SHARK 46782435Recalls & reported problems
Richard Wolf GmbH · CUTTING ELECTRODE MONO 24FR 12/30° for resectoscopes, telescope Ø 4 mm, for sheaths, intermittent/continuous irrigation, 24/26 Fr, loop: round, wire Ø 0.35 mm, fork color yellow, stem color transparent, PACK = 10 PCS, for Shark Resectoscope, sterile, for single use · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04055207049051), received between Oct 23, 2020 and May 23, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Fire2
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2023Fire; Fire
ON APRIL 24, 2023, RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) RECEIVED A MEDWATCH REPORT # (B)(4) REGARDING A WORKING ELEMENT PASSIVE MONO 12/30° PART 8674225, BATCH # UNKNOWN. ACCORDING TO THE RECEIVED INFORMATION, "WHILE USING THE RESECTOSCOPE. APPARENT THERE WAS A FIRE ON THE HAND PIECE TO THE RESECTOSCOPE DURING A PROSTATE PROCEDURE. NO HARM TO THE PATIENT OR SURGEON DURING THE EVENT. THE WORKING ELEMENT CAUGHT ON FIRE DURING USE." IT IS SUSPECTED THAT THE CUTTING ELECTRODE MONO 24FR 12/30°, PART ID: 46782435, BATCH # 21003532, WITHIN THE HAND PIECE (WORKING ELEMENT PASSIVE MONO 12/30° PART 8674225, BATCH # UNKNOWN) IS WHAT HAD IGNITED. THE REPORTED ISSUE CAUSED A 15 MINUTE DELAY IN THE PROCEDURE WHILE THE FIRE WAS PUT OUT AND THE BACK-UP DEVICE RETRIEVED TO COMPLETE THE SCHEDULED PROCEDURE. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16989269 ↗MalfunctionMay 23, 2023Fire; Fire
ON APRIL 24, 2023, RICHARD WOLF MEDICAL INSTRUMENTS CORP. (RWMIC) RECEIVED A MEDWATCH REPORT # (B)(4) REGARDING A WORKING ELEMENT PASSIVE MONO 12/30° PART 8674225, BATCH # UNKNOWN. ACCORDING TO THE RECEIVED INFORMATION, "WHILE USING THE RESECTOSCOPE. APPARENT THERE WAS A FIRE ON THE HAND PIECE TO THE RESECTOSCOPE DURING A PROSTATE PROCEDURE. NO HARM TO THE PATIENT OR SURGEON DURING THE EVENT. THE WORKING ELEMENT CAUGHT ON FIRE DURING USE." IT IS SUSPECTED THAT THE CUTTING ELECTRODE MONO 24FR 12/30°, PART ID: 46782435, BATCH # 21003532, WITHIN THE HAND PIECE (WORKING ELEMENT PASSIVE MONO 12/30° PART 8674225, BATCH # UNKNOWN) IS WHAT HAD IGNITED. THE REPORTED ISSUE CAUSED A 15 MINUTE DELAY IN THE PROCEDURE WHILE THE FIRE WAS PUT OUT AND THE BACK-UP DEVICE RETRIEVED TO COMPLETE THE SCHEDULED PROCEDURE. THERE IS NO REPORT OF INJURY TO THE PATIENT OR OTHER PERSONNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16989026 ↗MalfunctionOct 23, 2020Break; Break
THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION. THE DEVICE WAS RECEIVED VISUALLY EVALUATED. THE REPORTED CONDITION (LOOPS BURNED) WAS CONFIRMED. ACCORDING TO THE MANUFACTURER: "AFTER CONSULTATION WITH THE USER IT WAS DETERMINED THAT THE ELECTRODES WERE USED WITH A MEDTRONIC GENERATOR. THIS CAUSED THE USER TO USE 3-4 ELECTRODES PER TURP. THE GENERATOR WAS REPLACED AND IT WAS DETERMINED THAT THE MEDTRONIC GENERATOR WAS CAUSING THE LOOPS TO BURN OUT COMPLETELY. ACCORDING TO THE SUPPLEMENT BB-D366, SYSTEM OVERVIEW "SHARK" RESECTOSCOPES, THE SYSTEM COMPONENTS LISTED IN THE TABLES MUST ONLY BE USED IN COMPLIANCE WITH THE GA-D366 INSTRUCTIONS FOR USE. IN GENERAL, THE USER IS ADVISED IN THE CORRESPONDING INSTRUCTIONS FOR USE GA-D366/EN/2018-03 V5.0 PK 18-9297 UNDER SECTION 8 THAT A VISUAL AND FUNCTIONAL CHECK MUST BE CARRIED OUT BEFORE USE. DO NOT USE PRODUCTS THAT ARE DAMAGED, INCOMPLETE OR HAVE LOOSE PARTS. POSSIBLE DAMAGE OR DEVIATIONS CAN BE EASILY DETECTED BY HOSPITAL PERSONNEL.". PROBABLE ROOT CAUSE IS THOUGHT TO BE USER HANDLING ISSUE. THE ELECTRODES WERE SCRAPPED.
Patient problems: Insufficient Information
MAUDE report 10728745 ↗MalfunctionOct 23, 2020Break; Break
THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE INVESTIGATION. THE DEVICE WAS RECEIVED VISUALLY EVALUATED. THE REPORTED CONDITION (LOOPS BURNED) WAS CONFIRMED. ACCORDING TO THE MANUFACTURER: "AFTER CONSULTATION WITH THE USER IT WAS DETERMINED THAT THE ELECTRODES WERE USED WITH A MEDTRONIC GENERATOR. THIS CAUSED THE USER TO USE 3-4 ELECTRODES PER TURP. THE GENERATOR WAS REPLACED AND IT WAS DETERMINED THAT THE MEDTRONIC GENERATOR WAS CAUSING THE LOOPS TO BURN OUT COMPLETELY. ACCORDING TO THE SUPPLEMENT BB-D366, SYSTEM OVERVIEW "SHARK" RESECTOSCOPES, THE SYSTEM COMPONENTS LISTED IN THE TABLES MUST ONLY BE USED IN COMPLIANCE WITH THE GA-D366 INSTRUCTIONS FOR USE. IN GENERAL, THE USER IS ADVISED IN THE CORRESPONDING INSTRUCTIONS FOR USE GA-D366/EN/2018-03 V5.0 PK 18-9297 UNDER SECTION 8 THAT A VISUAL AND FUNCTIONAL CHECK MUST BE CARRIED OUT BEFORE USE. DO NOT USE PRODUCTS THAT ARE DAMAGED, INCOMPLETE OR HAVE LOOSE PARTS. POSSIBLE DAMAGE OR DEVIATIONS CAN BE EASILY DETECTED BY HOSPITAL PERSONNEL." PROBABLE ROOT CAUSE IS THOUGHT TO BE USER HANDLING ISSUE THE ELECTRODES WERE SCRAPPED.
Patient problems: Insufficient Information
MAUDE report 10727962 ↗
Other ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL devices
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22332DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22301DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page