Single Use 3-Lumen Sphincterotome V KD-V431M-0720 (UDI 04953170382680)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use 3-Lumen Sphincterotome V (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04953170382680), received between Dec 27, 2019 and Jul 21, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
- Material Split, Cut or Torn1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2023Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn
THE CUSTOMER REPORTED TO OLYMPUS, THE SINGLE USE 3-LUMEN SPHINCTEROTOME V HAD ITS CUTTING WIRE BREAK WHEN THE GENERATOR WAS CONNECTED TO GET ENERGY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17372357 ↗MalfunctionMar 22, 2023Break; Break
UPON FOLLOW UP, IT WAS REPORTED THAT THE PATIENT WAS NOT EXAMINED WITH XRAY OR OTHER IMAGING TECHNIQUE AND NO INTERVENTION WAS REQUIRED AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16589216 ↗MalfunctionMay 12, 2022Break; Break
A USER FACILITY REPORTED TO OLYMPUS THAT THE KNIFE WIRE BROKE DURING AN ENDOSCOPIC SPHINCTEROTOMY PROCEDURE AND THE POWER WAS TURNED ON. THE DEVICE WAS REPLACED AND THE PROCEDURE COMPLETED. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE PROBLEM, REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14378148 ↗MalfunctionOct 13, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY THE HEALTH PROFESSIONAL THAT DURING INSPECTION OF THE SUBJECT DEVICE BEFORE USE, THE WIRE KNIFE WAS BROKEN OFF DURING ACTIVATION. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THE SUBJECT DEVICE WAS USED IN COMBINATION WITH ANOTHER COMPANY'S ELECTROSURGICAL UNIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12622308 ↗MalfunctionDec 27, 2019Break; Break
THE SERVICE CENTER WAS INFORMED THAT ONE SINGLE USE 3-LUMEN SPHINCTEROTOME V CUTTING WIRE (KD-V431M-0720) WAS OBSERVED TO BE CUT IN THE CENTER WHEN THE WIRE WAS REMOVED FROM THE SCOPE. THE PHYSICIAN WAS PERFORMING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WHEN THE PHYSICIAN REQUESTED THAT THE TECHNICIAN BEND THE TOME. IT WAS REPORTED THERE WAS NO RESPONSE TO THE DISTAL END AS IT WAS NOT BENDING. UPON REMOVAL OF THE DEVICE, THE OBSERVATION WAS MADE THAT THE CENTER OF THE CUTTING WIRE WAS SEPARATED/DETACHED. IT WAS REPORTED THE PHYSICIAN COMPLETED THE PROCEDURE AS NORMALLY AND THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9527485 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page