Artis Q.zen floor, Model Number 10848353 recallZ-0978-2022 · Siemens Medical Solutions USA, Inc · initiated Feb 23, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
FDA root cause category: Device Design
What the company is doing
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 2/23/22 was sent to customers. Which steps have to be taken by the user to avoid the possible risks associated with this problem? As also described in the Operator Manual, we recommend that the system operator checks the water level of the cooling circuit at least every three months and refills water, if necessary: 1. Open the filling gland of the cooling unit. The water surface must be clearly visible above the cooling ribs. 2. Replenish with water (drinking water quality) if cooling liquid is lacking. Please inform the service technician in case of lacking cooling liquid. How will the corrective action be implemented and what is the efficiency of the corrective actions? The tube cooling unit will be equipped with a water level sensor and the software will be updated accordingly. In case the coolant level drops below a certain threshold the following message will be displayed: Tube cooling water level low: Refill as per Operator Manual . Siemens will correct the error with a hardware and software update via Update Instruction AX002/20/S and AX004/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand t
Recall details
- Recall number
- Z-0978-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 23, 2022
- Posted by the FDA
- Apr 21, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 33 systems
- Product code
- OWB
- FDA submissions
- K181407
- Distribution
- US Nationwide.
Product description
Artis Q.zen floor, Model Number 10848353
Code information (models, lots, serial numbers)
UDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125 Additional Serial Numbers as of 10/4/23: 143500
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
Other devices under the same FDA submission
11 devices cleared under K181407 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Artis zeego 10848283Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system5 recalls1 death7 reports
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- Artis zee floor 10094135Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls2 deaths94 reports
- Artis zee ceiling 10094137Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls91 reports
- Artis zeego 10280959Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls1 death24 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen biplane 10848355Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls5 reports
- Artis zee multi-purpose 10094139Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls26 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 89860).
- Z-0967-2022 · Open, ClassifiedArtis zee floor, Model Number 10094135
- Z-0968-2022 · Open, ClassifiedArtis zee ceiling, Model Number 10094137
- Z-0969-2022 · Open, ClassifiedArtis zee multi-purpose, Model Number 10094139
- Z-0970-2022 · Open, ClassifiedArtis zee biplane, Model Number 10094141
- Z-0971-2022 · Open, ClassifiedArtis zee floor MN, Model Number 10094142
- Z-0972-2022 · Open, ClassifiedArtis zee biplane MN, Model Number 10094143
- Z-0973-2022 · Open, ClassifiedArtis zeego, Model Number 10280959
- Z-0974-2022 · Open, ClassifiedArtis Q floor, Model Number 10848280
- Z-0975-2022 · Open, ClassifiedArtis Q ceiling, Model Number 10848281
- Z-0976-2022 · Open, ClassifiedArtis Q biplane, Model Number 10848282
- Z-0977-2022 · Open, ClassifiedArtis Q zeego, Model Number 10848283
- Z-0979-2022 · Open, ClassifiedArtis Q.zen ceiling, Model Number 10848354
- Z-0980-2022 · Open, ClassifiedArtis Q.zen biplane, Model Number 10848355
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page