Simplicity™ IIIRecalls & reported problems

St. Jude Medical, Inc. · Disposable Radiofrequency Electrode · Class 2

Adverse event reports
16
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 05415067028976), received between Mar 14, 2019 and Jun 10, 2024 — classified as 7 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Material Separation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation6
  • Insufficient Heating3
  • Temperature Problem3
  • Communication or Transmission Problem1
  • Fracture1
  • High impedance1
  • Impedance Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient8
  • Foreign Body Embolism1
  • Pain1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY ABLATION PROCEDURE ON (B)(6) 2024, THE PROBE WAS OVERHEATING AND PATIENT EXPERIENCED PAIN LOCAL. AS A RESULT, THE PROCEDURE WAS ABORTED..

    Patient problems: Pain

    MAUDE report 19506659 ↗
  • InjuryNov 8, 2023Insufficient Heating; Insufficient Heating

    IT WAS REPORTED DURING A PROCEDURE THE GENERATOR INDICATED AN ERROR THAT THE TEMPERATURE WAS RISING TOO SLOWLY. THE CASE WAS ABANDONED WITH NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18091054 ↗
  • InjuryJul 17, 2022Fracture; Fracture

    IT WAS REPORTED THE DEVICE BROKE WHILE THE PHYSICIAN WAS REPOSITIONING IT INSIDE THE PATIENT. THE PIECE THAT BROKE OFF INTO THE PATIENT WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15036807 ↗
  • InjuryOct 4, 2021Material Separation; Material Separation

    DURING A SACRAL RADIOFREQUENCY ABLATION PROCEDURE, THE PROBE FRACTURED AND SEPARATED IN THE PATIENT. AN 18G NEEDLE WAS USED TO TURN THE BROKEN PROBE AND THE BROKEN PROBE WAS REMOVED. THE PROBE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: Foreign Body Embolism

    MAUDE report 12569088 ↗
  • MalfunctionJul 12, 2021High impedance; Communication or Transmission Problem; High impedance; Communication or Transmission Problem

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING DETACHED ELECTRODES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12149297 ↗
  • MalfunctionJun 16, 2021Impedance Problem; Impedance Problem

    THIS REPORT IS TO ADVISE OF AN EVENT NOTED DURING ANALYSIS WHICH REVEALED A FRACTURED ELECTRODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12007653 ↗
  • MalfunctionMay 13, 2021Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED ELECTRODE FROM THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11820459 ↗
  • MalfunctionFeb 11, 2021Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING DETACHED ELECTRODES ON THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11314615 ↗
  • InjurySep 4, 2020Output Problem; Output Problem

    RELATED MANUFACTURING REF: 2182269-2020-00081 DURING A SIMPLICITY RF PROCEDURE WITH THE PATIENT ON THE TABLE AND PREPPED, AN ERROR MESSAGE WAS NOTED ON PORTS ONE AND TWO WHILE CONNECTED TO THE GENERATOR. AN ATTEMPT TO SWITCH PORT 1 AND 2 INTO PORT 3 WERE UNSUCCESSFUL AS THE ERRORS REMAINED AND THE PROCEDURE WAS CANCELLED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10493606 ↗
  • MalfunctionFeb 20, 2020Material Deformation; Material Deformation

    DURING A PROCEDURE, TWO OF THE THREE THERMOCOUPLE MODULES ON THE SIMPLICITY PROBE WERE DETACHED DURING THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9733334 ↗
  • MalfunctionAug 29, 2019Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8944453 ↗
  • MalfunctionAug 29, 2019Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8944437 ↗
  • MalfunctionAug 29, 2019Material Separation; Material Separation

    THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS CONFIRMING A DETACHED COMPONENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8944425 ↗
  • InjuryMar 14, 2019Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem

    RELATED MANUFACTURER REFERENCE: 2182269-2019-00028. DURING A PROCEDURE, THE PROBES WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A THIRD PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 8422703 ↗
  • InjuryMar 14, 2019Insufficient Heating; Temperature Problem; Insufficient Heating; Temperature Problem

    RELATED MANUFACTURER REFERENCE: 2182269-2019-00029. DURING A PROCEDURE, THE PROBES WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A THIRD PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 8422700 ↗
  • InjuryMar 14, 2019Temperature Problem; Temperature Problem

    DURING A PROCEDURE, THE PROBE WOULD NOT HEAT TO TEMPERATURE AND THERE WAS A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. CABLES AND SETTINGS WERE CHANGED WITH NO RESOLUTION. A SECOND PROBE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8422693 ↗

Other PROBE, RADIOFREQUENCY LESION devices

All PROBE, RADIOFREQUENCY LESION devices →