EnSite™ ENSITE-FF-4.5C-01Recalls & reported problems

St. Jude Medical, Inc. · Field Frame Cable · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067032232), received between Oct 23, 2025 and Nov 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Thermal Decomposition of Device; Thermal Decomposition of Device

    DURING PREPARATION FOR AN ATRIAL FIBRILLATION PROCEDURE, THE FIELD FRAME WAS PLUGGED IN INCORRECTLY, AND AFTER THIS THE CABLE AND PLUG OF THE FIELD FRAME WAS BURNED. THE FIELD FRAME AND FIELD FRAME CABLE WERE REPLACED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513567 ↗
  • MalfunctionOct 23, 2025Material Twisted/Bent; Material Twisted/Bent

    DURING AN ATRIAL FIBRILLATION PROCEDURE, THERE WAS A COMMUNICATION ISSUE WITH THE AMPLIFIER AND A BENT PIN ON THE FIELD FRAME CABLE RESULTING IN CANCELLATION OF THE PROCEDURE. THE FIELD FRAME GENERATOR WAS NOT ABLE TO CONNECT TO THE AMPLIFIER. DESPITE MULTIPLE RESTARTS OF THE AMPLIFIER AND DISPLAY WORKSTATION, AND AFTER VERIFYING ALL CONNECTIONS, IT WAS FOUND THAT ONE OF THE PINS ON THE FIELD FRAME CABLE WAS BROKEN. THE CABLE AND FIELD FRAME WERE EXCHANGED BUT THE ISSUE DID NOT RESOLVE. THE SYSTEM WAS SWITCHED TO NAVX MODE, HOWEVER THE CATHETERS WERE NOT VISIBLE. IT IS ALLEGED THAT THE AMPLIFIER WAS CAUSING THE ISSUE. DUE TO THE ISSUE NOT BEING RESOLVED, THE PROCEDURE WAS CANCELLED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23376241 ↗