Single Use 3-Lumen Sphincterotome V KD-V411M-0730 (UDI 04953170184055)Recalls & reported problems

Olympus Medical Systems Corp. · Single Use 3-Lumen Sphincterotome V · Class 2

All models of Single Use 3-Lumen Sphincterotome V (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus single-use sphincterotome is a disposable endoscopic cutting tool used to open ducts during ERCP procedures. Most of the 65 reports describe the cutting wire snapping, breaking or tearing, along with the plastic coating peeling or splitting; a few mention the tip melting or twisting. Reports also describe perforation of the duct wall, pancreatitis, and one case of electric shock. There are no deaths, 6 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04953170184055), received between Jun 12, 2020 and Aug 18, 2026 — classified as 6 injuries, 58 malfunctions.

The most frequently reported specific device problem is “Break” (43 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 5 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 5 reportsJan 2025: 0 reportsFeb 2025: 7 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break43
  • Peeled/Delaminated14
  • Material Split, Cut or Torn10
  • Melted4
  • Material Separation2
  • Material Twisted/Bent2
  • Thermal Decomposition of Device2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions56
  • Perforation4
  • No Consequences Or Impact To Patient3
  • Pancreatitis2
  • Electric Shock1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50305533 ↗
  • MalfunctionJan 27, 2026Break

    IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME KNIFE WIRE WAS FRACTURED. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190875 ↗
  • MalfunctionDec 23, 2025Break; Break

    IT WAS REPORTED THE SINGLE USE 2-LUMEN SPHINCTEROTOME HAD A FRACTURED KNIFE WIRE. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE (DURING SET UP IN ROOM / BEFORE PATIENT IN ROOM). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23881517 ↗
  • InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM THE AUTHOR.

    Patient problems: Pancreatitis

    MAUDE report 23720510 ↗
  • MalfunctionNov 3, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE FAILED TO BOW OR CUT, AND THE WIRE EVENTUALLY SNAPPED. THE ISSUE OCCURRED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23459032 ↗
  • MalfunctionJul 21, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V EXHIBITED WIRE BROKEN AND THE PROXIMAL END AND DAMAGED COATED PORTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22553826 ↗
  • MalfunctionJul 3, 2025Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COATED PORTION OF THE CUTTING WIRE ON THE SINGLE-USE 2-LUMEN SPHINCTEROTOME WAS TORN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22395254 ↗
  • MalfunctionMar 3, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21505672 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21339223 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338821 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338732 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338652 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338239 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338220 ↗
  • MalfunctionFeb 7, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21337769 ↗
  • MalfunctionDec 10, 2024Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20894907 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854239 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT WHILE USING THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20854234 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20853995 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME V THE CUTTING WIRE OF THE PAPILLOTOME BROKE OR WAS DAMAGED WHEN CUTTING UNDER TENSION. THE REPORTED MALFUNCTION OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20853843 ↗
  • MalfunctionOct 1, 2024Break; Break

    IT WAS REPORTED THE LUMEN SPHINCTEROTOME HAD THE KNIFE WIRE BREAK. THE ISSUE OCCURRED DURING AN UNKNOWN THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20348672 ↗
  • MalfunctionSep 17, 2024Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted

    IT WAS REPORTED THAT THE WIRE OF THE SINGLE-USE 2-LUMEN SPHINCTEROTOME WAS DAMAGED. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE AND WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20242654 ↗
  • MalfunctionSep 11, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE SINGLE USE 3-LUMEN SPHINCTEROTOME EXHIBITED COVER PEELING OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20196188 ↗
  • MalfunctionAug 6, 2024Break; Thermal Decomposition of Device; Break; Thermal Decomposition of Device

    IT WAS REPORTED THAT THE 3-LUMEN SPHINCTEROTOME V EXHIBITED A KNIFE WIRE THAT WAS DAMAGED. IT IS UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917343 ↗
  • MalfunctionJul 5, 2024Break; Break

    IT WAS REPORTED, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V'S CUTTING WIRE BROKE. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM AND THE INTENDED PROCEDURE WAS ABLE TO BE SUCCESSFULLY COMPLETED USING A SIMILAR DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19680952 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →