VIDEO SYSTEM OTV-SI (UDI 04953170170461)Recalls & reported problems

Olympus Medical Systems Corp. · "OTV-SI(B) OTV-SI OFFICE VID SYSTM W/PUMP" · Class 2

Quick summary

This Olympus OTV-SI office video system is a video unit for endoscopes. Its 143 reports often describe no picture, a flickering display, lamp errors or poor image quality, and most report no patient harm. There are no deaths and no injuries. No recalls name this device. Reports fell to 1 in the last 12 months from 7 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
143
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 143 adverse event reports that list this exact device (DI 04953170170461), received between Apr 6, 2020 and Mar 2, 2026 — classified as 143 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 4 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image32
  • Erratic or Intermittent Display11
  • Communication or Transmission Problem10
  • Optical Problem10
  • Poor Quality Image9
  • Display or Visual Feedback Problem8
  • Output Problem8
  • No Visual Prompts/Feedback7

Patient problems

  • No Clinical Signs, Symptoms or Conditions113
  • No Patient Involvement17
  • No Consequences Or Impact To Patient16

Recent adverse event reports

The 25 most recent of 143 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2026Communication or Transmission Problem

    IT WAS REPORTED THE VIDEO SYSTEM WAS GETTING A LAMP ERROR B. WAS NOT ABLE TO CLEAR IT EVEN THOUGH LAMP A WAS WORKING. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24488365 ↗
  • MalfunctionOct 28, 2025Activation Problem; Activation Problem

    IT WAS OBSERVED DURING THE DEVICE INVESTIGATION THAT THE SUBJECT DEVICE EXHIBITED A LAMP A ERROR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23407982 ↗
  • MalfunctionMay 23, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22078937 ↗
  • MalfunctionMay 9, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE SUBJECT DEVICE DISPLAYED LAMP A ERROR DURING A DIAGNOSTIC NASOPHARYNX SCOPE EXAM. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21995027 ↗
  • MalfunctionMay 2, 2025Optical Problem; Optical Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21953620 ↗
  • MalfunctionApr 24, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXHIBITED NO INPUT ON THE COMBINATION DEVICE USED ALONG WITH IT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21905087 ↗
  • MalfunctionFeb 7, 2025Corroded; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM EXHIBITED CONTACT PART ON THE VIDEO CONNECTOR RECEPTACLE IS CORRODED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21332933 ↗
  • MalfunctionNov 22, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20758474 ↗
  • MalfunctionOct 17, 2024Output Problem; Output Problem

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER EXHIBITED A LAMP ERROR. IT IS UNKNOWN, WHEN THE REPORTED ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20478112 ↗
  • MalfunctionOct 16, 2024Image Display Error/Artifact; Image Display Error/Artifact

    IT WAS REPORTED THAT THE VIDEO SYSTEM DISPLAYED A LAMP ERROR THAT WOULD NOT DISAPPEAR. THE ISSUE OCCURRED DURING AN UNKNOWN EVENT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20463327 ↗
  • MalfunctionOct 11, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT, THE VIDEO SYSTEM HAD BLACK LINES APPEAR ON THE IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20430410 ↗
  • MalfunctionSep 25, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE VIDEO SYSTEM HAD A LAMP ERROR. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20301494 ↗
  • MalfunctionSep 12, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED, THE VIDEO SYSTEM EXHIBITED INTERMITTENT VIDEO OUT PROBLEM. THE ISSUE OCCURRED, DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20202482 ↗
  • MalfunctionSep 4, 2024No Visual Prompts/Feedback; No Visual Prompts/Feedback

    IT WAS REPORTED THE VIDEO SYSTEM CENTER'S HALOGEN LAMP IS WORN OUT AND DOES NOT LIGHT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20135134 ↗
  • MalfunctionAug 21, 2024Output Problem; Output Problem

    IT WAS REPORTED, THE VIDEO SYSTEM EXHIBITED ERROR IN LAMP A. THERE WERE NO REPORTS OF PATIENT HARM. WHILE SPEAKING WITH THE OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC), THE CUSTOMER WAS ADVISED TO START BY CHANGING THE LAMP. THE CUSTOMER WAS THEN TRANSFERRED TO CUSTOMER SERVICE TO PLACE AN ORDER FOR A LAMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20017593 ↗
  • MalfunctionAug 6, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THAT THE VIDEO SYSTEM SMELL LIKE IT WAS BURNING WHEN THE POWER WAS ON. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19924914 ↗
  • MalfunctionAug 5, 2024No Visual Prompts/Feedback; No Visual Prompts/Feedback

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM EXHIBITED A LAMP ERROR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19910072 ↗
  • MalfunctionJul 10, 2024Failure to Power Up; No Visual Prompts/Feedback; Failure to Power Up; No Visual Prompts/Feedback

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM EXHIBITED THE POWER DOES NOT TURN ON DUE TO A FAULTY INTERLOCK SWITCH AND BOTH LAMPS ARE BLOWN DUE TO A LAMP ERROR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19711438 ↗
  • MalfunctionJun 28, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THAT THE VIDEO SYSTEM ANGLE OF THE SCOPE INSERTION PORT SOMETIMES PREVENTED THE IMAGE FROM BEING DISPLAYED ON THE MONITOR. THE ISSUE OCCURRED, DURING AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19636261 ↗
  • MalfunctionMay 9, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    THE EVENT OCCURRED DURING A DIAGNOSTIC PROCEDURE THAT WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19283212 ↗
  • MalfunctionFeb 23, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED THAT THE AUTO DISPLAY DOES NOT LIGHT UP DUE TO FRONT PANEL FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18767778 ↗
  • MalfunctionFeb 22, 2024Output Problem; Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM HAD A LAMP A ERROR. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18761199 ↗
  • MalfunctionJan 28, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER HAD NO SIGNAL OR IMAGE ON THE LIGHT GUIDE MONITOR. THE ISSUE WAS FOUND DURING PREPARATION FOR USE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER SIMILAR DEVICE. THE PROCEDURE WAS DELAYED 24 HOURS BUT THERE WERE NO REPORTS OF PATIENT HARM OR MEDICAL INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18592916 ↗
  • MalfunctionJan 26, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THE IMAGE ON THE VIDEO PROCESSOR WAS FREEZING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18592354 ↗
  • MalfunctionJan 22, 2024Defective Component; Defective Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE VIDEO SYSTEM CENTER HAD A DEAD BATTERY. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE LAMP A ERROR DUE TO LAMP A DAMAGE WAS FOUND. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18555555 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →