NA RFDEUK-10Recalls & reported problems
St. Jude Medical, Inc. · Disposable Radiofrequency Electrode · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05415067028990), received between Sep 1, 2022 and Sep 1, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Low impedance” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Low impedance6
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00199, 2184149-2022-00200, 2184149-2022-00201, 2184149-2022-00202. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING. THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337392 ↗MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00200, 2184149-2022-00201, 2184149-2022-00202, 2184149-2022-00203. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING. THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337360 ↗MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00199, 2184149-2022-00200, 2184149-2022-00202, AND 2184149-2022-00203. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING. THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337331 ↗MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00199. 2184149-2022-00201, 2184149-2022-00202, 2184149-2022-00203. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING. THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337325 ↗MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00199, 2184149-2022-00200, 2184149-2022-00201, 2184149-2022-00203. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING. THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337319 ↗MalfunctionSep 1, 2022Low impedance; Low impedance
RELATED MANUFACTURER REPORT NUMBERS: 2184149-2022-00197, 2184149-2022-00199, 2184149-2022-00200, 2184149-2022-00201, 2184149-2022-00202, 2184149-2022-00203. DURING THE PROCEDURE, THERE WAS A MISSING IMPEDANCE ERROR MESSAGE, AND THE GENERATOR STOPPED WORKING AFTER THE SENSORY AND MOTOR TESTING.THE PROCEDURE COULD NOT BE COMPLETED. HOWEVER, AFTER THE PROCEDURE, DURING TROUBLESHOOTING, THE ELECTRODE WAS REPLACED AND THE GENERATOR WAS FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337288 ↗
Other PROBE, RADIOFREQUENCY LESION devices
- NA TCN-15Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls10 reports
- NA TCN-10-3MBoston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION2 recalls9 reports
- NA CSK-TC10Boston Scientific Neuromodulation Corporation · PROBE, RADIOFREQUENCY LESION1 recall1 death58 reports
- NA CLASSICSt. Jude Medical, Inc. · PROBE, RADIOFREQUENCY LESION1 recall40 reports
- Intracept RLV 0055Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall30 reports
- Intracept RLV 0045Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall1 death23 reports
- Intracept RLV0065Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall14 reports
- Intracept RLV 0075Relievant Medsystems, Inc. · Probe, Radiofrequency Lesion1 recall3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.