Single Use Sphincterotome V(Distal Wireguided) KD-VC431Q-0720 (REF N5391530)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of Single Use Sphincterotome V(Distal Wireguided) (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170371172), received between Aug 10, 2022 and May 9, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 9, 2024Break; Break

    IT WAS REPORTED THE SPHINCTEROTOME WAS PRELOADED AND THE CUTTING/CAUTER SNAPPED WHEN INITIATED. THE ISSUE OCCURRED DURING A DIAGNOSTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THE PATIENT WAS SEDATED AND THERE WAS A 5-MINUTE DELAY IN THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19281667 ↗
  • MalfunctionMay 7, 2024Break; Break

    IT WAS REPORTED THE SPHINCTEROTOME EXPERIENCED SPLITTING OF THE CUTTING/CAUTER GUIDEWIRE. THE ISSUE OCCURRED DURING A DIAGNOSTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE. THE PROCEDURE WAS COMPLETED USING ANOTHER SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19264980 ↗
  • MalfunctionFeb 27, 2024Break; Break

    IT WAS REPORTED, THE IT WAS REPORTED, THE SINGLE USE SPHINCTEROTOME V'S CUTTING WIRE BROKE. THE ISSUE OCCURRED AT THE BEGINNING OF AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18789976 ↗
  • MalfunctionAug 10, 2022Break; Material Split, Cut or Torn; Break; Material Split, Cut or Torn

    THE CUSTOMER REPORTED THE CUTTING WIRE ON THE SUBJECT DEVICE BROKE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE. THE CUTTING WIRE WAS DAMAGED WITH NO SPARKS OBSERVED AND WAS STILL IN ONE PIECE. THE PROCEDURE WAS COMPLETED USING ANOTHER SIMILAR DEVICE. THE SUBJECT DEVICE WAS SENT TO OLYMPUS FOR EVALUATION. DURING INSPECTION AND TESTING, THE CUTTING WIRE WAS FOUND TO BE DAMAGED AND THE COATING ON THE WIRE WAS DAMAGED. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (DAMAGE TO CUTTING WIRE). THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15198227 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →