Autotome™ RX 49 M00545160Recalls & reported problems
Boston Scientific Corporation · Cannulating Sphincterotome · Class 2
All models of Autotome RX 49 (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Autotome RX 49 cannulating sphincterotome is used to cut tissue during endoscopy procedures such as ERCP. Reports mostly describe the metal cutting wire breaking during the procedure, plus some bent or deformed tips and difficulty positioning the device. There are 30 reports, with no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 08714729415152), received between Feb 14, 2019 and Dec 2, 2025 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Break” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break25
- Adverse Event Without Identified Device or Use Problem5
- Material Deformation3
- Use of Device Problem3
- Material Twisted/Bent2
- Positioning Problem2
- Appropriate Device Problem Term/Code Not Available1
- Failure to Align1
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- No Consequences Or Impact To Patient7
- No Known Impact Or Consequence To Patient4
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN A PROCEDURE ON (B)(6) 2025. IT WAS REPORTED THAT AFTER PERFORMING THE SPHINCTEROTOMY, AS THE DEVICE WAS REMOVED, IT WAS NOTICED THAT THE METAL CUTTING WIRE WAS BROKEN. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23698386 ↗MalfunctionSep 22, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS ATTEMPTED TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILIARY TRACT LITHOTOMY PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TIP WAS FRACTURED DURING THE INCISION OF THE PAPILLA OF THE DUODENAL SPHINCTER. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 49. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23113748 ↗MalfunctionSep 12, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS ATTEMPTED TO BE USED IN THE INTESTINAL TRACT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, NORMAL INTUBATION WAS PERFORMED. WHILE CUTTING THE PAPILLA WITH THE GENERATOR ENERGIZED, THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23044106 ↗MalfunctionAug 12, 2025Failure to Conduct; Material Deformation; Adverse Event Without Identified Device or Use Problem; Failure to Conduct; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DEVICE CONTACT WAS POOR AND THE POWER SUPPLY WAS NOT CONTINUOUS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF DEVICE CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786322 ↗MalfunctionOct 18, 2024Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE DUODENAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE WAS BROKEN. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 49. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THE COMPLAINANT REPORTED THAT THE DEVICE WAS ENERGIZED PRIOR TO BOWING. HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), PRECAUTION: THE SPHINCTEROTOME DOES NOT NEED TO BE ENERGIZED PRIOR TO PERFORMING SPHINCTEROTOMY. ENERGIZING THE CUTTING WIRE PRIOR TO USE WILL CAUSE PREMATURE CUTTING WIRE FATIGUE AND WILL COMPROMISE THE CUTTING WIRE'S INTEGRITY. ADDITIONAL INFORMATION RECEIVED ON OCTOBER 21, 2024: IT WAS REPORTED THAT THE DEVICE WAS NOT ENERGIZED PRIOR TO BOWING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20482231 ↗MalfunctionJan 2, 2024Break; Material Deformation; Material Split, Cut or Torn; Break; Material Deformation; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE PAPILLA DURING A PROCEDURE ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE DEVICE WAS ENERGIZED INITIALLY, THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427595 ↗MalfunctionAug 17, 2023Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2023. DURING THE PROCEDURE, THE DEVICE WAS ENERGIZED, AND WHEN THE PAPILLA WAS BEING CUT, THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17565824 ↗MalfunctionAug 17, 2022Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE PAPILLA OF VATER DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE. THE EXACT PROCEDURE DATE WAS UNKNOWN. DURING THE PROCEDURE, THE CUTTING WIRE BROKE OFF FROM THE CATHETER. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE EXPOSED CUTTING WIRE HAD A LITTLE INSIDE THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED. HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), "WARNING: VERIFY THAT THE CUTTING WIRE HAS EXITED THE ENDOSCOPE BY VISUALIZING IT ON THE ENDOSCOPE MONITOR. FAILURE TO DO SO MAY RESULT IN CONTACT BETWEEN THE CUTTING WIRE AND THE ENDOSCOPE WHILE ELECTRICAL CURRENT IS APPLIED. THIS MAY CAUSE GROUNDING, WHICH CAN RESULT IN PATIENT INJURY, OPERATOR INJURY, A BROKEN CUTTING WIRE, AND/OR DAMAGE TO THE ENDOSCOPE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15244001 ↗MalfunctionJun 1, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14565187 ↗MalfunctionApr 1, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE CHOLEDOCHAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN THE HANDLE WAS PULLED TO PERFORM SPHINCTEROTOMY IN THE CHOLEDOCHAL PAPILLA, THE CUTTING WIRE BROKE INSIDE THE PAPILLA. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13973500 ↗MalfunctionJan 20, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS INTENDED TO BE USED IN THE BILE DUCT IN AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING PREPARATION OUTSIDE THE PATIENT, THE CUTTING WIRE OF THE AUTOTOME RX 49 FRACTURED. IT WAS REPORTED THAT THERE WAS A BREAK IN THE CUTTING WIRE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13309005 ↗MalfunctionNov 7, 2021Failure to Align; Failure to Align
