EP-WorkMate™ Recording SystemRecalls & reported problems
St. Jude Medical, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05415067000965), received between Jun 26, 2018 and May 9, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch1
- Device Displays Incorrect Message1
- Output Problem1
- Unexpected Shutdown1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 9, 2025Output Problem; Output Problem
DURING A PROCEDURE, THE WORKMATE CLARIS COMPUTER DISPLAYED AN ERROR MESSAGE AND THE PROCEDURE WAS CANCELLED. THE OLD WORKMATE SYSTEM WAS REPLACED BY THE NEW ENSITE X VERSION OF WORKMATE CLARIS. THE ERROR MESSAGE THAT DISPLAYED WAS ERROR MESSAGE: "162-SYSTEM OPTIONS NOT SET. THE SYSTEM CONFIGURATION HAS CHANGED SINCE THE LAST BOOT. ADDITION OF A HARD DRIVE, ETC., OR LOSS OF POWER TO THE REAL TIME CLOCK HAS OCCURRED. PRESSING F1 WILL RECORD THE NEW CONFIGURATION. IF THIS MESSAGE PERSIST, YOU MAY NEED TO REPLACE THE ONBOARD BATTERY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21995035 ↗InjuryMay 22, 2019Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
DURING A PROCEDURE THE COMPUTER DISPLAYED A "WIN32" ERROR MESSAGE AND THE SOFTWARE STOPPED RUNNING. THE PROCEDURE WAS UNABLE TO BE COMPLETED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8633603 ↗InjuryNov 21, 2018Unexpected Shutdown; Unexpected Shutdown
DURING A PROCEDURE, THE COMPUTER SHUT DOWN AFTER A FEW MINUTES. A REBOOT WAS ATTEMPTED MULTIPLE TIMES WITH NO RESOLUTION. THE PATIENT WAS PREPPED WITH LOCAL TEMPORAL SEDATION AND CATHETERS INSERTED. THE PROCEDURE WAS CANCELLED AND THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8094552 ↗InjuryJun 26, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
DURING PREPARATION FOR AN ATRIAL FIBRILLATION ABLATION PROCEDURE A CANCELLATION OCCURRED DUE TO A SOFTWARE LOCKUP. FOLLOWING IV PLACEMENT THE PC WOULD FREEZE AND SHUTDOWN. THE ISSUE WAS NOT RESOLVED AND THE PROCEDURE WAS RESCHEDULED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7637512 ↗
Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page