Single Use 3-Lumen Sphincterotome V KD-V431M-0720 (REF N1090430)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use 3-Lumen Sphincterotome V · Class 2
All models of Single Use 3-Lumen Sphincterotome V (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170184178), received between May 18, 2022 and Dec 20, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 7, 2026Class IIOpen, Classified
Devices which did not undergo thermoforming could deform and lose performance.
Z-1275-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 20, 2023Break; Break
THE CUSTOMER REPORTED THE SPHINCTEROTOME WAS IN USE FOR A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY. THE DEVICE'S KNIFE WIRE BROKE WHEN ELECTRICITY WAS APPLIED. THE PROCEDURE WAS COMPLETED BY REPLACING THE DEVICE WITH ANOTHER ONE. NO ADVERSE EFFECTS TO PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18367592 ↗MalfunctionMar 28, 2023Break; Break
AN OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT, DURING A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY (EST), THE TIP KNIFE WIRE OF THEIR SINGLE USE 3-LUMEN SPHINCTEROTOME V BROKE, CAUSING SPARKS TO OCCUR. NO DEBRIS FELL OFF, AND THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16633010 ↗MalfunctionMay 18, 2022Break; Break
THE CUSTOMER REPORTED THE SINGLE USE 3-LUMEN SPHINCTEROTOME V WIRE BROKE WHEN THE PAPILLA WAS INCISED. THE EVENT OCCURRED DURING A THERAPEUTIC ENDOSCOPIC SPHINCTEROTOMY PROCEDURE AND A SIMILAR DEVICE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14427695 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page