Skip to content

Isoflex™ 1944-46Recalls & reported problems

St. Jude Medical, Inc. · Pacing Lead · Class 3 · Implant

All models of Isoflex (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Isoflex pacing lead is implanted in the heart. Its 329 reports describe leads that move out of place, noise that is misread as heartbeats, and leads that fail to pace or detect heartbeats, and some describe lead replacement surgery. The reports include 7 deaths and 203 injuries. No recalls name this device. Reports fell to 13 in the last 12 months from 26 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
329
Deaths reported
7
Injuries reported
202
Recalls naming it
0

Summary

The FDA's MAUDE database contains 329 adverse event reports that list this exact device (DI 05414734502313), received between Jul 2, 2018 and Sep 28, 2026 — classified as 7 deaths, 202 injuries, 120 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (89 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 4 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 3 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated89
  • Over-Sensing77
  • Signal Artifact/Noise77
  • Failure to Capture48
  • Adverse Event Without Identified Device or Use Problem35
  • Failure to Sense32
  • Pacing Problem32
  • Capturing Problem26

Patient problems

  • No Clinical Signs, Symptoms or Conditions156
  • No Known Impact Or Consequence To Patient94
  • Unspecified Infection27
  • Discomfort15
  • No Consequences Or Impact To Patient14
  • Insufficient Information9
  • Death5
  • Muscle Stimulation3

Recent adverse event reports

The 25 most recent of 329 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Over-Sensing

    It was reported that the asymptomatic patient presented remotely via Merlin.net. Upon review of transmission, it was found that the Right Ventricular (RV) lead exhibited oversensing noise resulted in pacing inhibition and triggering noise reversion and High Ventricular Rate (HVR) episodes. No intervention was performed. Patient condition was stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916323 ↗
  • InjuryAug 28, 2026Device Dislodged or Dislocated

    It was reported that patient presented to clinic and was found to have a dislodged atrial lead. The atrial lead was explanted and replaced. The patient was discharged with no reports of adverse consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50492231 ↗
  • InjuryAug 25, 2026Failure to Capture

    It was reported that the patient presented to the hospital for a routine generator upgrade procedure. Prior to the procedure, interrogation of the device revealed non-capture on the right atrial (RA) lead. The RA lead was capped and replaced on (b)(6) 2026. The patient was in stable condition prior to surgery, during surgery and post-surgery.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50442534 ↗
  • InjuryJul 30, 2026Failure to Capture; Device Dislodged or Dislocated

    It was reported that the patient presented to the clinic. Upon device interrogation, the atrial lead exhibited loss of capture. The RV lead was found to be dislodged, which was confirmed by fluoroscopy. The atrial lead was subsequently explanted and replaced. The patient remained stable throughout the procedure.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50078642 ↗
  • InjuryJul 8, 2026Therapy Delivered to Incorrect Body Area; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW-UP IN CLINIC WITH A COMPLAINT OF DISCOMFORT. EVALUATION REVEALED INTERMITTENT STIMULATION OF THE PATIENT'S RIGHT PHRENIC NERVE BY THE RIGHT ATRIAL (RA) LEAD. FLUOROSCOPY DURING THE PROCEDURE REVEALED THAT THE RA LEAD HAD DISLODGED. THE RA LEAD WAS SUBSEQUENTLY EXPLANTED AND REPLACED. THE PATIENT REMAINED STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: Discomfort

    MAUDE report 25756732 ↗
  • InjuryApr 22, 2026Failure to Capture; Under-Sensing; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A DEVICE UPGRADE PROCEDURE, THE ATRIAL LEAD EXHIBITED UNDER SENSING AND FAILED TO CAPTURE. THE ATRIAL LEAD WAS EXPLANTED AND REPLACED WITH A NEW LEAD. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24950104 ↗
  • MalfunctionApr 1, 2026Difficult to Fold, Unfold or Collapse; Product Quality Problem; Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR AN IMPLANT PROCEDURE. DURING THE PROCEDURE, IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD COULD NOT BE IMPLANTED, POTENTIALLY DUE TO THE DISTAL SCREW AT THE LEAD TIP AND AN ISSUE WITH THE LEAD DESIGN. THE RA LEAD WAS NOT USED AND REPLACED DURING THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24750983 ↗
  • MalfunctionMar 26, 2026Over-Sensing

    IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED OVERSENSING NOISE. IT COULD NOT BE DETERMINED WHETHER THE NOISE WAS ATTRIBUTED TO LEAD NOISE OR ELECTROMAGNETIC INTERFERENCE ON THE PACEMAKER (PM). THE PM WAS EXPLANTED AND REPLACED ON (B)(6) 2026. THE PATIENT CONDITION WAS NOT KNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699133 ↗
  • InjuryFeb 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE EMERGENCY ROOM WITH THE SKIN OF THE DEVICE POCKET APPEARING RED, SWOLLEN. AN INFECTION AT THE IMPLANTED DEVICE POCKET SITE. THE DEVICE WAS FOUND VISIBLE FROM THE OPENED INCISION SITE OF THE IMPLANTED DEVICE POCKET. THE PHYSICIAN EXPLANTED THE ENTIRE SYSTEM ¿ PACEMAKER, RIGHT VENTRICULAR LEAD AND ATRIAL LEAD DUE TO INFECTION. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.

    Patient problems: Unspecified Infection

    MAUDE report 24301580 ↗
  • MalfunctionDec 24, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY TO THE CLINIC VIA MERLIN.NET TRANSMISSION. TRANSMISSIONS SHOWED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE OVERSENSING. NO INTERVENTION WAS REPORTED. THE PATIENT HAD NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23886236 ↗
  • InjuryDec 9, 2025Failure to Capture; Under-Sensing; Device Dislodged or Dislocated; Failure to Capture; Under-Sensing; Device Dislodged or Dislocated

    NEW INFORMATION RECEIVED INDICATES THAT UNDER-SENSING WAS NOTED ON THE RIGHT ATRIAL (RA) LEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23762704 ↗
  • InjuryNov 7, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT OVER-SENSED NOISE WAS PRESENT ON BOTH THE RIGHT ATRIAL (RA) AND RIGHT VENTRICULAR (RV) LEADS. THE LEADS WERE EXPLANTED AND REPLACED. THE PATIENT WAS NOTED AS BEING STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23502835 ↗
  • InjuryNov 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO CLINIC WITH REDNESS, INFECTION AND SKIN EROSION AT DEVICE POCKET. THE INFECTION WAS NOT ASSOCIATED WITH ANY ABBOTT PRODUCT AND/OR PROCEDURE. THE PACEMAKER, RIGHT ATRIAL LEAD AND RIGHT VENTRICULAR LEAD WERE EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.

    Patient problems: Unspecified Infection

    MAUDE report 23460745 ↗
  • InjurySep 11, 2025Break; Over-Sensing; Low impedance; Break; Over-Sensing; Low impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A LEAD REVISION PROCEDURE. PRIOR TO THE PROCEDURE, IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE OVERSENSING, LOW PACING IMPEDANCE AND INSULATION BREACH. INFORMATION RECEIVED INDICATED THE INSULATION BREACH WAS NOTED DUE TO NOISE ON THE ATRIAL LEAD AND A DECREASE IN IMPEDANCE. THE RA LEAD WAS CAPPED AND REPLACED (B)(6) 2025. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23023469 ↗
  • InjurySep 10, 2025Capturing Problem; Capturing Problem

    IT WAS REPORTED THAT THE PRESENTED TO THE HOSPITAL FOR ATRIAL LEAD REVISION PROCEDURE DUE TO HIGH CAPTURE THRESHOLD. AFTER THE PROCEDURE WAS STARTED, INFECTION WAS NOTED AT THE PATIENT'S DEVICE POCKET. THE PACEMAKER, THE RIGHT ATRIAL LEAD AND THE RIGHT VENTRICULAR LEAD WERE EXPLANTED DUE TO THE INFECTION AND WERE REIMPLANTED ON THE RIGHT SIDE. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: Unspecified Infection

    MAUDE report 23023032 ↗
  • InjurySep 9, 2025Failure to Capture; Device Dislodged or Dislocated; Failure to Capture; Device Dislodged or Dislocated

