Skip to content

4K CAMERA HEAD OLYMPUS CH-S400-XZ-EARecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of 4K CAMERA HEAD (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus 4K camera head is the imaging unit used with endoscopes for gastroenterology and urology procedures. Its 2,353 reports often describe no image or display on the screen, communication problems and poor image quality, and mention 7 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,375
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,375 adverse event reports that list this exact device (DI 04953170375552), received between Jul 2, 2020 and Sep 30, 2026 — classified as 7 injuries, 2,368 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (1,147 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 38 reportsDec 2023: 41 reportsJan 2024: 49 reportsFeb 2024: 58 reportsMar 2024: 67 reportsApr 2024: 56 reportsMay 2024: 58 reportsJun 2024: 65 reportsJul 2024: 71 reportsAug 2024: 44 reportsSep 2024: 55 reportsOct 2024: 80 reportsNov 2024: 30 reportsDec 2024: 20 reportsJan 2025: 34 reportsFeb 2025: 10 reportsMar 2025: 30 reportsApr 2025: 44 reportsMay 2025: 29 reportsJun 2025: 31 reportsJul 2025: 38 reportsAug 2025: 56 reportsSep 2025: 16 reportsOct 2025: 44 reportsNov 2025: 36 reportsDec 2025: 32 reportsJan 2026: 31 reportsFeb 2026: 37 reportsMar 2026: 33 reportsApr 2026: 40 reportsMay 2026: 55 reportsJun 2026: 32 reportsJul 2026: 27 reportsAug 2026: 29 reportsSep 2026: 24 reportsOct 2026: 0 reports20242025202680
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image1,147
  • Communication or Transmission Problem469
  • Poor Quality Image370
  • Connection Problem104
  • Leak/Splash72
  • Output Problem65
  • Erratic or Intermittent Display62
  • Material Split, Cut or Torn62

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,334
  • No Patient Involvement25
  • No Consequences Or Impact To Patient12
  • Insufficient Information3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 2,375 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Gas/Air Leak; Material Split, Cut or Torn

    The customer returned the autoclavable camera head for a separate issue. During the initial evaluation of the device, it was observed that failed leak test and cut in video cable observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950406 ↗
  • MalfunctionSep 30, 2026No Display/Image; Poor Quality Image

    The customer reported that color bars were displayed on the screen during a therapeutic procedure involving an Olympus Autoclavable Camera Head. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947861 ↗
  • MalfunctionSep 29, 2026Gas/Air Leak

    The customer returned the Olympus camera head for a separate issue. During the initial evaluation of the returned device, a failed leak test was observed between the cos ring and rear cover. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936408 ↗
  • MalfunctionSep 29, 2026Erratic or Intermittent Display; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926943 ↗
  • MalfunctionSep 28, 2026No Device Output

    The customer reported that no functions were available during inspection for use involving an Olympus autoclavable camera head. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916466 ↗
  • MalfunctionSep 28, 2026Poor Quality Image; Communication or Transmission Problem

    The customer reported that the camera head picture was not clear during reprocessing. During the initial evaluation of the device, a camera head communication error (E224) was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916322 ↗
  • MalfunctionSep 25, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899956 ↗
  • MalfunctionSep 18, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50818709 ↗
  • MalfunctionSep 17, 2026No Display/Image; Connection Problem; Gas/Air Leak

    The customer reported whenever they were plugged into the processor, nothing happened during inspection for use of an Olympus autoclavable camera head. During the initial evaluation of the device, it was observed no image displayed on the monitor due to a faulty cable/connector; and water leaked between the rear cover and COS ring. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808882 ↗
  • MalfunctionSep 17, 2026No Display/Image; Loose or Intermittent Connection; Gas/Air Leak

    The customer reported that autoclavable camera head was not recognized during inspection for use. During the initial evaluation of the device, the unit failed the water leak test, and no image was observed. No patient involvement was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50802952 ↗
  • MalfunctionSep 17, 2026Poor Quality Image; Gas/Air Leak

    It was reported that the camera head had a blurry image. During evaluation the service center observe a leak. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50801455 ↗
  • MalfunctionSep 15, 2026Connection Problem; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50780308 ↗
  • MalfunctionSep 15, 2026No Display/Image

    The customer reported no picture issue during an unspecified procedure involving an Olympus camera head. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50773954 ↗
  • MalfunctionSep 14, 2026No Display/Image; Communication or Transmission Problem

    The customer reported light cord was not white balance involving an Olympus camera head. During initial evaluation it was observer, Error code E224 showing on image. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50755969 ↗
  • MalfunctionSep 14, 2026No Display/Image

    The customer reported no image, at all just a black screen during inspection for use involving an Olympus Autoclavable Camera Head. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50754832 ↗
  • MalfunctionSep 14, 2026Break; Poor Quality Image

    The customer reported bad image and defect in the glass involving an Olympus autoclavable camera head. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50751809 ↗
  • MalfunctionSep 11, 2026Display or Visual Feedback Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50741772 ↗
  • MalfunctionSep 11, 2026Poor Quality Image; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50739764 ↗
  • MalfunctionSep 11, 2026No Display/Image; Leak/Splash; Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50736874 ↗
  • MalfunctionSep 11, 2026Poor Quality Image; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50735597 ↗
  • MalfunctionSep 11, 2026Break; Poor Quality Image

    The customer reported a bad image and a defect in the glass involving an Olympus autoclavable camera head. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50728895 ↗
  • MalfunctionSep 10, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716736 ↗
  • MalfunctionSep 9, 2026Erratic or Intermittent Display; No Display/Image

    The customer reported the autoclavable camera head dull light source image flickering. During the initial evaluation of the returned device, no image was shown on the screen. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50682288 ↗
  • MalfunctionSep 2, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50555958 ↗
  • MalfunctionAug 31, 2026Blocked Connection

    Patient problems: Insufficient Information

    MAUDE report 50533176 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →