Single Use Sphincterotome V (Distal Wireguided) KD-VC431Q-0730Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use Sphincterotome V(Distal Wireguided) (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170371196), received between Mar 29, 2019 and Feb 16, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
AN OLYMPUS SALES REPRESENTATIVE (REP) REPORTED THAT HIS CUSTOMER EXPERIENCED AN OUT OF BOX FAILURE ON HIS OLYMPUS SINGLE-USE SPHINCTEROTOME V. ACCORDING TO THE REP, THE KNIFE WIRE FELL OFF UPON REMOVAL FROM THE PACKAGING. THIS EVENT HAD NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16385142 ↗MalfunctionMar 29, 2019Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT AT THE BEGINNING OF AN ERCP PROCEDURE DURING THE SPHINCTEROTOMY, THE WIRE ON THE CLEVERCUT SPHINCTEROTOME DISLODGED COMPLETELY FROM THE TOME. THE INTENDED PROCEDURE WAS COMPLETED WITH A NEW TOME. A 2ND DISPOSABLE GRASPING DEVICE, A RAPTOR, HAD TO BE OPENED AND USED TO RETRIEVE THE FREE FLOATING WIRE FROM THE BOWEL. A NEW CLEVERCUT WAS OPENED AND USED TO COMPLETE THE PROCEDURE. THE WIRE FRAGMENT WAS RETRIEVED USING A DISPOSABLE RAPTOR. THE PATIENT DID NOT REQUIRE A LONGER STAY. A 1-5 MINUTES DELAY OCCURRED IN THE PROCEDURE DUE TO THE TOME AND WIRE NEEDING TO BE WITHDRAWN AND THE RAPTOR BEING INTRODUCED TO RETRIEVE THE BROKEN WIRE. THE DEVICE WAS INSPECTED PRIOR TO USE AND THERE WERE NO ANOMALIES OBSERVED. THERE WERE NO ISSUES INSERTING THE REPORTED DEVICE INTO THE PATIENT. THE MODEL/SERIAL OF THE SCOPE USED DURING THE PROCEDURE WAS AN OLYMPUS ERCP SCOPE. THERE WERE NO REPORTED ISSUES WITH THE SCOPE AND IT DID NOT NEED TO BE REPLACED DURING THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8463316 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page