Single Use Preloaded Sphincterotome V KD-V611M-07201SRecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of Single Use Preloaded Sphincterotome V (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04953170403040), received between Jun 8, 2023 and Jun 8, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 8, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE DEVICE WIRE WAS TORN DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT. THE DEVICE WAS RETURNED TO SERVICE WHERE EVALUATION FOUND THE CUTTING WIRE WAS BROKEN. THIS REPORT IS BEING SUBMITTED FOR THE REPORTABLE MALFUNCTION FOUND AT EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17088227 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page