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE CUTTING WIRE WOULD NOT RELEASE THE BOW AFTER IT WAS BEING BOWED. IT WAS REPORTED THAT THERE WAS NO ANY VISIBLE DAMAGE NOTED ON THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12759525 ↗MalfunctionAug 23, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12355870 ↗MalfunctionAug 23, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12355775 ↗MalfunctionAug 10, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHILE CUTTING THROUGH THE PAPILLA, THE PHYSICIAN NOTICED THAT THE CUTTING WIRE OF THE AUTOTOME RX 49 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12304693 ↗MalfunctionJun 2, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, AS THE INCISION OF THE PAPILLA WAS PERFORMED, THE CUTTING WIRE BROKE. NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11920599 ↗MalfunctionMay 6, 2021Positioning Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC BILE DUCT STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN ATTEMPTING TO PERFORM ENDOSCOPIC SPHINCTEROTOMY (EST) IT WAS NOTICED THAT THE "BLADE DIRECTION" OF THE DEVICE WAS INCORRECT. EST WAS NOT PERFORMED, AND A TUBE STENT WAS PLACED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11782980 ↗MalfunctionApr 30, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHILE THE PAPILLA WAS CUT, THE CUTTING WIRE OF AUTOTOME RX 49 WAS BURNT AND MELTED INTO TWO PIECES. NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11754290 ↗MalfunctionApr 29, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE PHYSICIAN CUT THE PAPILLA, THE CUTTING WIRE OF AUTOTOME RX 49 WAS BURNT AND HAD DETACHED. NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11745539 ↗MalfunctionApr 5, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, AS THE PAPILLA IS BEING CUT, THE CUTTING WIRE BROKE. NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. REPORTEDLY, THE DEVICE WAS NOT ENERGIZED PRIOR TO BOWING AND WHEN NOT IN CONTACT WITH THE TISSUE. ALSO, THE EXPOSED CUTTING WIRE HAD FULLY EXITED THE ENDOSCOPE WHEN THE DEVICE WAS ENERGIZED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11613886 ↗MalfunctionNov 19, 2020Use of Device Problem; Material Twisted/Bent; Positioning Problem; Use of Device Problem; Material Twisted/Bent; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN A VERMIFORM APPENDIX DURING AN ENDOSCOPIC RETROGRADE APPENDICITIS THERAPY (ERAT) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE DEVICE WAS ENTERED INTO THE ILEOCECAL VALVE TO PREPARE FOR CANNULATION, THE DEVICE WAS FOUND TWISTED. ADDITIONALLY, THE ORIENTATION OF THE CATHETER TIP WAS INCORRECT WHEN THE DEVICE EXITED THE SCOPE AND THE CUTTING WIRE ORIENTATION WAS INCORRECT WITH RESPECT TO THE PLASTIC TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: ACCORDING TO THE COMPLAINANT, THE AUTOTOME RX 49 WAS USED IN THE VERMIFORM APPENDIX. HOWEVER, PER THE AUTOTOME INSTRUCTIONS FOR USE, THE DEVICE IS INDICATED FOR USE SELECTIVE CANNULATION OF THE COMMON BILE DUCTS (CBD) AND THE TRANSENDOSCOPIC SPHINCTEROTOMY OF THE PAPILLA OF VATER AND/OR THE SPHINCTER OF ODDI. THE DEVICE IS NOT INDICATED TO BE USED IN THE APPENDIX. ***ADDITIONAL INFORMATION NOVEMBER 29, 2020*** REPORTEDLY, THE TYPE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10871516 ↗MalfunctionJul 22, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CUTTING WIRE BROKE. REPORTEDLY, NO PART OF THE DEVICE WAS DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10309240 ↗MalfunctionJul 16, 2020Material Integrity Problem; Appropriate Device Problem Term/Code Not Available; Material Integrity Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PHYSICIAN NOTED A VISIBLE CRACK AT THE TIP OF THE DEVICE. THE PROCEDURE WAS NOT COMPLETED AS THE PATIENT COULD NOT ENDURE THE PROLONGED PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10284028 ↗MalfunctionMay 20, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 49 WAS USED IN PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THERE WAS A BREAK IN THE CUTTING WIRE WHEN IT WAS PULLED. NOTHING DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10075720 ↗MalfunctionJan 29, 2020Break; Break
NOTE: THIS REPORT PERTAINS TO A JAGTOME RX 49 AND AUTOTOME RX 49 DEVICES USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGTOME RX 49 AND AUTOTOME RX 49 WAS USED IN THE PAPILLA DURING EXTRACTION OF BILIARY STONE PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN CUT THE PAPILLA USING THE JAGTOME RX 49, THE CUTTING WIRE WAS BROKE AFTER TWO SECONDS OF CUTTING. THE PHYSICIAN USED ANOTHER SPHINCTEROTOME AN AUTOTOME RX 49 AND THE SAME PROBLEM OCCURRED, THE CUTTING WIRE BROKE. REPORTEDLY, NO PART OF THE DEVICE WAS DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A THIRD JAGTOME RX 49. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9642411 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page