Single Use Sphincterotome V(Distal Wireguided) KD-VC411Q-0720 (REF N5777530)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use Sphincterotome V(Distal Wireguided) · Class 2
All models of Single Use Sphincterotome V(Distal Wireguided) (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04953170399312), received between Nov 7, 2021 and Jul 5, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 5, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE SINGLE USE PRELOADED SPHINCTEROTOME EXHIBITED THE COVER WAS PEELED AND THE KNIFE WIRE WAS EXPOSED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19683608 ↗MalfunctionNov 22, 2022Break; Break
THE CUSTOMER REPORTED THAT DURING AN ENDOSCOPIC THORACIC SYMPATHECTOMY (EST) CASE, THE KNIFE WIRE BROKE DURING THE SECOND INCISION. THE PROCEDURE WAS COMPLETED BY REPLACING IT WITH ANOTHER EQUIVALENT PRODUCT, AND NO HEALTH DAMAGE OCCURRED. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15848773 ↗MalfunctionAug 24, 2022Break; Break
AS REPORTED, WHILE THE DOCTOR WAS USING KD-VC411Q-0720, THE KNIFE WIRE BROKE WHEN THE ELECTRICITY WAS APPLIED DURING THE DUODENAL PAPILLOTOMY ENDOSCOPIC SPHINCTEROTOMY (EST) THERAPEUTIC PROCEDURE. IT IS UNKNOWN WHETHER IT WAS AT THE TIME OF THE FIRST ENERGIZATION OR IT BROKE AFTER SEVERAL ENERGIZATIONS. THE DOCTOR REPLACED IT WITH ANOTHER OF THE SAME PRODUCT AND COMPLETED THE PROCEDURE. NO PATIENT HEALTH HAZARD. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT. THIRD PARTY HIGH FREQUENCY ABLATION DEVICE- BIO 3 WAS USED WITH THE DEVICE. THIS EVENT INCLUDES TWO REPORTS TO ADDRESS THE TWO KNIFE WIRE THAT BROKE DURING THE CASE WITH THE SAME DEVICE MODEL (KD-VC411Q-0720). REPORT WITH PATIENT IDENTIFIER (B)(6) (LOT NUMBER UNKNOWN). REPORT WITH PATIENT IDENTIFIER (B)(6) (LOT 23K). THIS REPORT IS FOR REPORT WITH PATIENT IDENTIFIER (B)(6) (LOT 23K).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15289313 ↗MalfunctionDec 12, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED FROM USER FACILITY THAT THE KNIFE BROKE WHEN ENERGIZED DURING THE ENDOSCOPIC PAPILLOTOMY (EST) PROCEDURE. ON DECEMBER 1, OMSC RECEIVED THE FOLLOWING INFORMATION OF THE PROCEDURE. THE HIGH FREQUENCY POWER SUPPLY WAS VIO300D (NON-OLYMPUS DEVICE). THE ENDOSCOPE WAS JF-260V. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12981965 ↗MalfunctionNov 7, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP.(OMSC) WAS INFORMED FROM USER FACILITY THAT THE CUTTING WIRE BROKE WHEN ENERGIZED. THE DEVICE WAS REPLACED AND THE PROCEDURE CONTINUED, BUT THE SECOND DEVICE ALSO BROKE WHEN ENERGIZED. IN EACH CASE, THE CUTTING WIRE BROKE DURING THE 2-3RD ENERGIZATION, AND THE BROKEN KNIFE WIRE DID NOT FALL INTO THE PATIENT'S BODY. THE PROCEDURE WAS ALMOST COMPLETED WITH THE SECOND DEVICE, AND THE PROCEDURE WAS COMPLETED WITHOUT ANY PROBLEMS BY QUARRYING AS IT WAS. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT. THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. IN THE EVALUATION OF OMSC THE FOLLOWING WAS CONFIRMED, (THE FIRST DEVICE). THE CUTTING WIRE WAS BROKEN. (THE SECOND DEVICE). THE CUTTING WIRE WAS NOT BROKEN. THEREFORE, BASED ON THE EVALUATION, OMSC CONCLUDED THAT THE SECOND DEVICE WAS NOT REPORTABLE EVENT. OMSC REPORT THE MDR FOR THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12764237 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page