THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip TB-0520FCRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04953170370519), received between Oct 5, 2016 and Jul 15, 2024 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 22, 2025Class IIOpen, Classified
Firm is initiating a removal due to continued reports of adverse events.
Z-1376-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break5
- Material Separation3
- Fracture2
- Detachment of Device or Device Component1
- Failure to Deliver Energy1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJul 15, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SUBJECT DEVICE FELL INTO THE PATIENT'S ABDOMINAL CAVITY AND WAS RETRIEVED WITH FORCEPS. THE PATIENT WAS NOT IMPACTED AS THE FALLING PART WAS REMOVED IMMEDIATELY AND IT TOOK 1-MINUTE TO RETRIEVE THE BROKEN SUBJECT DEVICE. THE PATIENT IS IN GOOD HEALTH AND HAS BEEN DISCHARGED FROM THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19746169 ↗MalfunctionJan 31, 2024Material Separation; Material Separation
IT WAS REPORTED, THE THUNDERBEAT PAD WAS BADLY WORN, THE FRONT END WAS COCKED AND COULD NOT BE USED NORMALLY. THERE WAS NO PART OF THE DEVICE THAT FELL OR FELL INTO THE PATIENT¿S BODY. THE THERAPEUTIC TOTAL HYSTERECTOMY WAS COMPLETED USING A REPLACEMENT DEVICE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18619098 ↗MalfunctionAug 29, 2023Material Separation; Material Separation
THE CUSTOMER REPORTED THAT THE TEFLON PAD ON THE ULTRASOUND KNIFE CAME OFF WITHIN 10 MINUTES DURING A THERAPEUTIC RIGHT LIVER RESECTION PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING ANOTHER DEVICE WITH A DELAY OF 15 MINUTES. THERE WAS NO REPORT OF PATIENT OR USER INJURY DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17646350 ↗MalfunctionMay 19, 2023Fracture; Fracture
A CUSTOMER REPORTED TO OLYMPUS, DURING A THERAPEUTIC PAROTID GLAND OPERATION THE THUNDERBEAT¿S NEEDLE CORE OF THE KNIFE HEAD WAS BROKEN AND INOPERABLE. THE PROCEDURE WAS DELAYED NO MORE THAN 15 MINUTES AND COMPLETED USING A REPLACEMENT DEVICE. THE SUBJECT DEVICE BROKE OUTSIDE OF THE PATIENT¿S BODY CAVITY. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16966777 ↗MalfunctionMay 18, 2023Fracture; Fracture
A CUSTOMER REPORTED TO OLYMPUS, DURING A THERAPEUTIC PAROTID GLAND SURGERY THE THUNDERBEAT¿S NEEDLE CORE OF THE KNIFE HEAD WAS BROKEN AND INOPERABLE. THE EVENT OCCURRED AT THE BEGINNING OF THE PROCEDURE. THE PROCEDURE WAS DELAYED NO MORE THAN 15 MINUTES AND COMPLETED USING A REPLACEMENT DEVICE. THE SUBJECT DEVICE BROKE OUTSIDE OF THE PATIENT¿S BODY. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16953913 ↗MalfunctionJan 28, 2022Material Separation; Material Separation
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, THIRTY MINUTES INTO A THERAPEUTIC OPEN RADICAL TREATMENT FOR ENDOMETRIAL CANCER PROCEDURE, U508 ALARMS FOR SHORT CIRCUIT WERE RECEIVED ON THE PROCESSOR FOR THE DEVICE IN THE FREE ILIAC ARTERY. THE DEVICE TEFLON PAD WAS WORN, SEPARATED, AND FELL OFF. THE SEPARATED PIECE WAS RETRIEVED FROM THE PATIENT. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13385346 ↗MalfunctionFeb 6, 2020Break; Break
DURING A PANCREATECTOMY WITH SPLENECTOMY, THE SUBJECT DEVICE WAS USED. AFTER 2 HOURS SINCE THE SURGERY BEGAN, THE PROBE OF THE SUBJECT DEVICE BROKE OFF OUTSIDE THE PATIENT. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9675114 ↗MalfunctionSep 25, 2019Break; Break
DURING AN UNSPECIFIED LAPAROSCOPIC PROCEDURE, THE SUBJECT DEVICE WAS USED. AFTER ONE ACTIVATION, THE TISSUE PAD OF THE SUBJECT DEVICE WAS DAMAGED AND SEPARATED. NO FURTHER INFORMATION WAS PROVIDED. ON SEPTEMBER 24, 2019, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE TISSUE PAD OF THE SUBJECT DEVICE WAS PARTIALLY SEPARATED. THIS IS THE REPORT REGARDING THE SEPARATION OF THE TISSUE PAD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9114750 ↗MalfunctionAug 2, 2018Failure to Deliver Energy; Failure to Deliver Energy
DURING AN OPEN SPLENECTOMY, THE SUBJECT DEVICE WAS USED. THE DEVICE DID NOT SEAL THE VESSEL. THERE WAS NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7741809 ↗MalfunctionOct 5, 2016Break; Break
THE SUBJECT DEVICE WAS USED DURING A RESECTION OF KIDNEY TUMOR. THE PTFE PAD OF THE DEVICE WAS PARTIALLY SEPARATED. THERE WAS NO PATIENT HARM REPORTED. NO FURTHER INFORMATION WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6002060 ↗MalfunctionOct 5, 2016Break; Break
THE SUBJECT DEVICE WAS USED DURING A NEPHRECTOMY. AFTER SHORT TIME USE, THE PTFE PAD OF THE DEVICE WAS PARTIALLY SEPARATED. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6002058 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page