Skip to content

Single Use Electrosurgical Knife KD-655 KD-655L (UDI 04953170377488)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of Single Use Electrosurgical Knife KD-655 (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04953170377488), received between Dec 15, 2023 and Dec 15, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Fire” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fire1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 15, 2023Fire; Fire

    THE CUSTOMER REPORTED THAT WHEN USING AN EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE, THERE WAS FIRE AT THE PLASTIC DISTAL END COVER CAUSING BURNS TO THE PATIENT¿S STOMACH. THE EVENT OCCURRED DURING THE PRE-LABELING PHASE OF A THERAPEUTIC PROCEDURE (ESD) AND THE C02 AIR PUMP WAS USED TO SUPPLY THE GAS AT THE TIME OF THE INCIDENT. THERE WAS NO PROCEDURAL DELAY, OR NO FURTHER PATIENT HARM REPORTED WITH THE EVENT. THE PROCEDURE WAS FINISHED WITH ANOTHER SET OF DEVICE. THE PATIENT WAS HOSPITALIZED FOR FURTHER TREATMENT. ADDITIONAL INFORMATION REGARDING THE EVENT HAS BEEN REQUESTED BUT NOT RECEIVED AT THIS TIME. THIS EVENT REQUIRES THREE REPORTS, AS ALL DEVICES WERE USED DURING THE EVENT: PATIENT IDENTIFIER (B)(6) IS FOR THE VIDEOSCOPE. PATIENT IDENTIFIER (B)(6) IS FOR THE ELECTROSURGICAL KNIFE. PATIENT IDENTIFIER C23555710 IS FOR THE HF UNIT

    Patient problems: Burn(s)

    MAUDE report 18336057 ↗

Other Unit, Electrosurgical, Endoscopic (With Or Without Accessories) devices

All Unit, Electrosurgical, Endoscopic (With Or Without Accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page