Single Use Preloaded Sphincterotome V(Distal Wireguided) KD-VC631Q-07203S (REF N5779430)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use Preloaded Sphincterotome V(Distal Wireguided) · Class 2
All models of Single Use Preloaded Sphincterotome V (Distal Wireguided) (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04953170399503), received between Apr 15, 2024 and Apr 8, 2026 — classified as 5 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Detachment of Device or Device Component5
- Device Fell1
- Difficult to Insert1
- Material Deformation1
- Peeled/Delaminated1
Patient problems
- Foreign Body In Patient5
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Difficult to Insert
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24823649 ↗InjuryDec 19, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) THE TIP BROKE AND FELL INTO THE PATIENT. THE CUSTOMER DID NOT RETRIEVE THE TIP.
Patient problems: Foreign Body In Patient
MAUDE report 23857424 ↗InjuryNov 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT WHEN USING THE SINGLE USE PRELOADED SPHINCTEROTOME DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY, THE PHYSICIAN NOTICED A MARK OR DOT ON THE X-RAY IMAGE. THE DEVICE WAS TAKEN OUT TO BE INSPECTED, NO DAMAGE WAS SEEN, AND THE DEVICE WAS INSERTED BACK INTO THE SCOPE. WHEN THE WIRE WAS INSERTED BACK INTO THE BILIARY DUCT UNDER X-RAY, THE MARK OR DOT WAS GONE. A SMALL PIECE OF METAL WAS OBSERVED NEXT TO THE DUCT. THE PHYSICIAN NOTED THAT THE PIECE OF METAL CAME FROM THE WIRE, AND THAT THERE WAS A PIECE MISSING AT THE JUNCTION OF THE SOFT PART OF THE WIRE AND WIRE CORE ITSELF, WHICH COULD BE THIS PIECE OF METAL. THE PHYSICIAN DETERMINED TO LEAVE THE METAL PIECE IN PLACE SINCE IT WILL BE VOIDED NATURALLY. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.
Patient problems: Foreign Body In Patient
MAUDE report 23612354 ↗InjuryNov 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT WHEN USING THE SINGLE USE PRELOADED SPHINCTEROTOME DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY, THE PHYSICIAN NOTICED A MARK OR DOT ON THE X-RAY IMAGE. THE DEVICE WAS TAKEN OUT TO BE INSPECTED, NO DAMAGE WAS SEEN, AND THE DEVICE WAS INSERTED BACK INTO THE SCOPE. WHEN THE WIRE WAS INSERTED BACK INTO THE BILIARY DUCT UNDER X-RAY, THE MARK OR DOT WAS GONE. A SMALL PIECE OF METAL WAS OBSERVED NEXT TO THE DUCT. THE PHYSICIAN NOTED THAT THE PIECE OF METAL CAME FROM THE WIRE, AND THAT THERE WAS A PIECE MISSING AT THE JUNCTION OF THE SOFT PART OF THE WIRE AND WIRE CORE ITSELF, WHICH COULD BE THIS PIECE OF METAL. THE PHYSICIAN DETERMINED TO LEAVE THE METAL PIECE IN PLACE SINCE IT WILL BE VOIDED NATURALLY. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.
Patient problems: Foreign Body In Patient
MAUDE report 23612332 ↗MalfunctionNov 3, 2025Material Deformation; Material Deformation
IT WAS REPORTED THE SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED) WAS STRIPPED AT THE TIP. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE AND WAS COMPLETED USING AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23453713 ↗InjuryOct 21, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE SINGLE USE PRELOADED SPHINCTEROTOME WIRE BROKE AND FELL INTO THE PATIENT¿S PANCREATIC DUCT. THE SURGEON ATTEMPTED TO RETRIEVE THE BROKEN WIRE WITH GRASPING FORCEPS BUT WAS UNSUCCESSFUL. THE PATIENT RETURNED FOR AN ADDITIONAL PROCEDURE TO TRY TO REMOVE THE PIECE OF BROKEN WIRE, HOWEVER THAT WAS AGAIN UNSUCCESSFUL. A FOLLOW-UP IS SCHEDULED TO DISCUSS POSSIBLE SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 23354548 ↗InjuryOct 21, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER. THIS SUPPLEMENTAL IS BEING SUBMITTED TO CORRECT CODING. UPDATES TO B2 AND F10.
Patient problems: Foreign Body In Patient
MAUDE report 23354542 ↗MalfunctionJun 24, 2025Break; Peeled/Delaminated; Break; Peeled/Delaminated
IT WAS REPORTED THE SINGLE USE PRELOADED SPHINCTEROTOME V IRON WIRE OF THE KNIFE IS BROKEN. THE ISSUE OCCURRED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 22322005 ↗MalfunctionMar 18, 2025Break; Break
IT WAS REPORTED THAT THE CUTTING WIRE BROKE WHEN ENERGY WAS APPLIED TO THE DEVICE. TWO CUTS WERE MADE AND DEVICE FAILED. THE EVENT OCCURRED DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21629250 ↗MalfunctionMay 22, 2024Device Fell; Device Fell
IT WAS REPORTED, THE SPHINCTEROTOME V¿S WIRE FELL OFF DURING A THERAPEUTIC PROCEDURE (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY). THE INTENDED PROCEDURE WAS COMPLETED USING THE SAME SET OF DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19373682 ↗MalfunctionApr 15, 2024
CLEVER CUT 3V USED DURING ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY), AT THE TIME OF THE SPHINCTEROTOMY DURING THE BOWING, THE WIRE BROKE WHILE AT THE AMPULLA INSIDE THE PATIENT IN THE SCOPE.
Patient problems: Insufficient Information
MAUDE report 19104395 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.