    NEW INFORMATION RECEIVED INDICATES THAT THE RA LEAD WAS EXPLANTED ON (B)(6) 2025. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23000578 ↗
  • DeathAug 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD DECEASED DUE TO AN UNKNOWN CAUSE. THERE IS NO KNOWN ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT SUGGESTS THAT THE DEATH WAS DEVICE RELATED. NO ADDITIONAL INFORMATION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22879788 ↗
  • InjuryAug 3, 2025Break; Over-Sensing; Pacing Problem; Break; Over-Sensing; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE HOSPITAL FOR LEAD REVISION. THE PATIENT'S RIGHT ATRIAL (RA) LEAD EXHIBITED OVER-SENSED NOISE RESULTING IN INAPPROPRIATE AUTOMATED MODE SWITCH (AMS) EPISODES DUE TO LEAD DAMAGE. THE RA LEAD WAS EXPLANTED AND REPLACED. THE PATIENT HAD NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22696641 ↗
  • InjuryJul 11, 2025Failure to Capture; Device Dislodged or Dislocated; Failure to Capture; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A RIGHT ATRIAL (RA) LEAD WAS DISLODGED. THE PHYSICIAN DECIDED TO EXPLANT THE RA LEAD AND A NEW LEAD WAS REPLACED. NO PATIENT SYMPTOMS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22477342 ↗
  • InjuryJun 25, 2025No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY VIA MERLIN. NET, THE RIGHT VENTRICLE (RV) EXHIBITED NOISE OVER SENSING RESULTED IN PACING INHIBITION. THE RIGHT ATRIAL (RA) LEAD HAD AN UNKNOWN MALFUNCTION. BOTH THE RA AND RV LEADS WERE CAPPED AND REPLACED ON (B)(6) 2025. THE PATIENT WAS RECOVERING POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22325966 ↗
  • MalfunctionJun 23, 2025Break; Over-Sensing; Break; Over-Sensing

    ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT LEAD DAMAGE WAS SUSPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22307864 ↗
  • InjuryMay 20, 2025Capturing Problem; Device Sensing Problem; Device Dislodged or Dislocated; Capturing Problem; Device Sensing Problem; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR INITIAL IMPLANT PROCEDURE. THE FOLLOWING DAY, HIGH CAPTURE THRESHOLD AND DECREASED SENSING WAS NOTED ON THE RIGHT ATRIAL (RA) LEAD. AN X-RAY WAS PERFORMED, AND DISLODGEMENT WAS CONFIRMED ON THE LEAD. THE PHYSICIAN ELECTED TO EXPLANT AND REPLACE THE RA LEAD. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22054054 ↗
  • InjuryMay 20, 2025Failure to Capture; Failure to Capture

    IT WAS REPORTED THAT A PATIENT PRESENTED TO HOSPITAL FOR A PACEMAKER UPGRADE TO CARDIAC RE-SYNCHRONIZATION THERAPY. DURING DEVICE-BASED TESTING IN PRE-OPERATION, IT WAS OBSERVED THE ATRIAL LEAD FAILED TO CAPTURE. CONCURRENTLY, THE PHYSICIAN ELECTED TO CAP AND REPLACED THE ATRIAL LEAD ON (B)(6) 2025. THE NEW PACEMAKER AND ATRIAL LEAD WERE SUCCESSFULLY IMPLANTED. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22050172 ↗
  • MalfunctionMay 9, 2025Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL FOR A GENERATOR REPLACEMENT PROCEDURE. AFTER THE POCKET WAS OPENED, IT WAS NOTED THAT THE RIGHT ATRIAL (RA) FAILED TO SENSE. THE RA LEAD WAS CAPPED AND REPLACED ON (B)(6) 2025. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21991489 ↗
  • InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL DUE TO INFECTION AT THE DEVICE SITE. THE PACEMAKER, RIGHT ATRIAL LEAD, RIGHT VENTRICULAR LEAD AND RIGHT VENTRICULAR CAPPED LEAD WERE EXPLANTED. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.

    Patient problems: Unspecified Infection

    MAUDE report 21931900 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page