Skip to content

CONNECTING TUBE MAJ-1513Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CONNECTING TUBE (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170234026), received between Sep 11, 2023 and Dec 17, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Break1
  • Device Reprocessing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THE CONNECTING TUBE FELL OFF DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23835323 ↗
  • InjurySep 16, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DIAGNOSTIC PROCEDURE, A RUBBER -LIKE OBJECT ATTACHED TO THE FORCEPS PORT OF THE CONNECTING TUBE HAD FALLEN OFF INTO THE PATIENT. THE RUBBER FROM THE CLEANING TUBE HAS BEEN RETRIEVED. IT IS BELIEVED THAT THE RUBBER BROKE OFF WHEN CLEANING WITH A CLEANING DEVICE MADE BY ANOTHER COMPANY. THE RUBBER BROKE OFF AND ENTERED THE ENDOSCOPE, CAUSING IT TO FALL OFF DURING THE EXAMINATION. THE FRAGMENTS WERE RETRIEVED USING THE SAME DEVICE, AND THE PROCEDURE WAS COMPLETED (WITH THE SAME DEVICE). NO REPORTS OF HEALTH HAZARD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20225004 ↗
  • MalfunctionSep 11, 2023Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE EVIS EXERA III BRONCHOVIDEOSCOPE WAS ATTEMPTED TO BE REPROCESSED ON THE OER-PRO ENDOSCOPE REPROCESSOR, BUT THE MAJ-1513/CONNECTING TUBE WOULD NOT STAY CONNECTED TO THE SCOPE. IN SPEAKING WITH OLYMPUS TECHNICAL ASSISTANCE SUPPORT (TAC) VIA PHONE, THE CUSTOMER NOTED THAT WHEN THE REPROCESSING SEQUENCE WAS COMPLETED, THE CONNECTOR HAD COME OFF THE SCOPE. THE CONNECTING TUBE COULD NOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER ONLY HAD ONE BF-P190 SCOPE AND ONE MAJ-1513 CONNECTING TUBE. THE PRODUCT WILL NOT BE RETURNED. THERE WERE NO REPORTS OF PATIENT HARM. RELATED PATIENT IDENTIFIER: (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17720728 ↗

Other ACCESSORIES, CLEANING, FOR ENDOSCOPE devices

All ACCESSORIES, CLEANING, FOR ENDOSCOPE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page