Single Use 3-Lumen Sphincterotome V KD-V411M-3030 (UDI 04953170184154)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use 3-Lumen Sphincterotome V (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170184154), received between Aug 25, 2022 and Jan 29, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2026Class IIOpen, Classified
Devices which did not undergo thermoforming could deform and lose performance.
Z-1292-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Material Split, Cut or Torn2
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
THE CUSTOMER REPORTED THE KNIFE OF THE SINGLE USE 3-LUMEN SPHINCTEROTOME WAS TURNED IN THE OPPOSITE DIRECTION. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE COATING WAS TORN. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18593965 ↗MalfunctionMay 18, 2023Break; Melted; Material Split, Cut or Torn; Break; Melted; Material Split, Cut or Torn
THE CUSTOMER REPORTED TO OLYMPUS THAT AFTER THE ENDOSCOPIC SPHINCTEROTOMY PROCEDURE BEGAN A SPARK OCCURRED AND THE SINGLE USE 3-LUMEN SPHINCTEROTOME V BROKE. THE PROCEDURE WAS COMPLETED AFTER THE DEVICE WAS REPLACED WITH A SIMILAR DEVICE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16954055 ↗MalfunctionAug 25, 2022Break; Break
IT WAS REPORTED THAT DURING A THERAPEUTIC SPHINCTEROTOMY THE KNIFE WIRE BROKE WHILE IT WAS ACTIVATED BUT NO DEVICE FRAGMENT FALL. THE PROCEDURE WAS COMPLETED USING A DIFFERENT BUT SIMILAR DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15298465 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page