Skip to content

SHARK/S-LINE 46300223Recalls & reported problems

Richard Wolf GmbH · BIVAP ELECTRODE BIPO 22FR 12/30° for Shark resectoscopes, telescope Ø 4 mm, suitable for sheaths from 22 Fr, electrode shape: torus, PACK = 3 pcs., sterile, in individual packaging, for single use · Class 2

All models of SHARK/S-LINE (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04055207048931), received between Jan 10, 2019 and May 6, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Fragmentation1
  • Melted1
  • Thermal Decomposition of Device1

Patient problems

  • Burn, Thermal1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2021Thermal Decomposition of Device; Material Fragmentation; Melted; Thermal Decomposition of Device; Material Fragmentation; Melted

    IT WAS REPORTED BY THE USER FACILITY TO RICHARD WOLF THAT "AFTER 40 MINUTES OF USE, THE INSULATION MELTED, ONE OF THE PRONGS BROKE, AND THE INSTRUMENT BECAME ENCRUSTED WITH A BLACK CHARRED MATERIAL." THE ITEM WAS DISCARDED AT THE FACILITY. ADDITIONALLY: WILL THE DEVICE BE RETURNED? NO. WAS THE DEVICE BEING USED ON A PATIENT WHEN THE REPORTING ISSUE OCCURRED? YES. WAS THERE ANY INJURY OR ILLNESS TO THE PATIENT DUE TO THE REPORTED ISSUE? NO. WAS THERE ANY INJURY OR ILLNESS TO ANY OTHER PERSONNEL DUE TO THE REPORTED ISSUE? NO. DID THE ISSUE CAUSE A DELAY IN THE PROCEDURE BEING PERFORMED? NO. DID THE DELAY PUT THE PATIENT AT RISK? N/A. WAS THERE A SIMILAR BACK-UP DEVICE AVAILABLE FOR USE? YES. WAS THE SCHEDULED PROCEDURE COMPLETED? YES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11787704 ↗
  • InjuryJan 10, 2019Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2018, RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) WAS SERVED WITH A PATIENT LAW SUIT INVOLVING A "BUTTON TYPE BIPOLAR PROSTATE RESECTOR AND/OR ELECTROSURGICAL GENERATOR". ON (B)(6) 2016, THE PATIENT UNDERWENT A PROCEDURE DESCRIBED AS "CYSTOSCOPY WITH URETHRAL DILATION, TRANSURETHRAL RESECTION/ABLATION OF PROSTATE WITH BIPOLAR". DURING SAID PROCEDURE, THE PATIENT WAS UNDER GENERAL ANESTHESIA AND SUFFERED THERMAL INJURY TO THEIR BLADDER AND URETHRA. AT THIS TIME, THE INVOLVEMENT OF RICHARD WOLF PRODUCTS HAS NOT BEEN CONFIRMED. HOWEVER, THE FOLLOWING ARE CONSIDERED SUSPECT PRODUCTS AND MDRS WILL BE SUBMITTED FOR EACH UNTIL THE USER FACILITY IS ABLE TO CONFIRM THAT NO RICHARD WOLF PRODUCTS WERE INVOLVED: 8680.225 WORKING ELEMENT PASSIVE BIPO 0/12/30 (MDR 1418479-2018-00051), 8680.224 WORKING ELEMENT PASSIVE BIPO 0/12/30 (MDR 1418479-2018-00052), 46300223 BIVAP ELECTRODE BIPO 22FR 12/30 (MDR), 46300243 BIVAP ELECTRODE BIPO 24-26FR 12/30 (1418479-2018-00054).

    Patient problems: Burn, Thermal

    MAUDE report 8237969 ↗

FDA clearance and classification

The FDA regulates product code FAS (electrode, electrosurgical, active, urological) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SHARK/S-LINE 46300223

Has the SHARK/S-LINE 46300223 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04055207048931), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SHARK/S-LINE 46300223?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 10, 2019 and May 6, 2021. In those reports, the most common specific device problems are “Material Fragmentation” (1), “Melted” (1) and “Thermal Decomposition of Device” (1); the most common patient problem is “Burn, Thermal” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SHARK/S-LINE 46300223?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the SHARK/S-LINE 46300223, and is it still sold?

The FDA's device database lists Richard Wolf GmbH as the company responsible for it (brand SHARK/S-LINE, model 46300223). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices

All ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page