VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of VISERA ELITE II VIDEO SYSTEM CENTER (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus VISERA Elite II video system is the central unit that processes images from endoscopes. Many reports describe communication failures and blank or abnormal displays, often found during inspection before use. The 158 reports include no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
158
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 158 adverse event reports that list this exact device (DI 04953170386060), received between Jan 11, 2022 and Nov 18, 2025 — classified as 158 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (50 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 5 reportsJan 2024: 3 reportsFeb 2024: 5 reportsMar 2024: 3 reportsApr 2024: 8 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem50
  • No Display/Image44
  • Electrical /Electronic Property Problem11
  • Output Problem11
  • Display or Visual Feedback Problem8
  • Defective Component7
  • Poor Quality Image7
  • Erratic or Intermittent Display3

Patient problems

  • No Clinical Signs, Symptoms or Conditions158
  • Electric Shock1

Recent adverse event reports

The 25 most recent of 158 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER HAD A LOOSENESS IN THE SOCKET, CAUSING AN UNSTABLE CONNECTION. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583879 ↗
  • MalfunctionNov 17, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23568987 ↗
  • MalfunctionSep 8, 2025Output Problem; Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED LIGHT-EMITTING DIODE UNIT HAS CARBONIZED, CAUSING THE LIGHT INTENSITY OF THE GENERAL LIGHTING TO FAIL TO MEET THE STANDARD VALUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22991819 ↗
  • MalfunctionAug 29, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22922311 ↗
  • MalfunctionJul 28, 2025Intermittent Loss of Power; Intermittent Loss of Power

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22639454 ↗
  • MalfunctionJul 28, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE SUBJECT DEVICE HAD INTERMITTENT IMAGE NOISE. THE EVENT OCCURRED DURING SET UP/INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22630549 ↗
  • MalfunctionFeb 19, 2025Image Display Error/Artifact; Image Display Error/Artifact

    DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED AN ABNORMAL IMAGE FROM ALL THE OUTPUT PORTS DUE TO A PRINTED CIRCUIT BOARD FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21412949 ↗
  • MalfunctionFeb 11, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXHIBITED A E333 ERROR. THE ISSUE OCCURRED AT INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356612 ↗
  • MalfunctionDec 27, 2024Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD ERROR 333. THE EVENT OCCURRED DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21028227 ↗
  • MalfunctionDec 27, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT DURING PREPARATION BEFORE USE THE SUBJECT DEVICE DISPLAYED ABNORMAL NOISE ON THE SCREEN, AND IT WAS UNABLE TO WORK NORMALLY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21026271 ↗
  • MalfunctionNov 15, 2024No Device Output; No Device Output

    IT WAS REPORTED, THE VISERA ELITE II VIDEO SYSTEM CENTER TOUCH PANEL WAS NOT FUNCTIONING. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694185 ↗
  • MalfunctionSep 19, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE VIDEO SYSTEM CENTER EXHIBITED AN INTERMITTENT DISPLAY ISSUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20264112 ↗
  • MalfunctionSep 18, 2024No Display/Image; Output Problem; No Display/Image; Output Problem

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD NO IMAGE AND THE SCREEN DID NOT LOAD UP. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20252458 ↗
  • MalfunctionSep 17, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD ERROR E333. THE ISSUE OCCURRED DURING PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20242621 ↗
  • MalfunctionAug 22, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THAT THE TOUCH DISPLAY WAS NOT WORKING FOR THE VIDEO SYSTEM CENTER. THE ISSUE OCCURRED, DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20030588 ↗
  • MalfunctionAug 12, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER HAD A LIGHT GUIDE CONNECTION DETECTION FUNCTION FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19962676 ↗
  • MalfunctionAug 5, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THE VIDEO SYSTEM CENTER¿S SCREEN WENT BLACK DURING AN UNSPECIFIED PROCEDURE AND DISPLAYED AN E226 SCOPE COMMUNICATION ERROR. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF DELAY DURING THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19914274 ↗
  • MalfunctionJul 18, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER HAD NO IMAGE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19770419 ↗
  • MalfunctionJun 12, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD ERROR E105. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19519340 ↗
  • MalfunctionJun 4, 2024Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER SOCKETS ARE DAMAGING THE PADDLE OF THE SCOPES AND CAMERA HEAD. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19464384 ↗
  • MalfunctionJun 1, 2024Failure to Power Up; Failure to Clean Adequately; Failure to Power Up; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE VIDEO SYSTEM CENTER EXHIBITED THAT IT WAS UNABLE TO POWER UP THE UNIT DUE TO DEFECTIVE POWER SUPPLY UNIT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19442744 ↗
  • MalfunctionMay 10, 2024Output Problem; Output Problem

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER HAD AN ERROR CODE E333 (TOUCH PANEL ERROR) DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19288223 ↗
  • MalfunctionMay 8, 2024Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THAT THE VIDEO SYSTEM CENTER HAD A BLURRY IMAGE AND EXPERIENCED AN E227 SCOPE COMMUNICATION ERROR. THE ISSUE WAS FOUND DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19271944 ↗
  • MalfunctionApr 29, 2024Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD E333 ERROR. THE ISSUE OCCURRED DURING PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19201431 ↗
  • MalfunctionApr 23, 2024Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THE VIDEO SYSTEM CENTER EXHIBITED AN IMAGE PROCESSOR ERROR (E230ERROR). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19169058 ↗

Other Endoscopic Video Imaging System/Component, Gastroenterology-Urology devices

All Endoscopic Video Imaging System/Component, Gastroenterology-Urology devices →