Skip to content

n.a. 27040DBRecalls & reported problems

Karl Storz GmbH & Co. KG · Working Element, bipolar · Class 2

Adverse event reports
14
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04048551075500), received between Sep 13, 2023 and May 8, 2026 — classified as 8 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device3
  • Device Handling Problem2
  • Display or Visual Feedback Problem2
  • Energy Output Problem2
  • Insufficient Device Problem Information2
  • Misconnection2
  • Smoking2
  • Arcing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Burn(s)2
  • Hemorrhage/Blood Loss/Bleeding2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT TWO INSTRUMENT HANDLES, REF. 27040DB: THE ONE USED WITH SNRK22: DURING USE, THE RESECTION LOOP MOVES OUT OF THE OPTICAL FIELD OF VIEW, EVEN THOUGH THE ASSEMBLY IS CORRECT. THE ONE USED WITH SN (B)(6): THE BIPOLAR RESECTION CABLE BURNED AT THE POINT OF INSERTION INTO THE HANDLE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25129087 ↗
  • MalfunctionMay 8, 2026Display or Visual Feedback Problem

    IT WAS REPORTED THAT TWO INSTRUMENT HANDLES, REF. (B)(4): THE ONE USED WITH SNRK22: DURING USE, THE RESECTION LOOP MOVES OUT OF THE OPTICAL FIELD OF VIEW, EVEN THOUGH THE ASSEMBLY IS CORRECT. THE ONE USED WITH SN OM03: THE BIPOLAR RESECTION CABLE BURNED AT THE POINT OF INSERTION INTO THE HANDLE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25127017 ↗
  • MalfunctionMar 13, 2026Arcing

    IT WAS REPORTED THAT "DURING SURGERY SPARKS OF FIRE CAME FROM THE WORKING ELEMENT AND THE ERBE CABLE 20196-118". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24596661 ↗
  • MalfunctionDec 30, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT "MALFUNCTION DETECTED IN THE ANESTHESIA DEVICE'S HEART RATE MONITOR. THE CABLE AND INSULATION ON THE WORKING ELEMENT ARE BURNT OUT, POSSIBLY DUE TO A SHORT CIRCUIT". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23928248 ↗
  • InjuryAug 25, 2025Device Handling Problem; Device Handling Problem

    IT WAS REPORTED THAT THE ELECTRODE WAS NOT FIXED TO THE WORKING ELEMENT AND CONTINUED TO SEPARATE. AS A RESULT OF INSPECTING BOTH PRODUCTS, IT WAS DETERMINED THAT THE ELECTRODE WAS DEFECTIVE. THE SYMPTOM OF THE DEFECT WAS THAT THERE SHOULD HAVE BEEN A FIXING GROOVE, BUT THE GROOVE WAS NOT PROPERLY CREATED. THERE WAS NO DIRECT IMPACT ON THE PATIENT, BUT SINCE THE SURGERY WAS EXTENDED BY ABOUT 40 MINUTES THE EVENT IS DEEMED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22884801 ↗
  • InjuryMay 25, 2025Misconnection; Device Handling Problem; Misconnection; Device Handling Problem

    IT WAS REPORTED THAT THE ELECTRODE WAS NOT FIXED TO THE WORKING ELEMENT AND CONTINUED TO SEPARATE. AS A RESULT OF INSPECTING BOTH PRODUCTS, IT WAS DETERMINED THAT THE ELECTRODE WAS DEFECTIVE. THE SYMPTOM OF THE DEFECT WAS THAT THERE SHOULD HAVE BEEN A FIXING GROOVE, BUT THE GROOVE WAS NOT PROPERLY CREATED. THERE WAS NO DIRECT IMPACT ON THE PATIENT, BUT SINCE THE SURGERY WAS EXTENDED BY ABOUT 40 MINUTES THE EVENT IS DEEMED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22082512 ↗
  • InjuryMay 2, 2025Misconnection; Device Dislodged or Dislocated; Misconnection; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE ELECTRODE WAS NOT FIXED TO THE WORKING ELEMENT AND CONTINUED TO SEPARATE. AS A RESULT OF INSPECTING BOTH PRODUCTS, IT WAS DETERMINED THAT THE ELECTRODE WAS DEFECTIVE. THE SYMPTOM OF THE DEFECT WAS THAT THERE SHOULD HAVE BEEN A FIXING GROOVE, BUT THE GROOVE WAS NOT PROPERLY CREATED. THERE WAS NO DIRECT IMPACT ON THE PATIENT, BUT SINCE THE SURGERY WAS EXTENDED BY ABOUT 40 MINUTES THE EVENT IS DEEMED REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21952240 ↗
  • InjuryAug 15, 2024Smoking; Temperature Problem; Smoking; Temperature Problem

