Single Use Preloaded Sphincterotome V (Distal Wireguided) KD-VC611Q-07301S (REF N5392330)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use Preloaded Sphincterotome V (Distal Wireguided) (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170371332), received between Oct 29, 2020 and Apr 5, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Material Split, Cut or Torn1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2024Break; Break
IT WAS REPORTED THE SINGLE USE PRELOADED SPINCTEROTOME EXPERIENCED BREAKAGE OF THE CUTTING WIRE. THE ISSUE OCCURRED DURING A THERAPEUTIC SPHINCTEROTOMY PROCEDURE. THE THERAPEUTIC PROCEDURE WAS COMPLETED WITH NO DELAY OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19048253 ↗MalfunctionMar 19, 2024Break; Break
IT WAS REPORTED THE SINGLE USE PRELOADED SPHINCTEROTOME CUTTING WIRE BROKE. THE ISSUE OCCURRED DURING A SPHINCTEROTOMY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18935638 ↗MalfunctionMar 18, 2024Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THAT THE SUBJECT DEVICE¿S CUTTING WIRE BROKE. THE ISSUE OCCURRED DURING A THERAPEUTIC SPHINCTEROTOMY. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE AND THERE WERE NO REPORTS OF DELAYS OR PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18924062 ↗MalfunctionFeb 15, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
THE CUSTOMER REPORTED THE CUTTING WIRE PROTECTION OF THE SINGLE USE PRELOADED SPHINCTEROTOME WAS ON THE DISTAL PART OF THE DEVICE AND THE THERAPEUTIC SPHINCTEROTOMY WAS UNABLE TO BE PERFORMED. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18710522 ↗MalfunctionMar 3, 2022Break; Break
BREAKAGE OF THE CUTTING WIRE OF THE SPHINCTEROTOME DURING A DIGESTIVE ENDOSCOPY PROCEDURE WAS REPORTED. ACCORDING TO THE REPORT, THIS INCIDENT HAS OCCURRED REPEATEDLY WITH DIFFERENT LOT NUMBERS (LOT NUMBER 14V AND 15V). INCREASED IN PROCEDURE (UNKNOWN TIME) WAS REPORTED DUE TO DEVICE CHANGE HOWEVER, THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. THIS REPORT IS FOR LOT 15V. THIS REPORT IS RELATED TO A REPORT WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13661112 ↗MalfunctionMar 3, 2022Break; Break
BREAKAGE OF THE CUTTING WIRE OF THE SPHINCTEROTOME DURING A DIGESTIVE ENDOSCOPY PROCEDURE WAS REPORTED. ACCORDING TO THE REPORT, THIS INCIDENT HAS OCCURRED REPEATEDLY WITH DIFFERENT LOT NUMBERS (LOT NUMBER 14V AND 15V). INCREASED PROCEDURE TIME (UNKNOWN TIME) WAS REPORTED DUE TO DEVICE CHANGE HOWEVER, THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. THIS REPORT IS FOR LOT 14V. THIS REPORT IS RELATED TO A REPORT WITH PATIENT (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13660953 ↗MalfunctionMay 12, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY THE HEALTH PROFESSIONAL THAT DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) USING THE SUBJECT DEVICE, THE CUTTING WIRE WAS BROKEN. THERE WERE NO MISSING PARTS. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11810531 ↗MalfunctionOct 29, 2020Break; Break
WE RECEIVED THE FOLLOWING REPORT. IT WAS IMPOSSIBLE TO SLIDE THE SPHINCTEROTOME (THE SUBJECT DEVICE) OVER THE GUIDEWIRE. THERE WAS NO PATIENT INJURY REPORTED. NO FURTHER INFORMATION WAS PROVIDED. THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. ON AUGUST 31, 2020, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE CUTTING WIRE WAS SNAPPED, AND THAT THE BROKEN SECTION OF THE CUTTING WIRE WAS MELTED AND BURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10753446 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page