BLUEENDO 02-33117Recalls & reported problems

TROKAMED GmbH · Grip Module V2.0, 5-times-use for Morcellator System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04251303811435), received between Jun 27, 2023 and Dec 5, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Shut Off” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Shut Off2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2023Failure to Shut Off; Failure to Shut Off

    THE PROBLEM WAS REPORTED TO US VIA CUSTOMER COMPLAINT (B)(4) ON (B)(6) 2023: WHEN USING THE MORCELLATOR, THE CUTTING MODULE CONTINUED TO ROTATE EVEN THOUGH THE ACTIVATION SWITCH OF THE HANDLE MODULE WAS RELEASED BY THE USER. THE UB-20230501_MORCELLATOR_SCHALTER INVESTIGATION REPORT WAS PREPARED TO ANALYZE THE CAUSE.THE CAUSE OF THE ERROR PATTERN WAS FOUND TO BE A FAULTY SWITCH FUNCTION, I.E. THE CONTACT DISTANCES OF THE SWITCH MAY BE TOO SMALL. THE SEAT OF THE FIXING SCREW OF THE ACTIVATION SWITCH IN THE GRIP MODULE WAS TESTED WITH AN UNSUITABLE TESTING DEVICE. AS A RESULT, PRODUCTS WITH TOO SMALL A SWITCH CONTACT DISTANCE (CONTACT SCREW TO SPRING PLATE) HAVE COME INTO CIRCULATION. WITH THESE PRODUCTS, IT IS THEREFORE POSSIBLE THAT THE GRIP MODULE TRANSMITS AN "ON" SIGNAL TO THE CONTROL UNIT EVEN AFTER THE ACTIVATION SWITCH HAS BEEN RELEASED AND THAT THE CUTTING MODULE CONTINUES TO ROTATE EVEN AFTER THE ACTIVATION SWITCH HAS BEEN RELEASED.REASON FOR FSCA: URGENT SAFETY INFORMATION IS BEING ISSUED TO REDUCE THE RISK OF THE OCCURRENCE OF A SIGNIFICANT DELAY IN THE MEDICAL PROCEDURE IF A USER HAS TO MAKE AN UNPLANNED CHANGE FROM HAND TO FOOT CONTROL DURING A PROCEDURE DUE TO THE SUDDE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265377 ↗
  • MalfunctionJun 27, 2023Failure to Shut Off; Failure to Shut Off

    THE PROBLEM WAS REPORTED TO US VIA CUSTOMER COMPLAINT (B)(4) ON (B)(6).2023: WHEN USING THE MORCELLATOR, THE CUTTING MODULE CONTINUED TO ROTATE EVEN THOUGH THE ACTIVATION SWITCH OF THE HANDLE MODULE WAS RELEASED BY THE USER. THE UB-20230501_MORCELLATOR_SCHALTER INVESTIGATION REPORT WAS PREPARED TO ANALYZE THE CAUSE.THE CAUSE OF THE ERROR PATTERN WAS FOUND TO BE A FAULTY SWITCH FUNCTION, I.E. THE CONTACT DISTANCES OF THE SWITCH MAY BE TOO SMALL. THE SEAT OF THE FIXING SCREW OF THE ACTIVATION SWITCH IN THE GRIP MODULE WAS TESTED WITH AN UNSUITABLE TESTING DEVICE. AS A RESULT, PRODUCTS WITH TOO SMALL A SWITCH CONTACT DISTANCE (CONTACT SCREW TO SPRING PLATE) HAVE COME INTO CIRCULATION. WITH THESE PRODUCTS, IT IS THEREFORE POSSIBLE THAT THE GRIP MODULE TRANSMITS AN "ON" SIGNAL TO THE CONTROL UNIT EVEN AFTER THE ACTIVATION SWITCH HAS BEEN RELEASED AND THAT THE CUTTING MODULE CONTINUES TO ROTATE EVEN AFTER THE ACTIVATION SWITCH HAS BEEN RELEASED.REASON FOR FSCA: URGENT SAFETY INFORMATION IS BEING ISSUED TO REDUCE THE RISK OF THE OCCURRENCE OF A SIGNIFICANT DELAY IN THE MEDICAL PROCEDURE IF A USER HAS TO MAKE AN UNPLANNED CHANGE FROM HAND TO FOOT CONTROL DURING A PROCEDURE DUE TO THE SUDDE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17210931 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →