Skip to content

TAKE-APART 26176HTRecalls & reported problems

Karl Storz GmbH & Co. KG · TAKE-APART® Forceps Insert, 5mm, 33cm · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04048551100059), received between May 22, 2023 and Jun 9, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Energy Output Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Energy Output Problem1
  • Output Problem1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2026Output Problem

    IT WAS REPORTED THAT "THE VET HAS STARTED CAUTERIZING THE FIRST SIDE, GOT HALFWAY THROUGH AND IT STOPPED WORKING. WE CHANGED DIATHERMY UNIT AND TRIED 3 DIFFERENT CABLES, AND IT STILL DIDN'T WORK. WE THEN TRIED OUR DIFFERENT CAUTERIZING INSTRUMENT, AND IT WORKED. THEY HAD A SPARE TAKE-APART BIPOLAR FORCEPS THAT WAS UNSTERILE AND HAD TO BE STERILIZED AGAIN. THE PROCEDURE TOOK 3 HOURS AS OPPOSED TO THE USUAL ONE AND A HALF HOURS". NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25451610 ↗
  • MalfunctionFeb 5, 2026Unexpected Shutdown

    IT WAS REPORTED THAT "THE FORCEPS HAVE STOPPED WORKING. THE INSERT LOOKS TO HAVE BEEN PULLED AWAY FROM THE HANDLE AND NOW A WIRE IS DISCONNECTED (SHE THINKS SOMEONE MAY NOT HAVE PUSHED THE BUTTON TO RELEASE THE INSERT). UNFORTUNATELY, DID NOT HAVE A SPARE AND ALREADY CAUTERIZED THE OVARY, WILL NEED TO CONVERT TO AN OPEN PROCEDURE". NO NEGATIVE IMPACT ON THE STATE OF HEALTH FOR A HUMAN IS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24267551 ↗
  • InjuryMay 22, 2023Energy Output Problem; Energy Output Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PRODUCT 26176HT TAKE-APART BIPOLAR FORCEPS INSERT. ACCORDING TO THE INFORMATION RECEIVED DURING A TUBAL LIGATION THE FORCEPS INSERT OF THE TAKE-APART SET MALFUNCTIONED CAUSING THE COMPLETE SEVERING OF THE PATIENT'S FALLOPIAN TUBE, UNKNOWN SIDE. CASE WAS COMPLETED BUT NOT UNTIL AFTER AN UNPLANNED SALPINGECTOMY WAS PERFORMED. PATIENT POST OPERATIVE CONDITION IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 16978942 ↗

Other LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) devices

All LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) devices →

Sources and updates