Simplicity™ IIIRecalls & reported problems
St. Jude Medical, Inc. · Disposable Radiofrequency Electrode · Class 2
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 05415067028976), received between Mar 14, 2019 and Jun 10, 2024 — classified as 7 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Material Separation” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation6
- Insufficient Heating3
- Temperature Problem3
- Communication or Transmission Problem1
- Fracture1
- High impedance1
- Impedance Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient8
- Foreign Body Embolism1
- Pain1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING A RADIOFREQUENCY ABLATION PROCEDURE ON (B)(6) 2024, THE PROBE WAS OVERHEATING AND PATIENT EXPERIENCED PAIN LOCAL. AS A RESULT, THE PROCEDURE WAS ABORTED..
Patient problems: Pain
MAUDE report 19506659 ↗InjuryNov 8, 2023Insufficient Heating; Insufficient Heating
IT WAS REPORTED DURING A PROCEDURE THE GENERATOR INDICATED AN ERROR THAT THE TEMPERATURE WAS RISING TOO SLOWLY. THE CASE WAS ABANDONED WITH NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18091054 ↗InjuryJul 17, 2022Fracture; Fracture
IT WAS REPORTED THE DEVICE BROKE WHILE THE PHYSICIAN WAS REPOSITIONING IT INSIDE THE PATIENT. THE PIECE THAT BROKE OFF INTO THE PATIENT WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15036807 ↗InjuryOct 4, 2021Material Separation; Material Separation
DURING A SACRAL RADIOFREQUENCY ABLATION PROCEDURE, THE PROBE FRACTURED AND SEPARATED IN THE PATIENT. AN 18G NEEDLE WAS USED TO TURN THE BROKEN PROBE AND THE BROKEN PROBE WAS REMOVED. THE PROBE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: Foreign Body Embolism
MAUDE report 12569088 ↗MalfunctionJul 12, 2021High impedance; Communication or Transmission Problem; High impedance; Communication or Transmission Problem
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING DETACHED ELECTRODES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12149297 ↗MalfunctionJun 16, 2021Impedance Problem; Impedance Problem
THIS REPORT IS TO ADVISE OF AN EVENT NOTED DURING ANALYSIS WHICH REVEALED A FRACTURED ELECTRODE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12007653 ↗MalfunctionMay 13, 2021Material Separation; Material Separation
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED ELECTRODE FROM THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11820459 ↗MalfunctionFeb 11, 2021Material Separation; Material Separation
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING DETACHED ELECTRODES ON THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11314615 ↗InjurySep 4, 2020Output Problem; Output Problem
RELATED MANUFACTURING REF: 2182269-2020-00081 DURING A SIMPLICITY RF PROCEDURE WITH THE PATIENT ON THE TABLE AND PREPPED, AN ERROR MESSAGE WAS NOTED ON PORTS ONE AND TWO WHILE CONNECTED TO THE GENERATOR. AN ATTEMPT TO SWITCH PORT 1 AND 2 INTO PORT 3 WERE UNSUCCESSFUL AS THE ERRORS REMAINED AND THE PROCEDURE WAS CANCELLED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10493606 ↗MalfunctionFeb 20, 2020Material Deformation; Material Deformation
DURING A PROCEDURE, TWO OF THE THREE THERMOCOUPLE MODULES ON THE SIMPLICITY PROBE WERE DETACHED DURING THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9733334 ↗MalfunctionAug 29, 2019Material Separation; Material Separation
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8944453 ↗MalfunctionAug 29, 2019Material Separation; Material Separation
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8944437 ↗MalfunctionAug 29, 2019Material Separation; Material Separation
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8944425 ↗InjuryMar 14, 2019Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem
RELATED MANUFACTURER REFERENCE: 2182269-2019-00028. DURING A PROCEDURE, THE PROBES WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A THIRD PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 8422703 ↗InjuryMar 14, 2019Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem
RELATED MANUFACTURER REFERENCE: 2182269-2019-00029. DURING A PROCEDURE, THE PROBES WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A THIRD PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 8422700 ↗InjuryMar 14, 2019Temperature Problem; Temperature Problem
DURING A PROCEDURE, THE PROBE WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A SECOND PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8422693 ↗
Other PROBE, RADIOFREQUENCY LESION devices
- NA TCN-15Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls10 reports
- NA TCN-10-3MBoston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls9 reports
- NA CSK-TC10Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall1 death58 reports
- NA CLASSICSt. Jude Medical, Inc. · PROBE, RADIOFREQUENCY LESION1 recall40 reports
- Intracept RLV 0055Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall30 reports
- Intracept RLV 0045Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall1 death23 reports
- Intracept RLV0065Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall14 reports
- Intracept RLV 0075Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.