Karl Storz GmbH & Co. KG · Hysteroscope Sheath, 5 mm · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04048551094167), received between Oct 17, 2018 and Jan 29, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Fragmentation2
  • Device Contamination with Chemical or Other Material1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Foreign Body In Patient2
  • Insufficient Information2
  • No Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 29, 2025Break; Break

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2024, THE PHYSICIAN WAS TRYING TO GET THE PRODUCT THROUGH THE CERVIX AND THE SCOPE BROKE. THE HYSTEROSCOPE SHEATH AND HOPKINS TELESCOPE 30°, 2.9 MM, 30 CM WERE ATTACHED (BOTH INSTRUMENTS WERE BEING USED DURING THE PROCEDURE).IT WAS REPORTED DURING THE HYTERSOCOPY PROCEDURE WAS PROLONGED FOR 30 MINUTES ON (B)(6) 2024, THE SURGEON WAS NOT ABLE TO COMPLETE THE PROCEDURE, THE PATIENT WAS REFERRED TO GYN CROSS REFERENCE MDR: 9610617-2025-00026.

    Patient problems: Insufficient Information

    MAUDE report 21270385 ↗
  • MalfunctionJan 14, 2025Break; Improper or Incorrect Procedure or Method; Break; Improper or Incorrect Procedure or Method

    IT WAS REPORTED DURING A MEDICAL PROCEDURE ON (B)(6) 2024, THE PHYSICAN WAS TRYING TO GET THEPRODUCT THROUGH THE CERVIX AND THE SCOPE BROKE. THE HYSTEROSCOPE SHEATH AND HOPKINS TELESCOPE 30°, 2.9 MM, 30 CM WERE ATTACHED (BOTH INSTRUMENTS WERE BEING USED DURING THE PROCEDURE). IT WAS NOT REPORTED IF THERE WAS ANY HARM TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED AFTER MULTIPLE ATTEMPTS. CROSS REFERENCE MDR: 9610617-2025-00026.

    Patient problems: Insufficient Information

    MAUDE report 21150664 ↗
  • InjuryOct 4, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE SURGEON OBSERVED DEBRIS POSSIBLY COMING FROM THE SHEATH. IT WAS AN APPEARANCE OF METALLIC FLAKES IS WAS SMALL ENOUGH THAT IT COULD NOT BE SEEN BY THE NAKED EYE ONLY THROUGH THE SCOPE AND CAMERA.

    Patient problems: Foreign Body In Patient

    MAUDE report 17873605 ↗
  • InjuryOct 4, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE SURGEON OBSERVED DEBRIS POSSIBLY COMING FROM THE SHEATH. IT WAS AN APPEARANCE OF METALLIC FLAKES IS WAS SMALL ENOUGH THAT IT COULD NOT BE SEEN BY THE NAKED EYE ONLY THROUGH THE SCOPE AND CAMERA.

    Patient problems: Foreign Body In Patient

    MAUDE report 17873511 ↗
  • MalfunctionOct 17, 2018Device Contamination with Chemical or Other Material

    ALLEGEDLY, DURING A HYSTEROSCOPY PROCEDURE, THE HOSPITAL REPORTED THEY SAW TISSUE FROM A PREVIOUS CASE COME OUT OF THE INSTRUMENT CHANNEL OF THE SHEATH WHEN AN INSTRUMENT WAS PUT THROUGH THE CHANNEL.

    Patient problems: No Information

    MAUDE report 7976394 ↗

Other LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) devices

All LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) devices →