Single Use Aspiration Needle NA-201SX-4022 (REF N5432330)Recalls & reported problems
Olympus Medical Systems Corp. · NA-201SX-4022 Single Use Aspiration Needle · Class 2
All models of Single Use Aspiration Needle (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus single-use aspiration needle is used to take samples during endobronchial ultrasound procedures. The 48 reports mostly describe the needle breaking or snapping off during procedures, bending, or puncturing the bronchoscope channel, sometimes leaving a fragment in the airway that had to be retrieved with forceps. Reports include 1 death and 11 injuries. There are no recalls naming this device, and reports in the past 12 months (5) are far fewer than the 21 in the prior12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 48
- Deaths reported
- 1
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 04953170389887), received between Aug 2, 2018 and Apr 17, 2026 — classified as 1 death, 11 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Break” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break16
- Detachment of Device or Device Component13
- Material Puncture/Hole7
- Material Twisted/Bent7
- Contamination3
- Material Erosion3
- Mechanical Problem3
- Deformation Due to Compressive Stress2
Patient problems
- No Clinical Signs, Symptoms or Conditions35
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient3
- Device Embedded In Tissue or Plaque2
- Confusion/ Disorientation1
- Cough1
- Dizziness1
- Fever1
Recent adverse event reports
The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SINGLE USE ASPIRATION NEEDLE WAS RETURNED SEVERED IN TWO PIECES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24911056 ↗MalfunctionFeb 24, 2026No Apparent Adverse Event
IT WAS REPORTED THE SINGLE-USE ASPIRATION NEEDLE PUNCTURED THE MANDIBLE IN THE NEEDLE. THIS OCCURRED DURING AN ENDOBRONCHIAL ULTRASOUND PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24429571 ↗MalfunctionJan 21, 2026Material Puncture/Hole; Manufacturing, Packaging or Shipping Problem
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24127905 ↗MalfunctionDec 12, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE EXHIBITED BROKE OFF. THE ISSUE OCCURRED DURING A DIAGNOSTIC BRONCHOSCOPE ENDOBRONCHIAL ULTRASOUND. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 23790149 ↗InjuryNov 10, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC ENDOBRONCHIAL ULTRASOUND-GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION UNDER SEDATION, THE PUNCTURE NEEDLE BROKE, LEAVING THE TIP LODGED IN THE SITE THAT THE USER WAS STAGING, LOWER RIGHT TRACHEAL, 4R. USING BRONCHOSCOPY, THE TIP WAS REMOVED WITH BIOPSY FORCEPS, WITHOUT FURTHER CONSEQUENCES FOR THE PATIENT. THE PROCEDURE WAS PROLONGED BY AT LEAST OR GREATER THAN 30 MINUTES AND COMPLETED WITH A BACK-UP DEVICE. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23513642 ↗MalfunctionSep 17, 2025Difficult to Flush; Retraction Problem
FROM STAFF: DOING AN EBUS [ENDOBRONCHIAL ULTRASOUND] PROCEDURE WITH DR. [REDACTED] AND EBUS NEEDLE WAS HARD TO FLUSH SPECIMEN OUT OF AND UNABLE TO THREAD STYLET BACK INTO EBUS NEEDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23072427 ↗MalfunctionSep 2, 2025Material Puncture/Hole; Material Twisted/Bent; Material Puncture/Hole; Material Twisted/Bent
IT WAS REPORTED THAT TWO NEEDLES WERE BENT. ADDITIONALLY, PRESSURE WAS NOTED WHEN PASSING THE NEEDLE INSIDE THE WORKING CHANNEL, RESULTING IN THE CHANNEL BEING PRICKED. THE REPORTED ISSUE OCCURRED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS), WHICH WAS COMPLETED USING ANOTHER DEVICE. THE CUSTOMER RETURNED FIVE (5) NEEDLES (SAME MODEL NUMBER, DIFFERENT LOT) RELATED TO THIS EVENT. THERE WERE NO REPORTS OF PATIENT HARM. THIS IS REPORT 3 OF 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22947036 ↗MalfunctionSep 2, 2025Material Puncture/Hole; Material Twisted/Bent; Material Puncture/Hole; Material Twisted/Bent
