunctionMar 26, 2026Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24697790 ↗
  • MalfunctionMar 25, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS IN OPTICAL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24684960 ↗
  • MalfunctionMar 24, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE RIGID OPTICAL LAPAROSCOPE HAD A MISSING LG LENS. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24682031 ↗
  • MalfunctionMar 17, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24614974 ↗
  • MalfunctionMar 9, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED CHIPPED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551508 ↗
  • MalfunctionMar 5, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED A CHIP ON THE DISTAL EDGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24523985 ↗
  • MalfunctionMar 2, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE EXHIBITED A CHIPPED OBJECTIVE COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24491340 ↗
  • MalfunctionFeb 26, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD DAMAGED EYEPIECE AND BROKEN BARREL ASSEMBLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455482 ↗
  • MalfunctionFeb 23, 2026Unstable

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD LOOSE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416344 ↗
  • MalfunctionFeb 23, 2026Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413617 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS OBSERVED DURING DEVICE EVALUATION, THE RIGID OPTICAL LAPAROSCOPE HAD BROKEN EYE PIECE FUNNEL CUP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24404841 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page