Single Use 3-Lumen Needle Knife V KD-V451M (UDI 04953170244148)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use 3-Lumen Needle Knife V · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170244148), received between Mar 28, 2022 and Sep 22, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Material Split, Cut or Torn1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Abdominal Pain1
- Fever1
- Hemorrhage/Blood Loss/Bleeding1
- Pneumonia1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Peeled/Delaminated
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50854562 ↗MalfunctionMar 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE COATED PORTION OF THE CUTTING KNIFE WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21660098 ↗MalfunctionJan 24, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
THE CUSTOMER REPORTED THE KNIFE OF THE SINGLE USE 3-LUMEN NEEDLE KNIFE PIERCED THE SHEATH AND PROTRUDED. THE ISSUE OCCURRED WHEN THE DEVICE WAS INSERTED INTO A SCOPE AND THE PHYSICIAN TRIED TO PROJECT THE INCISION KNIFE AT THE TIP. A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE WIRE SHEATHING HAD TURNED OVER TOWARD THE PROXIMAL SIDE. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18569545 ↗InjuryMar 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED: "CLINICOPATHOLOGICAL CHARACTERISTICS OF PATIENTS WITH PAIN AFTER ENDOSCOPIC SUBMUCOSAL DISSECTION FOR GASTRIC EPITHELIAL NEOPLASM" BY JUNG-WOOK KIM ET AL. LITERATURE SUMMARY: THIS RETROSPECTIVE STUDY WAS AIMED TO EXPLORE, BOTH THE INCIDENCE OF EPIGASTRIC PAIN AFTER GASTRIC ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) AND CLINICOPATHOLOGIC FACTORS ASSOCIATED WITH THE PAIN IN A LARGE NUMBER OF PATIENTS. THE STUDY RETROSPECTIVELY ANALYZED A PROSPECTIVELY COLLECTED REGISTRY OF CLINICAL, ENDOSCOPIC, AND PATHOLOGIC RESULTS OF 1266 PATIENTS WHO UNDERWENT ESD FOR GASTRIC ADENOMA OR CANCER. THE DATA WAS ANALYZED INCLUDED AGE, SEX, PATHOLOGY, SPECIMEN AND TUMOR SIZE, PROCEDURE TIME, AND TUMOR LOCATION. A VISUAL ANALOGUE SCALE (VAS) WAS USED TO ASSESS PAIN IMMEDIATELY AFTER, AND 2, 12, AND 24 H AFTER ESD WHILE THE PRIMARY OUTCOME WAS THE USE OF PAINKILLERS (VAS SCORE > 4). OF 1226 PATIENTS, 461 (36.4%) NEEDED A PAINKILLER AT LEAST ONCE AFTER ESD, WERE CONSIDERED AS THE PAIN GROUP. THE STUDY CONCLUDED THAT THE FACTORS INFLICTING PAIN AFTER ESD WERE INCLUDED FEMALE SEX, PROCEDURE TIME, AND LESION LOCATION. ALSO SUGGESTED THAT THE ENDOSCOPISTS SHOU
Patient problems: Abdominal Pain; Fever; Hemorrhage/Blood Loss/Bleeding; Pneumonia
MAUDE report 13929560 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page