Skip to content

EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TLRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA III COLONOVIDEOSCOPE (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III colonoscope is a flexible camera scope used to look inside the colon. Many of the 178 reports describe scopes that did not clean properly or had residue inside. The reports include 174 malfunctions, 4 injuries and no deaths. Reports from the last 12 months number 72, up from 15 in the 12 months before. Two recalls are open, covering repaired scopes missing protective adhesive and water left in scope channels.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
184
Deaths reported
0
Injuries reported
4
Recalls naming it
2

Summary

The FDA's MAUDE database contains 184 adverse event reports that list this exact device (DI 04953170420702), received between Oct 19, 2021 and Sep 30, 2026 — classified as 4 injuries, 180 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (98 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K192793); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 6 reportsJan 2024: 8 reportsFeb 2024: 9 reportsMar 2024: 7 reportsApr 2024: 5 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 5 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 8 reportsJan 2026: 16 reportsFeb 2026: 7 reportsMar 2026: 12 reportsApr 2026: 8 reportsMay 2026: 5 reportsJun 2026: 6 reportsJul 2026: 2 reportsAug 2026: 5 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 11, 2024Class IIOpen, Classified

    Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

    Z-1212-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately98
  • Communication or Transmission Problem21
  • Microbial Contamination of Device20
  • No Display/Image15
  • Device Reprocessing Problem12
  • Residue After Decontamination8
  • Poor Quality Image5
  • Contamination4

Patient problems

  • No Clinical Signs, Symptoms or Conditions182
  • Inflammation2

Recent adverse event reports

The 25 most recent of 184 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Erratic or Intermittent Display; No Display/Image; Poor Quality Image; Communication or Transmission Problem; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954835 ↗
  • MalfunctionSep 29, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939151 ↗
  • MalfunctionSep 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937750 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, dark residue was observed in the small light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50913005 ↗
  • MalfunctionSep 23, 2026No Display/Image; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50875827 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    The customer returned the Colonovideoscope for a separate issue. During the initial evaluation of the returned device, cylinder and channel with white residue were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859588 ↗
  • MalfunctionSep 9, 2026No Display/Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50692703 ↗
  • MalfunctionSep 2, 2026Erratic or Intermittent Display; No Display/Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50561423 ↗
  • MalfunctionAug 20, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50351401 ↗
  • MalfunctionAug 18, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50295734 ↗
  • MalfunctionAug 17, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50277819 ↗
  • MalfunctionAug 14, 2026Poor Quality Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50254694 ↗
  • MalfunctionAug 10, 2026Failure to Clean Adequately

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device it was observed, white residue in air water channel and air water cylinder were observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50194135 ↗
  • MalfunctionJul 30, 2026Optical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50082528 ↗
  • MalfunctionJul 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50068321 ↗
  • MalfunctionJun 26, 2026Residue After Decontamination

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING DEVICE EVALUATION, THE REPAIR CENTER TECHNICIAN IDENTIFIED THE DEVICE HAD WHITE RESIDUE IN AIR WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25638215 ↗
  • MalfunctionJun 26, 2026Residue After Decontamination

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT DEVICE HAD WHITE RESIDUE IN THE AIR WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25635891 ↗
  • MalfunctionJun 24, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE REPAIR TECHNICIAN FOUND WHITE RESIDUE INSIDE THE CYLINDER AND CHANNEL ALSO DEBRIS IN THE LENS, CAUSE NOT IDENTIFIED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25607179 ↗
  • MalfunctionJun 19, 2026Failure to Clean Adequately

    THE CUSTOMER REPORTED THAT DEBRIS HAD BEEN NOTICED ON THE CLEANING BRUSHES DURING REPROCESSING INVOLVING AN OLYMPUS COLONOVIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25568430 ↗
  • MalfunctionJun 5, 2026No Display/Image

    THE CUSTOMER REPORTED NO IMAGE DURING A REPROCESSING INVOLVING AN OLYMPUS COLONOVIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414713 ↗
  • MalfunctionJun 4, 2026Flare or Flash; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED BURN MARKS ON THE DISTAL END PLASTIC COVER AND WHITE RESIDUE INSIDE THE AIR/WATER SUPPLY CHANNEL AND AUXILIARY WATER FLOW CHANNEL AND ALSO IN THE AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25394589 ↗
  • MalfunctionMay 26, 2026Failure to Clean Adequately

    THE COLONOVIDEOSCOPE WAS RETURNED TO OLYMPUS FOR ASSET RETURN HOWEVER AN MDR REPORTABLE FAILURE WAS FOUND DURING INSPECTION OF THE DEVICE, THE AIR/WATER CYLINDER HAD WHITE FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25279577 ↗
  • MalfunctionMay 22, 2026Failure to Clean Adequately

    THE COLONOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF NO CUSTOMER ALLEGATION. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, WHITE RESIDUE IN THE AIR WATER CHANNEL AND CYLINDER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25261912 ↗
  • MalfunctionMay 19, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD HIGH GLUE IN C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25218776 ↗
  • MalfunctionMay 14, 2026No Display/Image

    IT WAS REPORTED THE COLONOVIDEOSCOPE HAD NO IMAGE. THE ISSUE WAS FOUND DURING THE MIDDLE OF AN ENDOSCOPY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25172375 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K192793. One submission often covers many models, so check the product and lot details in each record.

  • Z-1213-2024 · Jan 11, 2024 · Open, Classified
    Colonoscope, Model Number PCF-HQ190L.

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page