CONNECTING TUBE MAJ-2115Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of CONNECTING TUBE (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170404108), received between Nov 19, 2022 and Apr 10, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2026Class IIOpen, Classified
XXX
Z-1903-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Connection Problem2
- Break1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 10, 2024Break; Break
AN OLYMPUS REPRESENTATIVE REPORTED THAT THE CONNECTOR TUBE OF THE REPROCESSOR CONTINUED TO BREAK OFF. THE REPORTED ISSUE WAS OBSERVED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19078545 ↗MalfunctionOct 30, 2023Connection Problem; Connection Problem
IT WAS REPORTED THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D) DEVICE WAS SEEN IN CLINIC. PROGRAMMING WAS PERFORMED TO PACE FASTER THAN THE CURRENT RATE TO SLOW DOWN VENTRICULAR TACHYCARDIA (VT). THERE WERE STORED EPISODES OF PACEMAKER MEDIATED TACHYCARDIA (PMT) WHICH WERE DETERMINED TO NOT BE TRUE PMT RATHER THE PATIENT'S SLOW VENTRICULAR TACHYCARDIA (VT). THE PATIENT HAD A SICK HEART, AND THE ATRIUM APPEARED TO BE DRIVING THE RATES. TS RECOMMENDED THAT RECOMMENDED THAT THE PMT ALGORITHM TO BE PROGRAMMED OFF AND THAT THE PATIENT'S ATRIAL RATE NEEDS TO BE ADDRESSED BY THE FOLLOWING PHYSICIAN. THIS DEVICE CURRENTLY REMAINS IN SERVICE. NO ADVERSE EFFECTS WERE REPORTED. ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PATIENT WAS SEEN IN THE CLINIC WITH THE HISTORY OF SLOW VT. DURING THE VISIT IN CLINIC, THERE WAS GOOD CAPTURE ON RV AND LV LEADS. TECHNICAL SERVICES (TS) REVIEWED AND STATED THAT RV SENSE GOT CLOSE ENOUGH TO THE AP-SR THAT IT FALLS INTO CROSS CHAMBER BLANKING PERIOD [RV] AND IS FUNCTIONALLY UNDERSENSED AND THERE WAS FUNCTIONAL LOSS OF CAPTURE (LOC). TS STATED THAT THERE WAS NORMAL OPERATION WITH JUST THE PRODUCT OF THE TIMING AND T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029502 ↗MalfunctionOct 30, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT WHILE REPROCESSING VARIOUS OLYMPUS CONNECTING TUBES, THE CLIPS BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT DURING THESE EVENTS. THIS REPORT IS 4 OF 14. PLEASE SEE THE FOLLOWING PATIENT IDENTIFIERS FOR THE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029427 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2138). (B
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086535 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115): THIS REPORT. (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17086399 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2118) ;(B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118) ;(B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115): THIS REPORT; (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17085896 ↗MalfunctionJun 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330), (B)(6) (MODEL NUMBER: MAJ-2330), (B)(6) (MODEL NUMBER: MAJ-2330), (B)(6) (MODEL NUMBER: MAJ-2330), (B)(6) (MODEL NUMBER: MAJ-2118), (B)(6) (MODEL NUMBER: MAJ-2118), (B)(6) (MODEL NUMBER: MAJ-2118), (B)(6) (MODEL NUMBER: MAJ-2118), (B)(6) (MODEL NUMBER: MAJ-2115): THIS REPORT, (B)(6) (MODEL NUMBER: MAJ-2115), (B)(6) (MODEL NUMBER: MAJ-2115), (B)(6) (MODEL NUMBER: MAJ-2115), (B)(6) (MODEL NUMBER: MAJ-2114), (B)(6) (MODEL NUMBER: MAJ-2114), (B)(6) (MODEL NUMBER: MAJ-2114), (B)(6) (MODEL NUMBER: MAJ-2114), (B)(6) (MODEL NUMBER: MAJ-2116), (B)(6) (MODEL NUMBER: MAJ-2116), (B)(6) (MODEL NUMBER: MAJ-2116), (B)(6) (MODEL NUMBER:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17085823 ↗MalfunctionNov 19, 2022Separation Problem; Separation Problem
CUSTOMER CONFIRMED ALL THE CONNECTORS BROKE AT THE METAL CLIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15831088 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.
- Z-0638-2022 · Dec 13, 2021 · TerminatedOER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor
Other Accessories, Cleaning, For Endoscope devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page