EnSite™ ENSITE-PRSFRNT-01Recalls & reported problems

St. Jude Medical, Inc. · Patient Reference Sensor · Class 2

All models of EnSite (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05415067033284), received between Apr 5, 2023 and Mar 6, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2025Communication or Transmission Problem; Communication or Transmission Problem

    DURING A CONGENITAL ATRIOVENTRICULAR REENTRANT TACHYCARDIA PROCEDURE, THERE WAS A DELAY DUE TO A COMMUNICATION ISSUE WITH THE PRS CABLES. THE PRS WAS INITIALLY VISIBLE BUT DID NOT STAY VISIBLE. THE SYSTEM WAS REBOOTED, AND THE PRS-A CABLE AND FIELD FRAME WERE BOTH MOVED AND EXCHANGED, BUT THE ISSUE PERSISTED. THE PRS-P CABLE COULD NOT BE MOVED DUE TO GA. THE CASE WAS REVERTED TO NAVX TO SUCCESSFULLY PERFORM THE PROCEDURE WITHOUT ADVERSE CONSEQUENCES. THE PRS SENSORS WERE TESTED AFTER THE PROCEDURE, AND EXCHANGING THE PRS-P CABLE MADE THE SENSORS VISIBLE WITH SOME FLICKERING. EXCHANGING THE PRS-A CABLE MADE SENSORS VISIBLE. BOTH CABLES SEEMED TO CAUSE AN ISSUE AS THE PRS-P CAUSED A SENSOR ISSUE, AND THE PRS-A WOULD INTERMITTENTLY DO THE SAME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21542485 ↗
  • MalfunctionApr 5, 2023Communication or Transmission Problem; Communication or Transmission Problem

    DURING DEVICE PREPARATION FOR A PREMATURE VENTRICULAR CONTRACTIONS PROCEDURE, AFTER THE PATIENT WAS PREPPED FOR THE PROCEDURE, IT WAS REPORTED THAT THERE WAS A PRS A NOT CONNECTED MESSAGE RESULTING IN A DELAY. DISCONNECTING, AND RECONNECTING DID NOT RESOLVE THE ISSUE. SWAPPING PORTS DID NOT RESOLVE THE ISSUE EITHER. THE ISSUE WAS THEN RESOLVED BY USING A PRS FROM A DIFFERENT HOSPITAL. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16686677 ↗

Other Computer, diagnostic, programmable devices