Single Use 3-Lumen Sphincterotome V KD-V411M-0730 (UDI 04953170184055)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use 3-Lumen Sphincterotome V · Class 2
All models of Single Use 3-Lumen Sphincterotome V (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus single-use sphincterotome is a disposable endoscopic cutting tool used to open ducts during ERCP procedures. Most of the 65 reports describe the cutting wire snapping, breaking or tearing, along with the plastic coating peeling or splitting; a few mention the tip melting or twisting. Reports also describe perforation of the duct wall, pancreatitis, and one case of electric shock. There are no deaths, 6 injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 64
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04953170184055), received between Jun 12, 2020 and Aug 18, 2026 — classified as 6 injuries, 58 malfunctions.
The most frequently reported specific device problem is “Break” (43 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break43
- Peeled/Delaminated14
- Material Split, Cut or Torn10
- Melted4
- Material Separation2
- Material Twisted/Bent2
- Thermal Decomposition of Device2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions56
- Perforation4
- No Consequences Or Impact To Patient3
- Pancreatitis2
- Electric Shock1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50305533 ↗MalfunctionJan 27, 2026Break
IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME KNIFE WIRE WAS FRACTURED. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24190875 ↗MalfunctionDec 23, 2025Break; Break
IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME HAD A FRACTURED KNIFE WIRE. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23881517 ↗InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE AUTHOR.
Patient problems: Pancreatitis
MAUDE report 23720510 ↗MalfunctionNov 3, 2025Break; Break
IT WAS REPORTED THAT THE SUBJECT DEVICE FAILED TO BOW OR CUT, AND THE WIRE EVENTUALLY SNAPPED. THE ISSUE OCCURRED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23459032 ↗MalfunctionJul 21, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V EXHIBITED WIRE BROKEN AND THE PROXIMAL END AND DAMAGED COATED PORTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22553826 ↗MalfunctionJul 3, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COATED PORTION OF THE CUTTING WIRE ON THE SINGLE-USE 2-LUMEN SPHINCTEROTOME WAS TORN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22395254 ↗MalfunctionMar 3, 2025Break; Break
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21505672 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21339223 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21338821 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21338732 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21338652 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21338239 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21338220 ↗MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21337769 ↗MalfunctionDec 10, 2024Break; Break
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20894907 ↗MalfunctionDec 5, 2024Break; Break
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20854239 ↗MalfunctionDec 5, 2024Break; Break
IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20854234 ↗MalfunctionDec 5, 2024Break; Break
IT WAS REPORTED THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20853995 ↗MalfunctionDec 5, 2024Break; Break
IT WAS REPORTED THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20853843 ↗MalfunctionOct 1, 2024Break; Break
IT WAS REPORTED THE LUMEN SPHINCTEROTOME HAD THE KNIFE WIRE BREAK. THE ISSUE OCCURRED DURING AN UNKNOWN THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20348672 ↗MalfunctionSep 17, 2024Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted
IT WAS REPORTED THAT THE WIRE OF THE SINGLE-USE 2-LUMEN SPHINCTEROTOME WAS DAMAGED. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE AND WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20242654 ↗MalfunctionSep 11, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME EXHIBITED COVER PEELING OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20196188 ↗MalfunctionAug 6, 2024Break; Thermal Decomposition of Device; Break; Thermal Decomposition of Device
IT WAS REPORTED THAT THE 3-LUMEN SPHINCTEROTOME V EXHIBITED A KNIFE WIRE THAT WAS DAMAGED. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917343 ↗MalfunctionJul 5, 2024Break; Break
IT WAS REPORTED, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V'S CUTTING WIRE BROKE. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM AND THE INTENDED PROCEDURE WAS ABLE TO BE SUCCESSFULLY COMPLETED USING A SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19680952 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page