Durata™ 7121Q-52Recalls & reported problems

St. Jude Medical, Inc. · Defibrillation Lead · Class 3 · Implant

Quick summary

This St. Jude Medical Durata 7121Q-52 is a defibrillator lead, a wire placed in the heart to deliver pacing and shock therapy. The 35 reports mostly describe leads picking up noise or signal from the body, or readings that were too high or too low, sometimes prompting clinicians to replace the lead. There are 23 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
23
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 05414734503211), received between Nov 19, 2018 and Dec 3, 2025 — classified as 23 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (15 reports).

3 other recalls cover devices cleared under the same FDA submission (P950022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing15
  • High impedance11
  • Inappropriate/Inadequate Shock/Stimulation9
  • Signal Artifact/Noise6
  • Low impedance4
  • Break2
  • Capturing Problem2
  • Device Sensing Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Known Impact Or Consequence To Patient4
  • Discomfort2
  • No Consequences Or Impact To Patient2
  • Shock from Patient Lead(s)2
  • Therapeutic Effects, Unexpected2
  • Insufficient Information1
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2025High impedance; High impedance

    IT WAS REPORTED THAT INTERMITTENT HIGH LEAD IMPEDANCE MEASUREMENTS ON THE RIGHT VENTRICULAR (RV) LEAD WAS NOTED VIA REVIEW OF REMOTE TRANSMISSION. THERE WERE INTERMITTENT OUT-OF-RANGE VALUES FOR THE RV TO SVC AND SVC TO CAN VECTORS WITHIN THE LAST 6 MONTHS. NO INTERVENTION WAS PERFORMED, AND NO PATIENT SYMPTOMS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23709278 ↗
  • InjuryDec 22, 2023High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICLE (RV) LEAD EXHIBITED HIGH PACING IMPEDANCES. THE LEAD WAS CAPPED AND REPLACED ON (B)(6) 2023. THE PATIENT WAS IN STABLE CONDITION PRIOR TO AND POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18387851 ↗
  • InjuryJun 27, 2023Capturing Problem; Capturing Problem

    NEW INFORMATION RECEIVED INDICATED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXTRACTED AND REPLACED DUE TO ELEVATED CAPTURE THRESHOLDS. THAT PATIENT WAS ASYMPTOMATIC BEFORE, DURING, AND AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17209159 ↗
  • InjuryMay 2, 2023Over-Sensing; Low impedance; Over-Sensing; Low impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR RIGHT VENTRICULAR (RV) LEAD AND RIGHT ATRIAL (RA) LEAD REVISION. IT WAS PREVIOUSLY NOTED THAT THE RV LEAD EXHIBITED NOISE OVER-SENSING AND LOW PACING IMPEDANCE AND THE RA LEAD EXHIBITED INAPPROPRIATE AUTOMATIC MODE SWITCH DUE TO NOISE OVER-SENSING. BOTH LEADS WERE CAPPED AND REPLACED ON (B)(6) 2023. PATIENT WAS EXPERIENCING LOW BLOOD PRESSURE DURING PROCEDURE. THE HYPOTENSION WAS RESOLVED, AND PATIENT CONDITION WAS STABLE POST PROCEDURE.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 16850908 ↗
  • MalfunctionApr 7, 2023Over-Sensing; Over-Sensing

    IT WAS REPORTED DURING REMOTE FOLLOW UP THAT THE RIGHT VENTRICULAR LEAD EXHIBITED OVERSENSING DUE TO NOISE. PROGRAMMING CHANGES WERE RECOMMENDED BUT NOT YET COMPLETED. THE PATIENT WAS STABLE AND ASYMPTOMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16704197 ↗
  • InjuryFeb 24, 2023Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED FAR P WAVE OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD RESULTING IN INAPPROPRIATE THERAPY. NO CHANGES OR INTERVENTION WAS REPORTED. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16434274 ↗
  • MalfunctionFeb 14, 2023High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA MERLIN.NET. REVIEW OF THE TRANSMISSION REVEALED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH IMPEDANCE. PROGRAMMING CHANGES WERE PERFORMED. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16375256 ↗
  • InjuryDec 21, 2022Capturing Problem; Capturing Problem

