CONNECTING TUBE MAJ-2114Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of CONNECTING TUBE (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04953170404092), received between Nov 19, 2022 and Jul 3, 2024 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2026Class IIOpen, Classified
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
Z-1902-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break6
- Detachment of Device or Device Component3
- Connection Problem2
- Loose or Intermittent Connection1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2024Break; Break
IT WAS REPORTED, THE CONNECTING TUBE WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19670854 ↗MalfunctionApr 10, 2024Break; Break
AN OLYMPUS REPRESENTATIVE REPORTED THAT THE OER-ELITE (ENDOSCOPE REPROCESSOR) SEPARATED HOOK PART CONNECTOR TUBE CONTINUED TO BREAK OFF. THE REPORTED ISSUE WAS OBSERVED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19078472 ↗MalfunctionFeb 12, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THE TAB BROKE ON THE CONNECTING TUBE MAKING IT IMPOSSIBLE TO CONNECT TO THE ENDOSCOPE REPROCESSOR. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18692525 ↗MalfunctionFeb 12, 2024Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THE TAB BROKE ON THE REPROCESSOR CONNECTING TUBE MAKING IT IMPOSSIBLE TO CONNECT TO THE ENDOSCOPE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18692515 ↗MalfunctionOct 30, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT WHILE REPROCESSING VARIOUS OLYMPUS CONNECTING TUBES, THE CLIPS BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT DURING THESE EVENTS. THIS REPORT IS 8 OF 14. PLEASE SEE THE FOLLOWING PATIENT IDENTIFIERS FOR THE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18030000 ↗MalfunctionOct 30, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT WHILE REPROCESSING VARIOUS OLYMPUS CONNECTING TUBES, THE CLIPS BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT DURING THESE EVENTS. THIS REPORT IS 8 OF 14. PLEASE SEE THE FOLLOWING PATIENT IDENTIFIERS FOR THE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029996 ↗MalfunctionSep 28, 2023Break; Break
THE OLYMPUS REPRESENTATIVE REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE CONNECTING TUBE WING BROKE AFTER A FEW USES. THE ISSUE WAS FOUND IN THE IN CLEANING AND DISINFECTION ROOM. THERE WAS NO PATIENT INVOLVEMENT AND NO USER INJURY. RELATED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17833293 ↗MalfunctionSep 28, 2023Break; Break
THE OLYMPUS REPRESENTATIVE REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE CONNECTING TUBE WING BROKE AFTER A FEW USES. THE ISSUE WAS FOUND IN THE IN CLEANING AND DISINFECTION ROOM. THERE WAS NO PATIENT INVOLVEMENT AND NO USER INJURY. RELATED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17833129 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2138). (B
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086988 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114): THIS REPORT (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2138); (B)(6) (MODEL NUMBER: M
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086920 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330), (B)(6) (MODEL NUMBER: MAJ-2118), (B)(6) (MODEL NUMBER: MAJ-2115), (B)(6) (MODEL NUMBER: MAJ-2114), (B)(6) (MODEL NUMBER: MAJ-2116), (B)(6) (MODEL NUMBER: MAJ-2138).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086612 ↗MalfunctionMay 31, 2023Loose or Intermittent Connection; Loose or Intermittent Connection
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BIN. THERE IS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THE REPORTED PROBLEM WAS FOUND DURING USE. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17033411 ↗MalfunctionNov 19, 2022Separation Problem; Separation Problem
THE CUSTOMER REPORTED TO OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC), THERE WERE MULTIPLES OER-ELITE CONNECTORS THAT WERE ¿BAD¿. THE CUSTOMER REPORTED THAT IT WAS AN OUT OF BOX FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15831066 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.
- Z-0638-2022 · Dec 13, 2021 · TerminatedOER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor
Other Accessories, Cleaning, For Endoscope devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page