Single Use 3-Lumen Sphincterotome V KD-V411M-3030 (UDI 04953170382673)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use 3-Lumen Sphincterotome V (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04953170382673), received between Oct 4, 2021 and May 10, 2024 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Poor Quality Image13
- Break4
- Material Split, Cut or Torn1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 10, 2024Break; Material Split, Cut or Torn; Break; Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SPHINCTEROTOME EXHIBITED THE KNIFE WIRE WAS BROKEN AND THE WIRE COATING WAS TORN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19291609 ↗MalfunctionApr 30, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE LAPARO-THORACO VIDEOSCOPE HAD POOR QUALITY IMAGE. THE ISSUE OCCURRED DURING A PROCEDURE. THE CUSTOMER NOTED THERE MAY BE A POSSIBLE INTERNAL ISSUE AND THERE WAS CONDENSATION FORMING ON THE INSIDE OF THE TIP OF THE SCOPE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE SCOPE IMAGE FAILED A FOG TEST. THE REPORT IS BEING SUBMITTED DUE THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19211959 ↗MalfunctionApr 30, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE LAPARO-THORACO VIDEOSCOPE HAD POOR IMAGE QUALITY. THE ISSUE OCCURRED DURING A PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS VERY CLOUDY. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19211917 ↗MalfunctionApr 25, 2024Poor Quality Image; Peeled/Delaminated; Poor Quality Image; Peeled/Delaminated
THE CUSTOMER REPORTED THE LAPARO-THORACO VIDEOSCOPE HAD NO ANGULATION WHEN THE CONTROL LEVER WAS TURNED. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY AND THE LIGHT GUIDE LENS ADHESIVE WAS PEELED OFF. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECTS FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19183378 ↗MalfunctionApr 25, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS GRAINY. THE ISSUE OCCURRED WHEN PREPARING THE SCOPE FOR USE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19183260 ↗MalfunctionApr 25, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS CLOUDY. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS VERY CLOUDY DUE TO FLUID IN THE OBJECTIVE LENS. THE REPORT IS BEING SUBMITTED DUE TO THE FAULT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19183245 ↗MalfunctionApr 10, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE TIP OF THE LAPARO-THORACO VIDEOSCOPE WAS LEAKING. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY DUE TO A DAMAGED OBJECTIVE LENS AND CHARGED COUPLED DEVICE (CCD) UNIT. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19079008 ↗MalfunctionApr 10, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS BLURRY AND THERE WAS AN AIR LEAK. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY DUE TO A DAMAGED OBJECTIVE LENS. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19078928 ↗MalfunctionApr 10, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE BENDING SECTION COVER OF THE LAPARO-THORACO VIDEOSCOPE HAD A HOLE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY DUE TO A DAMAGED OBJECTIVE LENS AND CHARGED COUPLED DEVICE (CCD) UNIT. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19078741 ↗MalfunctionApr 9, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS BAD. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY DUE TO A DAMAGED OBJECTIVE LENS. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19069873 ↗MalfunctionMar 18, 2024Poor Quality Image; Poor Quality Image
IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS FOGGY. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18923722 ↗MalfunctionMar 18, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS BLURRY. THE ISSUE OCCURRED DURING A PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18919641 ↗MalfunctionMar 14, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS BLURRY. THE ISSUE WAS FOUND DURING A PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY DUE TO A DAMAGED CHARGED COUPLED DEVICE (CCD) UNIT. THE REPORT IS BEING SUBMITTED DUE TO DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18901004 ↗MalfunctionMar 6, 2024Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED THE IMAGE OF THE LAPARO-THORACO VIDEOSCOPE WAS FUZZY AND ONE LIGHT WAS OUT. THE ISSUE OCCURRED DURING A PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT. DURING DEVICE EVALUATION AT OLYMPUS, IT WAS FOUND THE IMAGE WAS FOGGY. THE REPORT IS BEING SUBMITTED DUE TO THE DEFECT FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18846509 ↗MalfunctionJan 26, 2024Break; Break
IT WAS REPORTED, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V'S WIRE BROKE. THE ISSUE OCCURRED DURING A NIPPLE INCISION FOR AN ENDOSCOPIC SPHINCTEROTOMY. THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18588184 ↗MalfunctionOct 5, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY A NURSE THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY USING THE SUBJECT DEVICE (KD-V411M-0330), THE CUTTING WIRE WAS BROKEN OFF ON ACTIVATION. THE USER CHANGED FROM THE SUBJECT DEVICE FROM ANOTHER DEVICE (KD-V411M-0725). KD-V411M-0725 WAS ALSO BROKEN OFF ON ACTIVATION. THE INTENDED PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE FIRST DEVICE (KD-V411M-0330).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12580712 ↗MalfunctionOct 4, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED FROM USER FACILITY THAT THE KNIFE WIRE BROKE WHEN ENERGIZED DURING A ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12566769 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page