IT WAS REPORTED THAT TWO NEEDLES WERE BENT. ADDITIONALLY, PRESSURE WAS NOTED WHEN PASSING THE NEEDLE INSIDE THE WORKING CHANNEL, RESULTING IN THE CHANNEL BEING PRICKED. THE REPORTED ISSUE OCCURRED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS), WHICH WAS COMPLETED USING ANOTHER DEVICE. THE CUSTOMER RETURNED FIVE (5) NEEDLES (SAME MODEL NUMBER, DIFFERENT LOT) RELATED TO THIS EVENT. THERE WERE NO REPORTS OF PATIENT HARM. THIS IS REPORT 5 OF 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22946628 ↗MalfunctionSep 2, 2025Material Puncture/Hole; Material Twisted/Bent; Material Puncture/Hole; Material Twisted/Bent
IT WAS REPORTED THAT TWO NEEDLES WERE BENT. ADDITIONALLY, PRESSURE WAS NOTED WHEN PASSING THE NEEDLE INSIDE THE WORKING CHANNEL, RESULTING IN THE CHANNEL BEING PRICKED. THE REPORTED ISSUE OCCURRED DURING A DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS), WHICH WAS COMPLETED USING ANOTHER DEVICE. THE CUSTOMER RETURNED FIVE (5) NEEDLES (SAME MODEL NUMBER, DIFFERENT LOT) RELATED TO THIS EVENT. THERE WERE NO REPORTS OF PATIENT HARM. THIS IS REPORT 4 OF 5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22946133 ↗MalfunctionAug 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED DURING A DIAGNOSTIC EBUS (ENDOBRONCHIAL ULTRASOUND) PROCEDURE, THE NEEDLE WAS WORKING FINE (HAD BEEN USED ON A LYMPH NODE STATION BEFORE) AND WHEN THE PHYSICIAN WAS FINISHING UP ON A NODE, HE PULLED BACK ON THE NEEDLE TO LOCK IT BEFORE REMOVING FROM THE WORKING CHANNEL AND THE WHOLE PLASTIC TOP OF NEEDLE SEPARATED FROM THE BOTTOM BY THE SHEATH ADJUSTER. THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882487 ↗InjuryJun 27, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN EBUS (ENDOBRONCHIAL ULTRASOUND) PROCEDURE FOR A DIAGNOSTIC TRANSBRONCHIAL BIOPSY USING A SINGLE-USE ASPIRATION NEEDLE, THE SHEATH OF THE NEEDLE BROKE OFF AND BECAME LODGED IN THE RIGHT MAIN BRONCHUS DURING THE SECOND PUNCTURE AS THE NEEDLE WAS BEING WITHDRAWN. THE PROCEDURE WAS PROLONGED BY APPROXIMATELY FIVE MINUTES DUE TO THE REPLACEMENT OF THE EBUS DEVICE WITH A REGULAR BRONCHO FIBERSCOPE AND THE SUBSEQUENT REMOVAL OF THE ELEMENT FROM THE BRONCHUS. THE SHEATH WAS REMOVED WITH FORCEPS, AND THE BRONCHIAL TREE WAS EXAMINED, WITH NO OTHER FOREIGN BODIES FOUND. DUE TO THE SITUATION, THE SINGLE USE ASPIRATION NEEDLE WAS NOT USED FURTHER AND THE PROCEDURE WAS COMPLETED WITH USE OF A SIMILAR DEVICE. IN ADDITION, THE SINGLE USE ASPIRATION NEEDLE WAS INSPECTED PRIOR TO USE WITH NO ABNORMALITIES OBSERVED, AND THE MANDREL THAT SECURES THE SINGLE USE ASPIRATION NEEDLE WAS ALSO EXAMINED, WITH NO ABNORMALITIES OBSERVED. THERE WAS NO REPORT OF PATIENT INJURY OR HARM AS A RESULT OF THE DEVICE ISSUE. THE PATIENT'S CURRENT HEALTH STATUS WAS DESCRIBED AS SIGNIFICANTLY DEBILITATED, WITH EXERTIONAL DYSPNEA; HOWEVER, THERE IS CURRENTLY NO EVIDENCE TO SUGGEST THAT THIS CO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22347793 ↗MalfunctionJun 18, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT DURING A DIAGNOSTIC NEEDLE BIOPSY PROCEDURE, THE SUBJECT DEVICE WAS UNABLE TO ATTACH TO THE CONNECTION SLIDE OF THE SINGLE-USE VALVE. THE SLIDE DID NOT LOCK, AND THE NEEDLE PIERCED THE ECHO ENDOSCOPE SHEATH. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22284263 ↗MalfunctionJun 13, 2025Contamination; Material Erosion; Contamination; Material Erosion
IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE HAD A BLACK ABRASION MARK ON THE COMPRESS DUE TO A LEAKAGE OF SALINE SOLUTION. THE ISSUE OCCURRED DURING AN ENDOBRONCHIAL ULTRASOUND-GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION. THE PROCEDURE WAS COMPLETED WITHOUT DELAY WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22205574 ↗MalfunctionJun 13, 2025Contamination; Material Erosion; Contamination; Material Erosion
IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE HAD A BLACK ABRASION MARK ON THE COMPRESS DUE TO A LEAKAGE OF SALINE SOLUTION. THE ISSUE OCCURRED DURING AN ENDOBRONCHIAL ULTRASOUND-GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION. THE PROCEDURE WAS COMPLETED WITHOUT DELAY WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22205551 ↗MalfunctionJun 13, 2025Contamination; Material Erosion; Contamination; Material Erosion
IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE HAD A BLACK ABRASION MARK ON THE COMPRESS DUE TO A LEAKAGE OF SALINE SOLUTION. THE ISSUE OCCURRED DURING AN ENDOBRONCHIAL ULTRASOUND-GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION. THE PROCEDURE WAS COMPLETED WITHOUT DELAY WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22205507 ↗InjuryJun 11, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DIAGNOSTIC ENDOSCOPIC BRONCHIAL ULTRASOUND (EBUS), THE TIP PF THE SUBJECT DEVICE BROKE OFF INSIDE THE PATIENT. THE CLINICIAN WAS ASPIRATING FLUID FROM THE PATIENT FOR CYTOLOGY. THE SUBJECT DEVICE WAS WITHDRAWN AFTER ONE SUCCESSFUL PASS. WHEN THE CLINICIAN ATTEMPTED TO INSERT THE NEEDLE AGAIN, IT WOULDN'T GO ALL THE WAY IN. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. DURING THE FINAL INSPECTION OF THE PROCEDURE, THE CLINICIAN VISUALLY EXAMINED THE AREA INSIDE THE PATIENT AND SAW A SHINY OBJECT. THE BROKEN PIECE WAS EXTRACTED WITH FORCEPS. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22191153 ↗InjuryJun 11, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE AN UPDATE TO THE FOLLOWING FIELDS:D3, D4, D9, F9 AND F14.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22191152 ↗MalfunctionJun 2, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT PRIOR TO A BRONCHOSCOPY PROCEDURE, THE SUBJECT DEVICE WAS INSPECTED AFTER BEING TAKEN OUT OF THE BOX, AND THE DOCTOR NOTICED THE LOCK WAS STUCK AND THE NEEDLE WAS NOT ABLE TO BE DEPLOYED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22128555 ↗MalfunctionMay 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22034663 ↗MalfunctionMay 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22018356 ↗MalfunctionMay 2, 2025Material Puncture/Hole; Deformation Due to Compressive Stress; Material Twisted/Bent; Material Puncture/Hole; Deformation Due to Compressive Stress; Material Twisted/Bent
IT WAS REPORTED THAT DURING A DIAGNOSTIC PUNCTURE PROCEDURE, THE CATHETER OF THE NEEDLE WAS KINKED, AND THE NEEDLE DAMAGED THE WORKING CHANNEL. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21954641 ↗InjuryMar 19, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT A SINGLE-USE ASPIRATION NEEDLE BROKE INSIDE THE PATIENT, LEAVING A PIECE OF APPROXIMATELY 3 CM LODGED IN THE LEFT MAIN BRONCHUS, WHICH WAS RECOVERED USING A CLAMP. THERE WAS A PROCEDURAL DELAY OF LESS THAN 30 MINUTES WHILE THE PATIENT WAS SEDATED. A BACKUP DEVICE WAS USED TO COMPLETE THE DIAGNOSTIC ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21646576 ↗MalfunctionMar 17, 2025Failure to Align; Failure to Align
IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE WENT OUT OF ADJUSTMENT DURING THE FIRST PUNCTURE, EVEN THOUGH IT HAD BEEN ADJUSTED BEFOREHAND. THIS LED TO THE OPERATOR CHANNEL OF THE ECHO-ENDOSCOPE BEING PIERCED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21625482 ↗MalfunctionMar 13, 2025Deformation Due to Compressive Stress; Detachment of Device or Device Component; Deformation Due to Compressive Stress; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21595712 ↗InjuryJan 2, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A BRONCHIAL PUNCTURE PROCEDURE, THE SINGLE USE ASPIRATION NEEDLE BROKE OFF IN THE PATIENT¿S LUNG. THE DEVICE WAS CHANGED TO USE A BRONCHOSCOPE WITH AN OPERATIONAL CHANNEL AND THE NEEDLE WAS RETRIEVED WITH A BRONCHIAL BIOPSY FORCEPS. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21057636 ↗
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall118 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall29 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page