    IT WAS REPORTED THAT A PATIENT PRESENTED TO CLINIC FOR FOLLOW UP. DEVICE INTERROGATION REVEALED HIGH CAPTURE THRESHOLD ON THE RIGHT VENTRICULAR (RV) LEAD. ON (B)(6) 2022, THE PHYSICIAN ELECTED TO CAP AND REPLACE THE RV LEAD. THE PATIENT WAS DISCHARGED FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16024819 ↗
  • InjuryDec 21, 2022Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED DURING REMOTE FOLLOW UP THAT THE RIGHT VENTRICULAR LEAD HAD DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING. THIS INAPPROPRIATE DEFIBRILLATION THERAPY WAS DUE TO OVERSENSING NOISE. PATIENT CONDITION WAS UNKNOWN. THE LEAD WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16022982 ↗
  • MalfunctionApr 20, 2022Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW UP VIA MERLIN.NET WITH A RIGHT VENTRICULAR LEAD THAT EXHIBITED NOISE OVER-SENSING. THE PATIENT WAS BROUGH IN TO THE CLINIC; PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14164413 ↗
  • MalfunctionApr 13, 2022Signal Artifact/Noise; Break; Signal Artifact/Noise; Break

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A ROUTINE GENERATOR CHANGE PROCEDURE. DURING THE PROCEDURE, IT WAS NOTED THAT THE RIGHT VENTRICULAR LEAD HAD INSULATION DAMAGE. THE PHYSICIAN ELECTED TO REPAIR THE LEAD. THE PATIENT WAS IN STABLE CONDITION BEFORE, DURING, AND AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14107081 ↗
  • InjuryJan 20, 2022Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LEAD FRACTURE, HIGH PACING IMPEDANCE, AND OVERSENSED NOISE WHICH LED TO INAPPROPRIATE SHOCK. THE RV LEAD WAS CAPPED AND REPLACED ON (B)(6) 2022. THE PATIENT WAS DISCHARGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13309019 ↗
  • InjuryDec 16, 2021Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE IN-PATIENT PRESENTED WITH INAPPROPRIATE THERAPY. THE RIGHT VENTRICULAR (RV) LEAD WAS OVER-SENSING NOISE CAUSING INAPPROPRIATE ANTI-TACHYCARDIA PACING. THE PHYSICIAN CAPPED AND REPLACED THE RV LEAD ON (B)(6) 2021. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13020589 ↗
  • MalfunctionNov 22, 2021Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY VIA (B)(6). REVIEW OF THE TRANSMISSION REVEALED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED OVERSENSING NOISE. PROGRAMMING CHANGES WERE PERFORMED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12858166 ↗
  • MalfunctionNov 22, 2021Break; Break

    RELATED MANUFACTURER REFERENCE NUMBER: 2017865-2021-37390. IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR HAD MIGRATED. IT WAS NOTED THAT THE DEVICE HAD ¿FALLEN¿ AND WAS BELOW THE LEFT BREAST. DURING THE SCHEDULED PROCEDURE, IT WAS OBSERVED THAT THE RIGHT VENTRICULAR LEAD HAD A MINOR INSULATION BREACH. THE DEVICE WAS EXPLANTED, AND THE PHYSICIAN ELECTED TO PUT SOME MEDICAL ADHESIVE OVER THE ABRASION. THE PATIENT WAS IN RECOVERING CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12858039 ↗
  • InjuryNov 22, 2021Over-Sensing; Therapy Delivered to Incorrect Body Area; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Therapy Delivered to Incorrect Body Area; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT A PATIENT PRESENTED IN CLINIC FOR A FOLLOW-UP APPOINTMENT. THE PATIENT'S RIGHT VENTRICULAR (RV) WAS CAUSING EXTRACARDIAC STIMULATION AND OVERSENSING NOISE LEADING TO INAPPROPRIATE ATP. THE PATIENT WAS ASYMPTOMATIC. ON (B)(6) 2021, THE RV LEAD WAS CAPPED AND REPLACED. THE PATIENT WAS STABLE THROUGHOUT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12856199 ↗
  • InjurySep 9, 2021Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY ROOM (ER) DUE TO DISCOMFORT FROM INAPPROPRIATE SHOCKS. IT WAS NOTED THIS ANOMALY WAS DUE TO THE RIGHT VENTRICULAR (RV) LEAD OVER-SENSING NOISE. THE RV LEAD WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE THROUGHOUT THE PROCEDURE.