    IT WAS REPORTED THAT A PROBLEM REGARDING UROLOGY BIPOLAR WORKING ELEMENT 27040DB BURNING AND SMOKING WHILE IN USE AFTER THE CASE THE SURGEON HAS SAID HE HAS ALSO GOT SORE FINGER, IT WAS REPORTED THERE WAS NO HARM TO THE PATIENT. HOWEVER, DUE TO THE MENTIONED SORE FINGER OF THE SURGEON, THIS CASE IS DEEMED REPORTABLE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 19995219 ↗
  • InjuryJul 25, 2024Energy Output Problem; Overheating of Device; Energy Output Problem; Overheating of Device

    ACCORDING TO THE INFORMATION RECEIVED, DURING A BIPOLAR PHOTOVAPORIZATION OF THE PROSTATE: BLACK SMOKE AND FLAME ON THE CONNECTION OF THE BIPOLAR CABLE AND THE RESECTOSCOPE OCCURRED. NO NEGATIVE EFFECTS ON THE PATIENT WERE REPORTED. BUT THE SURGEON SUFFERED BURNS.

    Patient problems: Burn(s)

    MAUDE report 19829376 ↗
  • InjuryJul 25, 2024Overheating of Device; Overheating of Device

    ACCORDING TO THE INFORMATION RECEIVED, DURING A BIPOLAR PHOTOVAPORIZATION OF THE PROSTATE: BLACK SMOKE AND FLAME ON THE CONNECTION OF THE BIPOLAR CABLE AND THE RESECTOSCOPE OCCURRED. NO NEGATIVE EFFECTS ON THE PATIENT WERE REPORTED. BUT THE SURGEON SUFFERED BURNS.

    Patient problems: Burn(s)

    MAUDE report 19829323 ↗
  • MalfunctionMay 23, 2024Sparking; Flare or Flash; Sparking; Flare or Flash

    IT WAS REPORTED THAT THE WORKING ELEMENT AND LEAD SPARKED DURING THE CASE. IT HAS BEEN CONFIRMED THAT THE ELECTRODE WAS INSIDE THE PATIENT WHEN THE INCIDENT OCCURRED, THEY STOPPED THE PROCEDURE STRAIGHTWAY. NO HARM TO THE PATIENT OR THE SURGEON WAS NOTED. ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE RIGHT AFTER THE INCIDENT. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19384671 ↗
  • MalfunctionApr 11, 2024Fire; Smoking; Fire; Smoking

    IT WAS REPORTED THAT SMOKE HAD COME OUT DURING THE INTERVENTION AND THE DEVICE GOT FIRE. THERE WAS A SURGICAL PROLONGATION "DELAY" BECAUSE THEY HAD TO CHANGE THE RESECTION AREA PER OPERATION." PROCEDURE WAS COMPLETED SUCCESSFULLY, NO MEDICAL/ SURGICAL/ DIAGNOSTICAL INTERVENTION WAS REQUIRED. NO ADVERSE CONSEQUENCES REPORTED. NO DEATH OR (UNANTICIPATED) SERIOUS DETERIORATION IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19088626 ↗
  • InjurySep 13, 2023Energy Output Problem; Insufficient Device Problem Information; Energy Output Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A WORKING ELEMENT. THERE IS NO FAILURE DESCRIPTION AVAILABLE. ACCORDING TO THE INFORMATION RECEIVED THERE WAS AN ADDITIONAL MEDICAL INTERVENTION NECESSARY DUE TO THE INCIDENT REPORTED AND A POOR HAEMOSTASIS OCCURRED. DUE TO LIMITED INFORMATION AND THE INFORMATION MENTIONED ABOVE THIS CASE WAS RATED DEEMED REPORTABLE. ADDITIONAL INFORMATION WAS ALREADY REQUESTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WILL BE UPDATED UNSOLICITED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17738497 ↗
  • InjurySep 13, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH A WORKING ELEMENT. THERE IS NO FAILURE DESCRIPTION AVAILABLE. ACCORDING TO THE INFORMATION RECEIVED THERE WAS AN ADDITIONAL MEDICAL INTERVENTION NECESSARY DUE TO THE INCIDENT REPORTED AND A POOR HAEMOSTASIS OCCURRED . DUE TO LIMITED INFORMATION AND THE INFORMATION MENTIONED ABOVE THIS CASE WAS RATED DEEMED REPORTABLE. ADDITIONAL INFORMATION WAS ALREADY REQUESTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WILL BE UPDATED UNSOLICITED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17738496 ↗

Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices

All ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.