CONNECTING TUBE MAJ-2330Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of CONNECTING TUBE (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04953170404207), received between Nov 19, 2022 and May 20, 2026 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K190969); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2026Class IIOpen, Classified
Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
Z-1909-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break7
- Detachment of Device or Device Component6
- Connection Problem2
- Device Reprocessing Problem1
- Failure to Clean Adequately1
- Separation Failure1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2026Device Reprocessing Problem; Use of Device Problem; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION AT THE CUSTOMER SITE, THAT THE CONNECTING TUBE EXHIBITED VISIBLE DEBRIS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232904 ↗MalfunctionFeb 27, 2026Break
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ENDOSCOPE REPROCESSOR HAD A BROKEN CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24471621 ↗MalfunctionNov 15, 2024Break; Break
IT WAS REPORTED THAT THE CONNECTING TUBE HAD A BROKEN CONNECTOR FOR THE ENDOSCOPE REPROCESSOR. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20692374 ↗MalfunctionNov 5, 2024Break; Break
IT WAS REPORTED THE CONNECTING TUBE TUBING CLIPS BROKE OFF. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20615307 ↗MalfunctionJul 12, 2024Break; Break
IT WAS REPORTED THE CONNECTING TUBE WAS CRACKED AND BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19735556 ↗MalfunctionJun 26, 2024Break; Break
IT WAS REPORTED THE CONNECTING TUBE WAS FRAGILE AND BREAKING, THUS NEEDING TO BE REPLACED FREQUENTLY. THE ISSUE OCCURRED DURING PREPARATION FOR USE IN UNSPECIFIED REPROCESSING PROCEDURES. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19614750 ↗MalfunctionJun 26, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE CONNECTING TUBE WERE BROKEN. IT WAS REPORTED THAT THE CONNECTOR'S RIGH AND LEFT LEG CONTINUED TO FALL OFF. IT WAS ALSO REPORTED THAT OER-ELITE ENDOSCOPE REPROCESSOR WAS INTERMITTENTLY RECEIVING AN EO61 ERROR CODE, THE ISSUE WAS DUE TO THE CONNECTING TUBE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17203846 ↗MalfunctionJun 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THAT HIS OLYMPUS CONNECTING TUBES WERE BREAKING DURING REPROCESSING. ACCORDING TO THE INITIAL REPORTER, THE CONNECTOR¿S RIGHT AND LEFT LEGS WERE FALLING OFF. THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT. FOR REFERENCE TO RELATED EVENTS, PLEASE SEE REPORTS WITH THE FOLLOWING PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17143434 ↗MalfunctionJun 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330). (B)(6) (MODEL NUMBER: MAJ-2330): THIS REPORT. (B)(6) (MODEL NUMBER: MAJ-2330). B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2118). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2115). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2114). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116). (B)(6) (MODEL NUMBER: MAJ-2116).. (B)(6) (MODEL NUMBER:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17043284 ↗MalfunctionJun 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THE CONNECTORS BECAME DETACHED TOO EASILY AND AT TIMES, THE CONNECTOR HOOK BECOMES DETACHED FROM THE CONNECTOR. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) (MODEL NUMBER: MAJ-2330): THIS REPORT (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2330); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2118); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2115); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2114); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: MAJ-2116); (B)(6) (MODEL NUMBER: M
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17042668 ↗MalfunctionMay 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS, THE CONNECTING TUBE SIDE CONNECTORS CRACK AND FALL OFF. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) , (MAJ-2117). (B)(6) , (MAJ-2111). (B)(6) , (MAJ-2110). (B)(6) , (MAJ-2330, THIS REPORT). (B)(6) , (MAJ-2110).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17035125 ↗MalfunctionMay 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS, THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17013823 ↗MalfunctionMay 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED TO OLYMPUS, THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THE CUSTOMER STATED THE CONNECTORS ARE FALLING APART. THERE WAS A METAL CLIP IN THE CLASPS THAT FALLS OFF, CAUSING THE PLASTIC ¿WINGS¿ TO COME OFF THE CONNECTOR. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE MEDICAL DEVICE REPORT (MDR) IS RELATED TO THE FOLLOWING PATIENT IDENTIFIERS: (B)(6) .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17013281 ↗MalfunctionMar 31, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING REPROCESSING, THERE HAS BEEN AN ONGOING ISSUE WITH THE BROKEN CONNECTING TUBE WHILE BEING USED WITH THE OER-ELITE PREPROCESSOR. THE CONNECTORS ARE BREAKING ON THE PREPROCESSOR CONNECTOR SIDE. THE CONNECTING TUBE CANNOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT. THIS COMPLAINT IS RELATED TO PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16658847 ↗MalfunctionFeb 8, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED THE SUBJECT DEVICE COULD NOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16330254 ↗MalfunctionFeb 8, 2023Connection Problem; Connection Problem
THE CUSTOMER REPORTED THE SUBJECT DEVICE COULD NOT BE CONNECTED TO THE CHANNEL CONNECTOR IN THE REPROCESSING BASIN. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16327257 ↗MalfunctionNov 19, 2022Separation Failure; Separation Failure
CUSTOMER CONFIRMED ALL THE CONNECTORS BROKE AT THE METAL CLIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15831102 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190969. One submission often covers many models, so check the product and lot details in each record.
- Z-0638-2022 · Dec 13, 2021 · TerminatedOER-Elite Endoscope Reprocessor-for use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes, their accessories, and endoscope reprocessor accessor
Other Accessories, Cleaning, For Endoscope devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page