Skip to content

CONNECTING TUBE MAJ-2117Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CONNECTING TUBE (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170404122), received between May 31, 2023 and Jun 16, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 25, 2026Class IIOpen, Classified

    Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.

    Z-1905-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT HIS OLYMPUS CONNECTING TUBES WERE BREAKING DURING REPROCESSING. ACCORDING TO THE INITIAL REPORTER, THE CONNECTOR¿S RIGHT AND LEFT LEGS WERE FALLING OFF. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT. FOR REFERENCE TO RELATED EVENTS, PLEASE SEE REPORTS WITH THE FOLLOWING PATIENT IDENTIFIERS: (B)(6), (B)(6), (B)(6), (B)(6), (B0(6), (B)(6), (B)(6), (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17143537 ↗
  • MalfunctionMay 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO OLYMPUS, THE CONNECTING TUBE SIDE CONNECTORS CRACK AND FALL OFF. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6), (MAJ-2117, THIS REPORT) (B)(6), (MAJ-2111) (B)(6), (MAJ-2110) (B)(6), (MAJ-2330) (B)(6), (MAJ-2110).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17035121 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.

  • Z-0638-2022 · Dec 13, 2021 · Terminated
    OER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor

Other Accessories, Cleaning, For Endoscope devices

All Accessories, Cleaning, For Endoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page