HD CAMERA HEAD OTV-S7ProH-HD-L08ERecalls & reported problems
Olympus Medical Systems Corp. · "OTV-S7PROH-HD-L08E HD URO CAMERA" · Class 2
All models of HD CAMERA HEAD (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus HD urology camera head has 1,410 FDA adverse event reports, with 1 injury, 1,409 malfunctions and no deaths. The reports mostly describe no image on the display, poor image quality, and the camera head failing leak tests, with some describing torn material or loose connections. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,392
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,392 adverse event reports that list this exact device (DI 04953170268267), received between Nov 10, 2020 and Jul 14, 2026 — classified as 1 injury, 1,391 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (535 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image535
- Poor Quality Image321
- Material Split, Cut or Torn102
- Leak/Splash73
- Loose or Intermittent Connection63
- Break61
- Erratic or Intermittent Display60
- Gas/Air Leak43
Patient problems
- No Clinical Signs, Symptoms or Conditions1,388
- No Consequences Or Impact To Patient2
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 1,392 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2026Detachment of Device or Device Component
THE CUSTOMER REPORTED MISSING CAMERA HEAD LENS DURING REPROCESSING INVOLVING AN OLYMPUS CAMERA HEAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25815898 ↗MalfunctionJul 14, 2026Leak/Splash
THE CUSTOMER RETURNED THE CAMERA HEAD, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, A FAILED LEAK TEST WAS OBSERVED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25815204 ↗MalfunctionJul 14, 2026No Display/Image
THE CUSTOMER REPORTED NO IMAGE DURING INSPECTION FOR USE INVOLVING AN OLYMPUS CAMERA HEAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25813982 ↗MalfunctionJul 13, 2026Leak/Splash
THE CUSTOMER RETURNED THE CAMERA HEAD TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING THE REPAIR CENTER TECHNICIAN FOUND FAILED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25801582 ↗MalfunctionJul 8, 2026Loose or Intermittent Connection
THE CUSTOMER REPORTED CAMERA LOOKS WIGGLEY DURING A INSPECTION FOR USE PROCEDURE INVOLVING AN OLYMPUS CAMERA HEAD. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25750833 ↗MalfunctionJun 28, 2026Gas/Air Leak
THE CUSTOMER RETURNED THE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT FAILED LEAK TEST WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25647679 ↗MalfunctionJun 26, 2026Scratched Material; Material Split, Cut or Torn
THE CUSTOMER RETURNED THE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE EXHIBITED A LENS IN WHICH THE CHARGE COUPLED DEVICE (CCD) COVER GLASS HAD DEEP SCRATCHES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25643878 ↗MalfunctionJun 22, 2026Scratched Material
THE CUSTOMER RETURNED THE CAMERA HEAD TO THE SERVICE CENTER FOR EVALUATION RELATED TO SEPARATE ISSUE. DURING TESTING, THE SERVICE TECHNICIAN IDENTIFIED THE DEVICE HAD LENS HAD MULTIPLE DEEP SCRATCHES ON CHARGED COUPLED DEVICE COVER GLASS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25580028 ↗MalfunctionJun 17, 2026Break
THE CUSTOMER REPORTED CHARGED COUPLED DEVICE DAMAGE, INVOLVING AN OLYMPUS CAMERA HEAD. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25537804 ↗MalfunctionJun 17, 2026Material Integrity Problem
THE CUSTOMER REPORTED PULLED WIRE DURING REPROCESSING INVOLVING AN OLYMPUS CAMERA HEAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25533378 ↗MalfunctionJun 16, 2026Gas/Air Leak
THE CUSTOMER RETURNED THE CAMERA HEAD, WHICH IS AN OLYMPUS ASSET FOR A SEPARATE ISSUE. DURING THE EVALUATION OF THE DEVICE, LEAKAGE WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE FINDING WAS DUE TO KINKS AND PUNCTURED CABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25527059 ↗MalfunctionJun 12, 2026Poor Quality Image
THE CUSTOMER REPORTED CLOUDY IMAGE DURING REPROCESSING INVOLVING AN OLYMPUS CAMERA HEAD. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25487075 ↗MalfunctionJun 10, 2026Gas/Air Leak
THE CUSTOMER RETURNED THE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPORT TECHNICIAN INDICATES THAT LEAK IN CABLE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25457570 ↗MalfunctionMay 22, 2026Gas/Air Leak
THE AUTOCLAVABLE CAMERA HEAD WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF CORD CUT. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, CAMERA FAILED LEAK TEST FROM CABLE CONNECTOR WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260198 ↗MalfunctionMay 20, 2026Detachment of Device or Device Component
IT WAS REPORTED THE AUTOCLAVABLE CAMERA HEAD HAD SOME PLASTIC PIECES AROUND THE LENS THAT CAME APART. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229409 ↗MalfunctionMay 11, 2026Poor Quality Image
IT WAS REPORTED THAT THE CAMERA HEAD HAD A BLURRY PICTURE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25135204 ↗MalfunctionMay 8, 2026Loose or Intermittent Connection; Poor Quality Image
IT WAS REPORTED THE CAMERA HEAD PART THAT HOOKS TO THE SCOPE IS LOOSE AND THE IMAGE IS HARD TO SEE. THIS ISSUE OCCURRED DURING CYSTOSCOPY WITH LASER LITHOTRIPSY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25127449 ↗MalfunctionMay 7, 2026Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED YELLOW LINE ON THE IMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25111998 ↗MalfunctionApr 30, 2026No Display/Image
IT WAS REPORTED THAT THE CAMERA HEAD COULD NOT DISPLAY AN IMAGE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25043633 ↗MalfunctionApr 27, 2026Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED A CUT ON INSULATION CABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25010631 ↗MalfunctionApr 27, 2026No Display/Image
IT WAS REPORTED THE CAMERA HEAD EXHIBITED NO PICTURE. THE ISSUE OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25001073 ↗MalfunctionApr 25, 2026Poor Quality Image
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24991875 ↗MalfunctionApr 22, 2026Leak/Splash
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED LEAK FROM CABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24956318 ↗MalfunctionApr 22, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED CHARGED COUPLE DEVICE COVER GLASS IS MISSING IN THE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24952721 ↗MalfunctionApr 16, 2026Poor Quality Image
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24895484 ↗
FDA clearance and classification
Cleared through 510(k) K083155 on Jan 13, 2009 (decision: Substantially Equivalent). Submitted by Olympus Medical Systems Corporation as “Hd Camera Head Otv-S7proh-Hd-L08e”. The FDA review took 81 days. FDA record ↗
The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the HD CAMERA HEAD OTV-S7ProH-HD-L08E
Has the HD CAMERA HEAD OTV-S7ProH-HD-L08E been recalled?
No FDA recall record in our data names this exact device identifier (DI 04953170268267), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the HD CAMERA HEAD OTV-S7ProH-HD-L08E?
The FDA's MAUDE database lists 1,392 adverse event reports for this device, received between Nov 10, 2020 and Jul 14, 2026. In those reports, the most common specific device problems are “No Display/Image” (535), “Poor Quality Image” (321) and “Material Split, Cut or Torn” (102). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HD CAMERA HEAD OTV-S7ProH-HD-L08E?
None of the reports are classified as deaths, 1 as an injury and 1,391 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HD CAMERA HEAD OTV-S7ProH-HD-L08E FDA approved?
It was cleared through 510(k) K083155 on Jan 13, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HD CAMERA HEAD OTV-S7ProH-HD-L08E, and is it still sold?
The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand HD CAMERA HEAD, model OTV-S7ProH-HD-L08E). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,484 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,290 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology526 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths409 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall183 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology120 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page