    Patient problems: Discomfort

    MAUDE report 12447651 ↗
  • MalfunctionSep 8, 2021Degraded; High impedance; Degraded; High impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A FOLLOW-UP IN CLINIC. DURING INTERROGATION, IT WAS OBSERVED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH DEFIBRILLATION LEAD IMPEDANCE, AND DEGRADATION. PROGRAMMING CHANGES WERE MADE. THE PATIENT WAS STABLE THROUGHOUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12443739 ↗
  • MalfunctionJul 9, 2021Defective Device; Defective Device

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

    Patient problems: Insufficient Information

    MAUDE report 12144691 ↗
  • InjuryJul 7, 2021Over-Sensing; Low impedance; Device Sensing Problem; Over-Sensing; Low impedance; Device Sensing Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOW PACING IMPEDANCE AND R-WAVE AMPLITUDE VARIATION. NO CHANGES OR INTERVENTION WAS PERFORMED; THE PATIENT WOULD CONTINUE TO BE MONITORED. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12129836 ↗
  • MalfunctionApr 13, 2021Low impedance; Low impedance

    IT WAS REPORTED THAT THE PATIENT PRESENTED REMOTELY TO THE CLINIC ON (B)(6) 2021. REVIEW OF TRANSMISSIONS REVEALED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD LOW DEFIBRILLATION LEAD IMPEDANCE. PHYSICIAN BROUGHT THE PATIENT IN CLINIC FOR A CHECK. DURING EXAMINATION OF RV LEAD, ALL THE FUNCTIONS WERE NORMAL AND THE LEAD DEFIBRILLATION IMPEDANCE WAS WITHIN RANGE. PHYSICIAN MADE PROGRAMMING CHANGES TO THE LEAD TO AVOID THIS ISSUE IN THE FUTURE. THE PATIENT WAS IN STABLE CONDITION BEFORE, DURING AND AFTER THE INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11655686 ↗
  • InjuryMar 30, 2021High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT WAS NOTIFIED VIA VIBRATORY ALERT ON AN UNKNOWN DATE OF AN EVENT OF HIGH DEFIBRILLATION IMPEDANCE ON THE RIGHT VENTRICULAR (RV) LEAD. ON (B)(6) 2021 THE PHYSICIAN CAPPED AND REPLACED THE RV LEAD TO RESOLVE THE ISSUE. THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11593233 ↗
  • InjuryMar 29, 2021Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED NOISE. THE LEAD WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: Discomfort; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11580574 ↗
  • InjuryFeb 11, 2021High impedance; High impedance

    ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE ISSUE WAS DISCOVERED IN CLINIC FOR A FOLLOW UP. THE PATIENT'S PACING IMPEDANCE WAS ELEVATED. THE PATIENT WAS STABLE THROUGHOUT THE EXPLANT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11318459 ↗
  • InjuryJan 25, 2021High impedance; High impedance

    IT WAS REPORTED THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH PACING IMPEDANCE. THE PATIENT PRESENTED IN CLINIC FOR A PROCEDURE ON (B)(6) 2021 WHERE THE LEAD WAS EXPLANTED AND REPLACED. THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11227383 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P950022. One submission often covers many models, so check the product and lot details in each record.

  • Z-0660-2016 · Nov 3, 2015 · Terminated
    Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Car
  • Z-0457-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co
  • Z-0458-2012 · Nov 28, 2011 · Terminated
    St. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous

Other Permanent defibrillator electrodes devices