Single Use Preloaded Sphincterotome V(Distal Wireguided) KD-VC631Q-07203S (REF N5394330)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use Preloaded Sphincterotome V (Distal Wireguided) (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170371738), received between Dec 22, 2021 and May 3, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Flare or Flash2
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 3, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) AT THE TIME OF THE SPHINCTEROTOMY DURING BOWING, THE WIRE OF THE SINGLE USE PRELOADED SPHINCTEROTOME BROKE AT THE AMPULLA INSIDE THE PATIENT IN THE SCOPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19244241 ↗MalfunctionOct 18, 2022Break; Flare or Flash; Break; Flare or Flash
DEVICE WAS REPORTED TO FAIL UPON INITIAL CLINICAL USE. OLYMPUS AREA SALESPERSON CUSTOMER REPRESENTATIVE (REP.) REPORTED THE CUTTING WIRE OF THE KD-VC631Q-07203S HAD BROKEN AND BURN IN HALF AT THE DISTAL DURING A THERAPEUTIC UNSPECIFIED PROCEDURE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15621892 ↗MalfunctionOct 18, 2022Break; Flare or Flash; Break; Flare or Flash
DEVICE WAS REPORTED TO FAIL UPON INITIAL CLINICAL USE. OLYMPUS AREA SALESPERSON CUSTOMER REPRESENTATIVE (REP.) REPORTED THE CUTTING WIRE OF THE KD-VC631Q-07203S HAD BROKEN AND BURN IN HALF AT THE DISTAL DURING A PROCEDURE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15621881 ↗MalfunctionDec 22, 2021Break; Break
THE CUSTOMER REPORTED THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE, WHILE INSIDE THE PATIENT THE SPHINCTEROTOME WAS ACTIVATED AND THE GUIDEWIRE BROKE AT THE PROXIMAL END. THE WIRE REMAINED ATTACHED AT THE DISTAL END; HOWEVER, COULD NOT BE RETRACTED THROUGH THE UNIT. THE WIRE FORMED A HOOK SHAPE WHICH CONCERNED THE DOCTOR. THE SPHINCTEROTOME WAS SUCCESSFULLY EXTRACTED FROM THE PATIENT BY PULLING THE WIRE INTO THE SCOPE AND WITHDRAWING THE SCOPE FROM THE PATIENT. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13059774 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page