Skip to content

UroMax Ultra M0062251080Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2

All models of UroMax Ultra (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729341185), received between Jan 15, 2019 and Feb 11, 2025 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation4
  • Material Puncture/Hole3
  • Leak/Splash2
  • Burst Container or Vessel1
  • Inflation Problem1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AUROMAX ULTRA DILATION BALLOON CATHETER WAS ABOUT TO BE USED DURING A RETROGRADE INTRARENAL SURGERY (RIRS). DURING PREPARATION, IT WAS FOUND THAT THE BALLOON WAS TORN AS SOON AS THEY REMOVED THE PACKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21356529 ↗
  • MalfunctionSep 4, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING PREPARATION A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS FOUND BENT AT THE END OF THE BALLOON WHERE IT WAS ROLLED UP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20141069 ↗
  • MalfunctionMay 3, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS USED DURING A RETROGRADE INTRARENAL SURGERY (RIRS) PROCEDURE. DURING THE PROCEDURE, THE CONTRAST MEDIA WAS LEAKED FROM THE BALLOON CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A PINHOLE IN THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19243678 ↗
  • MalfunctionMar 11, 2024Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS USED DURING A URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CATHETER SHOWS BENDING DAMAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18882208 ↗
  • MalfunctionFeb 20, 2024Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS TO BE USED DURING A URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2023. DURING PREPARATION, THE BALLOON EXPANSION DID NOT WORK. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18741415 ↗
  • MalfunctionAug 22, 2023Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS TO BE USED DURING A URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2023. DURING PREPARATION, THE CATHETER WAS BENT IN THE MIDDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. THE DEVICE WAS NOT RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17599279 ↗
  • MalfunctionDec 19, 2022Leak/Splash; Leak/Splash

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON CATHETER WAS USED IN THE RIGHT URETER DURING AN URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS LEAKING CONTRAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16003424 ↗
  • MalfunctionFeb 24, 2021Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS USED IN THE RIGHT URETER DURING A URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE CONTRAST WAS INJECTED TO THE BALLOON IT WAS NOTICED THAT THERE WAS NO PRESSURE. THE C-ARM WAS CHECKED AND NOTED TO HAVE A LEAK FROM A SMALL HOLE THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11372349 ↗
  • MalfunctionAug 12, 2020Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON CATHETER WAS USED IN THE RIGHT URETER DURING A UTEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON HAD A FINE HOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10395229 ↗
  • MalfunctionJul 23, 2019Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA URETERAL DILATATION BALLOON WAS USED IN THE RIGHT URETER DURING A URETERAL BALLOON DILATION PERFORMED ON (B)(6), 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST AS SOON AS THE PRESSURE WAS GRADUALLY INCREASED TO EXPAND THE BALLOON.THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA URETERAL DILATATION BALLOON DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8815831 ↗
  • MalfunctionJan 15, 2019Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON DEVICE WAS USED IN THE LEFT URETER DURING A URETERAL STRICTURE BALLOON DILATION PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, IT WAS NOTED THAT THE CATHETER WAS KINKED. ADDITIONALLY, IT WAS ALSO NOTED THAT CATHETER WOULD NOT PASS THROUGH THE GUIDEWIRE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8246924 ↗

Other DILATOR, URETHRAL devices

All DILATOR, URETHRAL devices →

Sources and updates