OPTISURE™ LDP220Q-58Recalls & reported problems
St. Jude Medical, Inc. · Defibrillation Lead · Class 3 · Implant
All models of OPTISURE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 05414734507448), received between Jan 10, 2019 and Aug 27, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P950022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Damaged Prior to Use1
- Failure to Capture1
- High Capture Threshold1
- High impedance1
- Inappropriate/Inadequate Shock/Stimulation1
- Low impedance1
- Over-Sensing1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Shock from Patient Lead(s)1
- Syncope1
- Therapeutic Effects, Unexpected1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2024High impedance; Low impedance; High impedance; Low impedance
IT WAS REPORTED THAT A PATIENT PRESENTED REMOTELY WITH INSTANCES OF HIGH AND LOW DEFIBRILLATION IMPEDANCE VALUES WHEN COMPARED TO THE CHRONIC VALUE. NO INTERVENTION OR ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20078928 ↗MalfunctionMar 27, 2020Over-Sensing; Over-Sensing
NEW INFORMATION RECEIVED NOTES THAT POST PACED T-WAVE OVER-SENSING ON THE RIGHT VENTRICULAR (RV) LEAD REOCCURRED. NO INTERVENTION WAS PERFORMED. PATIENT CONDITION WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 9892262 ↗MalfunctionSep 13, 2019Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR LEAD EXHIBITED NO CAPTURE. PERFORMING A CHEST X-RAY AND LEAD REPLACEMENT WERE DISCUSSED. THE PATIENT WAS STABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9036996 ↗InjuryApr 15, 2019Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold
IT WAS REPORTED THAT A PATIENT PRESENTED IN THE EMERGENCY ROOM WITH AN IMPLANTABLE CARDIOVERTER DEVICE (ICD) THAT EXHIBITED EPISODES OF INAPPROPRIATE SHOCK AND ANTI TACHYCARDIA PACING (ATP). PATIENT COMPLAINED OF BRIEF SYNCOPE WITH EACH EPISODE. UPON REVIEW OF THE PATIENT'S ELECTROGRAM (EGM), CLINICIAN NOTED THAT THE PATIENT HAD ATRIAL FIBRILLATION (A-FIB) WITH RAPID VENTRICULAR RESPONSE (RVR), OCCASIONALLY WITH ABERRANCY. CLINICIAN WAS UNCLEAR WHETHER THIS WAS VENTRICULAR TACHYCARDIA (VT) OR A-FIB WITH ABERRANCY DUE TO STABLE, REGULAR RHYTHM WITH DIFFERENT MORPHOLOGY. CLINICIAN ALSO NOTED ELEVATED PACING THRESHOLD IN THE RIGHT VENTRICULAR LEAD. PROGRAMMING CHANGES WERE MADE. PATIENT WAS STABLE AFTER PROGRAMMING.
Patient problems: Shock from Patient Lead(s); Syncope; Therapeutic Effects, Unexpected
MAUDE report 8516503 ↗MalfunctionJan 10, 2019Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT DURING AN IMPLANT, ONE OF THE FINS WAS COMPLETELY DAMAGED ON THE TINED LEAD. THE LEAD WAS NOT USED. THE PATIENT CONDITION IS STABLE..
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8234015 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P950022. One submission often covers many models, so check the product and lot details in each record.
- Z-0660-2016 · Nov 3, 2015 · TerminatedOptisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Car
- Z-0457-2012 · Nov 28, 2011 · TerminatedSt. Jude Medical Riata (8Fr), Models: 1560, 1561, 1562, 1570, 1571, 1572, 1580, 1581, 1582, 1590, 1591, 1592 Product Usage: These Silicone transvenous high voltage (HV) leads co
- Z-0458-2012 · Nov 28, 2011 · TerminatedSt. Jude Medical Riata ST Silicone Endocardial Defibrillation Leads Riata(7Fr), Models: 7000, 7001, 7002, 7010, 7011, 7040, 7041, 7042 Product Usage: These Silicone transvenous
Other Permanent defibrillator electrodes devices
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes333 deaths30,581 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M62Medtronic, Inc. · Permanent defibrillator electrodes221 deaths14,405 reports
- Capsurefix® Novus 5076-52Medtronic, Inc. · Permanent defibrillator electrodes111 deaths9,860 reports
- CapSureFix Novus MRI™ SureScan® 5076-52Medtronic, Inc. · Permanent defibrillator electrodes80 deaths8,280 reports
- Sprint Quattro Secure S MRI™ SureScan™ 6935M55Medtronic, Inc. · Permanent defibrillator electrodes116 deaths7,750 reports
- Durata™ 7122Q-58St. Jude Medical, Inc. · Permanent defibrillator electrodes151 deaths6,202 reports
- Durata™ 7122Q-65St. Jude Medical, Inc. · Permanent defibrillator electrodes86 deaths4,773 reports
- Capsurefix® Novus 5076-58Medtronic, Inc. · Permanent defibrillator electrodes47 deaths4,587 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page